血管假体的动态内部顺应性。

ASAIO transactions Pub Date : 1991-07-01
R Tu, J McIntyre, C Hata, C L Lu, E Wang, R C Quijano
{"title":"血管假体的动态内部顺应性。","authors":"R Tu,&nbsp;J McIntyre,&nbsp;C Hata,&nbsp;C L Lu,&nbsp;E Wang,&nbsp;R C Quijano","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A new technique was used to measure dynamic internal compliance of a blood vessel or vascular graft subjected to dynamic internal pressure. The internal compliance can be broken into three categories: the volumetric compliance (CV), defined as (dV/V)/dP; the longitudinal compliance (CL), defined as (dL/L)/dP; and radial compliance (CR), defined as (dR/R)/dP. It can be shown mathematically that CV = 2 CR + CL. Thus, measuring any two of the three entities will also give the value for the third. A Dynatek DCT1 dynamic compliance tester was used for measuring the compliance of DenaflexTM biologic grafts and fresh bovine internal thoracic arteries, from which the Denaflex grafts were obtained by fixation. Volumetric compliance was obtained with the test sample mounted in a loose loop that allowed the sample to move both radially and longitudinally. By mounting the sample in a straight fashion that limited longitudinal movement, the radial compliance was determined. The longitudinal compliance was then calculated from the above relationship. Test results show that the fresh bovine artery had an average volumetric compliance of 26.1%/100 mmHg, radial compliance of 9.5%/100 mmHg, and longitudinal compliance of 7.2%/100 mmHg. The Denaflex vascular graft showed a reduction in longitudinal and radial compliance, compared to the fresh raw artery, as a result of extensive fixation.</p>","PeriodicalId":77493,"journal":{"name":"ASAIO transactions","volume":"37 3","pages":"M470-2"},"PeriodicalIF":0.0000,"publicationDate":"1991-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Dynamic internal compliance of a vascular prosthesis.\",\"authors\":\"R Tu,&nbsp;J McIntyre,&nbsp;C Hata,&nbsp;C L Lu,&nbsp;E Wang,&nbsp;R C Quijano\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A new technique was used to measure dynamic internal compliance of a blood vessel or vascular graft subjected to dynamic internal pressure. The internal compliance can be broken into three categories: the volumetric compliance (CV), defined as (dV/V)/dP; the longitudinal compliance (CL), defined as (dL/L)/dP; and radial compliance (CR), defined as (dR/R)/dP. It can be shown mathematically that CV = 2 CR + CL. Thus, measuring any two of the three entities will also give the value for the third. A Dynatek DCT1 dynamic compliance tester was used for measuring the compliance of DenaflexTM biologic grafts and fresh bovine internal thoracic arteries, from which the Denaflex grafts were obtained by fixation. Volumetric compliance was obtained with the test sample mounted in a loose loop that allowed the sample to move both radially and longitudinally. By mounting the sample in a straight fashion that limited longitudinal movement, the radial compliance was determined. The longitudinal compliance was then calculated from the above relationship. Test results show that the fresh bovine artery had an average volumetric compliance of 26.1%/100 mmHg, radial compliance of 9.5%/100 mmHg, and longitudinal compliance of 7.2%/100 mmHg. The Denaflex vascular graft showed a reduction in longitudinal and radial compliance, compared to the fresh raw artery, as a result of extensive fixation.</p>\",\"PeriodicalId\":77493,\"journal\":{\"name\":\"ASAIO transactions\",\"volume\":\"37 3\",\"pages\":\"M470-2\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1991-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ASAIO transactions\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ASAIO transactions","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本文提出了一种测量血管或血管移植物在动态内压作用下动态内顺应性的新方法。内部顺应性可分为三类:体积顺应性(CV),定义为(dV/V)/dP;纵向顺应度(CL),定义为(dL/L)/dP;径向柔度(CR),定义为(dR/R)/dP。CV = 2cr + CL可以用数学方法表示。因此,测量三个实体中的任意两个也会得到第三个实体的值。采用Dynatek DCT1型动态顺应性测试仪测量DenaflexTM生物移植物与新鲜牛胸内动脉的顺应性,固定获得Denaflex移植物。体积顺应性与测试样品安装在一个松散的环路,允许样品径向和纵向移动。通过以限制纵向运动的直线方式安装样品,确定了径向顺应性。然后根据上述关系计算纵向柔度。实验结果表明,牛新鲜动脉的平均体积顺应性为26.1%/100 mmHg,径向顺应性为9.5%/100 mmHg,纵向顺应性为7.2%/100 mmHg。与新鲜的生动脉相比,由于广泛的固定,Denaflex血管移植物显示出纵向和径向顺应性的降低。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Dynamic internal compliance of a vascular prosthesis.

A new technique was used to measure dynamic internal compliance of a blood vessel or vascular graft subjected to dynamic internal pressure. The internal compliance can be broken into three categories: the volumetric compliance (CV), defined as (dV/V)/dP; the longitudinal compliance (CL), defined as (dL/L)/dP; and radial compliance (CR), defined as (dR/R)/dP. It can be shown mathematically that CV = 2 CR + CL. Thus, measuring any two of the three entities will also give the value for the third. A Dynatek DCT1 dynamic compliance tester was used for measuring the compliance of DenaflexTM biologic grafts and fresh bovine internal thoracic arteries, from which the Denaflex grafts were obtained by fixation. Volumetric compliance was obtained with the test sample mounted in a loose loop that allowed the sample to move both radially and longitudinally. By mounting the sample in a straight fashion that limited longitudinal movement, the radial compliance was determined. The longitudinal compliance was then calculated from the above relationship. Test results show that the fresh bovine artery had an average volumetric compliance of 26.1%/100 mmHg, radial compliance of 9.5%/100 mmHg, and longitudinal compliance of 7.2%/100 mmHg. The Denaflex vascular graft showed a reduction in longitudinal and radial compliance, compared to the fresh raw artery, as a result of extensive fixation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信