M Katayama, K Miyasaka, I Hirose, M Shimada, R Mimura, H Akasu
{"title":"微型膜式氧合器在新生儿ECMO中的应用。","authors":"M Katayama, K Miyasaka, I Hirose, M Shimada, R Mimura, H Akasu","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A new miniature membrane oxygenator (Kuraray KMO, size, 0.3 m2, with a priming volume of 47 ml, compliance of less than 0.1 ml/100 mmHg, and pressure loss of 45 mmHg) with improved gas transfer and mechanical durability was developed and tested. The membrane material is a hollow fiber double layer polyolefin. The testing procedures determined by the AAMI were followed, and the results showed improved O2 and CO2 transfer (70 ml/min and 55 ml/min, respectively). Hemolysis was within acceptable limits, and plasma leakage was undetectable after 7 days of perfusion. Clinical study demonstrated satisfactory performance.</p>","PeriodicalId":77493,"journal":{"name":"ASAIO transactions","volume":"37 3","pages":"M407-8"},"PeriodicalIF":0.0000,"publicationDate":"1991-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Miniature membrane oxygenator for use in neonatal ECMO.\",\"authors\":\"M Katayama, K Miyasaka, I Hirose, M Shimada, R Mimura, H Akasu\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A new miniature membrane oxygenator (Kuraray KMO, size, 0.3 m2, with a priming volume of 47 ml, compliance of less than 0.1 ml/100 mmHg, and pressure loss of 45 mmHg) with improved gas transfer and mechanical durability was developed and tested. The membrane material is a hollow fiber double layer polyolefin. The testing procedures determined by the AAMI were followed, and the results showed improved O2 and CO2 transfer (70 ml/min and 55 ml/min, respectively). Hemolysis was within acceptable limits, and plasma leakage was undetectable after 7 days of perfusion. Clinical study demonstrated satisfactory performance.</p>\",\"PeriodicalId\":77493,\"journal\":{\"name\":\"ASAIO transactions\",\"volume\":\"37 3\",\"pages\":\"M407-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1991-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ASAIO transactions\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ASAIO transactions","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Miniature membrane oxygenator for use in neonatal ECMO.
A new miniature membrane oxygenator (Kuraray KMO, size, 0.3 m2, with a priming volume of 47 ml, compliance of less than 0.1 ml/100 mmHg, and pressure loss of 45 mmHg) with improved gas transfer and mechanical durability was developed and tested. The membrane material is a hollow fiber double layer polyolefin. The testing procedures determined by the AAMI were followed, and the results showed improved O2 and CO2 transfer (70 ml/min and 55 ml/min, respectively). Hemolysis was within acceptable limits, and plasma leakage was undetectable after 7 days of perfusion. Clinical study demonstrated satisfactory performance.