M. Kowara, A. Kołodzińska, K. Ozierański, Diana Paskudzka, Klaudia Wołynkiewicz, Edyta Jakubik, Maciej Kubik, Karolina Rybak, M. Grabowski, G. Opolski
{"title":"导联阻抗、起搏阈值和心室起搏百分比随导联年龄的变化——对随访患者的横断面研究。","authors":"M. Kowara, A. Kołodzińska, K. Ozierański, Diana Paskudzka, Klaudia Wołynkiewicz, Edyta Jakubik, Maciej Kubik, Karolina Rybak, M. Grabowski, G. Opolski","doi":"10.24255/hbj/128011","DOIUrl":null,"url":null,"abstract":"Background Patients with permanent pacemakers should be followed up regularly and parameters recorded during the follow-up visits might have potential for further diagnostic studies. Material and methods A pilot cross-sectional study on 127 consecutive patients who underwent pacemaker follow-up was performed. Results The study revealed that patients with older ventricular leads (more than 10 years) have higher lead impedance in comparison with patients with newer ventricular leads (less than 10 years) – 512 Ω vs 461 Ω (median values), with no differences in pacing thresholds, percentage of ventricular pacing, presence of ventricular high rate episodes or status of absolute pacemaker dependency. Conclusions This study demonstrates that parameters obtained from a regular pacemaker check-up differ according to variables such as lead age, which opens a new field for studies investigating other variables (e.g. disease biomarkers) in correlation with parameters from the pacemaker check-up.","PeriodicalId":247417,"journal":{"name":"Heart Beat Journal","volume":"99 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2021-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Lead impedance, pacing threshold and percentage of ventricular pacing according to the lead age – a cross-sectional study upon patients during follow-up.\",\"authors\":\"M. Kowara, A. Kołodzińska, K. Ozierański, Diana Paskudzka, Klaudia Wołynkiewicz, Edyta Jakubik, Maciej Kubik, Karolina Rybak, M. Grabowski, G. Opolski\",\"doi\":\"10.24255/hbj/128011\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background Patients with permanent pacemakers should be followed up regularly and parameters recorded during the follow-up visits might have potential for further diagnostic studies. Material and methods A pilot cross-sectional study on 127 consecutive patients who underwent pacemaker follow-up was performed. Results The study revealed that patients with older ventricular leads (more than 10 years) have higher lead impedance in comparison with patients with newer ventricular leads (less than 10 years) – 512 Ω vs 461 Ω (median values), with no differences in pacing thresholds, percentage of ventricular pacing, presence of ventricular high rate episodes or status of absolute pacemaker dependency. Conclusions This study demonstrates that parameters obtained from a regular pacemaker check-up differ according to variables such as lead age, which opens a new field for studies investigating other variables (e.g. disease biomarkers) in correlation with parameters from the pacemaker check-up.\",\"PeriodicalId\":247417,\"journal\":{\"name\":\"Heart Beat Journal\",\"volume\":\"99 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Heart Beat Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.24255/hbj/128011\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Heart Beat Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.24255/hbj/128011","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
背景:装有永久性起搏器的患者应定期随访,随访期间记录的参数可能有进一步诊断研究的潜力。材料与方法对127例连续接受起搏器随访的患者进行初步横断面研究。结果研究显示,与新心室导联(小于10年)的患者相比,年长心室导联(超过10年)的患者具有更高的导联阻抗——512 Ω vs 461 Ω(中位数),在起搏阈值、心室起搏百分比、心室高速率发作的存在或绝对起搏器依赖状态方面没有差异。结论:本研究表明,定期起搏器检查获得的参数根据导联年龄等变量而不同,这为研究与起搏器检查参数相关的其他变量(如疾病生物标志物)开辟了一个新的领域。
Lead impedance, pacing threshold and percentage of ventricular pacing according to the lead age – a cross-sectional study upon patients during follow-up.
Background Patients with permanent pacemakers should be followed up regularly and parameters recorded during the follow-up visits might have potential for further diagnostic studies. Material and methods A pilot cross-sectional study on 127 consecutive patients who underwent pacemaker follow-up was performed. Results The study revealed that patients with older ventricular leads (more than 10 years) have higher lead impedance in comparison with patients with newer ventricular leads (less than 10 years) – 512 Ω vs 461 Ω (median values), with no differences in pacing thresholds, percentage of ventricular pacing, presence of ventricular high rate episodes or status of absolute pacemaker dependency. Conclusions This study demonstrates that parameters obtained from a regular pacemaker check-up differ according to variables such as lead age, which opens a new field for studies investigating other variables (e.g. disease biomarkers) in correlation with parameters from the pacemaker check-up.