{"title":"光比色计校准功能","authors":"I. Zakharov, O. Botsiura, A.Yu. Pakhomova","doi":"10.1109/MMA52675.2021.9610950","DOIUrl":null,"url":null,"abstract":"The paper considers the procedure of photocolorimeter calibration using the example of measuring the mass concentration of total iron in water. The color intensity of the solution, proportional to the mass concentration of iron, is measured using a photocolorimeter at a wavelength of 400–430 nm. The main steps of photocolorimeter calibration are: preparation of calibration solutions; measuring the concentration of calibration solutions with a photocolorimeter, obtaining the calibration dependence; identification of the calibration dependence; using the calibration dependence to find the concentration of the test solution of total iron in water. To obtain calibration solutions, the state standard sample is used, which is diluted with water using pipettes and measuring flasks to the required concentration. The uncertainty of identification of the calibration dependence and its use for finding the concentration of the studied solution of total iron in water is evaluated.","PeriodicalId":287017,"journal":{"name":"2021 XXXI International Scientific Symposium Metrology and Metrology Assurance (MMA)","volume":"29 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2021-09-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Photocolorimeter calibration features\",\"authors\":\"I. Zakharov, O. Botsiura, A.Yu. Pakhomova\",\"doi\":\"10.1109/MMA52675.2021.9610950\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The paper considers the procedure of photocolorimeter calibration using the example of measuring the mass concentration of total iron in water. The color intensity of the solution, proportional to the mass concentration of iron, is measured using a photocolorimeter at a wavelength of 400–430 nm. The main steps of photocolorimeter calibration are: preparation of calibration solutions; measuring the concentration of calibration solutions with a photocolorimeter, obtaining the calibration dependence; identification of the calibration dependence; using the calibration dependence to find the concentration of the test solution of total iron in water. To obtain calibration solutions, the state standard sample is used, which is diluted with water using pipettes and measuring flasks to the required concentration. The uncertainty of identification of the calibration dependence and its use for finding the concentration of the studied solution of total iron in water is evaluated.\",\"PeriodicalId\":287017,\"journal\":{\"name\":\"2021 XXXI International Scientific Symposium Metrology and Metrology Assurance (MMA)\",\"volume\":\"29 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-09-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2021 XXXI International Scientific Symposium Metrology and Metrology Assurance (MMA)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/MMA52675.2021.9610950\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2021 XXXI International Scientific Symposium Metrology and Metrology Assurance (MMA)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/MMA52675.2021.9610950","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本文以测量水中总铁的质量浓度为例,讨论了光比色计的标定过程。溶液的颜色强度与铁的质量浓度成正比,使用波长为400-430 nm的光比色计进行测量。光比色计校准的主要步骤是:校准溶液的制备;用光比色计测量校准溶液的浓度,得到校准依赖关系;标定相关性的识别;利用校准依赖关系求出水中总铁测试溶液的浓度。为了得到校准溶液,使用国家标准样品,用移液管和量瓶用水稀释到所需的浓度。评价了标定依赖性鉴定的不确定度及其用于测定所研究的水中总铁溶液浓度的不确定度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Photocolorimeter calibration features
The paper considers the procedure of photocolorimeter calibration using the example of measuring the mass concentration of total iron in water. The color intensity of the solution, proportional to the mass concentration of iron, is measured using a photocolorimeter at a wavelength of 400–430 nm. The main steps of photocolorimeter calibration are: preparation of calibration solutions; measuring the concentration of calibration solutions with a photocolorimeter, obtaining the calibration dependence; identification of the calibration dependence; using the calibration dependence to find the concentration of the test solution of total iron in water. To obtain calibration solutions, the state standard sample is used, which is diluted with water using pipettes and measuring flasks to the required concentration. The uncertainty of identification of the calibration dependence and its use for finding the concentration of the studied solution of total iron in water is evaluated.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信