{"title":"基于袋鼠模式的婴儿培养箱湿度PID分析","authors":"S. Setiawan, D. Andayani, A. Pudji, L. Soetjiatie, Alievya Brillianty Anugrah Kusuma","doi":"10.35882/jeeemi.v4i1.6","DOIUrl":null,"url":null,"abstract":"The main factor that affects the parameters of the workings of humidity in the baby incubator is the sensor used to read the humidity in the room. The purpose of this study is to analyze the humidity sensor in the baby incubator using 2 humidity sensors (DHT11 and DHT22) in the different location. The manufacture of this device used an after-only design, with a comparison device of INCU Analyzer, DHT11, and DHT22 sensors. Based on the measurement, DHT11 produced a value of 46%, while DHT22 produced a value of 55.45% with BPFK standards of 50%-70%. Based on the results of measurements using the INCU Analyzer, the average error value for DHT11 is 16.05%, while DHT22 is 3.47%. Therefore, the results showed that the DHT22 sensor was more accurate to be used in baby incubators because the measurement results were under BPFK standards and produced a low error value. This can be further implemented in a baby incubator making to improve the health and safety of the babies","PeriodicalId":369032,"journal":{"name":"Journal of Electronics, Electromedical Engineering, and Medical Informatics","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-01-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Analysis Of Baby Incubator Humidity Based PID with Kangaroo Mode\",\"authors\":\"S. Setiawan, D. Andayani, A. Pudji, L. Soetjiatie, Alievya Brillianty Anugrah Kusuma\",\"doi\":\"10.35882/jeeemi.v4i1.6\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The main factor that affects the parameters of the workings of humidity in the baby incubator is the sensor used to read the humidity in the room. The purpose of this study is to analyze the humidity sensor in the baby incubator using 2 humidity sensors (DHT11 and DHT22) in the different location. The manufacture of this device used an after-only design, with a comparison device of INCU Analyzer, DHT11, and DHT22 sensors. Based on the measurement, DHT11 produced a value of 46%, while DHT22 produced a value of 55.45% with BPFK standards of 50%-70%. Based on the results of measurements using the INCU Analyzer, the average error value for DHT11 is 16.05%, while DHT22 is 3.47%. Therefore, the results showed that the DHT22 sensor was more accurate to be used in baby incubators because the measurement results were under BPFK standards and produced a low error value. This can be further implemented in a baby incubator making to improve the health and safety of the babies\",\"PeriodicalId\":369032,\"journal\":{\"name\":\"Journal of Electronics, Electromedical Engineering, and Medical Informatics\",\"volume\":\"1 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-01-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Electronics, Electromedical Engineering, and Medical Informatics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.35882/jeeemi.v4i1.6\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Electronics, Electromedical Engineering, and Medical Informatics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.35882/jeeemi.v4i1.6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

影响婴儿恒温箱湿度工作参数的主要因素是用于读取室内湿度的传感器。本研究的目的是利用2个湿度传感器(DHT11和DHT22)在不同位置对婴儿培养箱中的湿度传感器进行分析。本装置的制作采用后置设计,采用INCU分析仪、DHT11、DHT22传感器的比较装置。根据测量,DHT11产生的值为46%,DHT22产生的值为55.45%,BPFK标准为50%-70%。根据INCU分析仪的测量结果,DHT11的平均误差值为16.05%,DHT22的平均误差值为3.47%。因此,结果表明,DHT22传感器测量结果在BPFK标准下,误差值较低,更适合用于婴儿保育箱。这可以在婴儿保温箱制作中进一步实施,以提高婴儿的健康和安全
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Analysis Of Baby Incubator Humidity Based PID with Kangaroo Mode
The main factor that affects the parameters of the workings of humidity in the baby incubator is the sensor used to read the humidity in the room. The purpose of this study is to analyze the humidity sensor in the baby incubator using 2 humidity sensors (DHT11 and DHT22) in the different location. The manufacture of this device used an after-only design, with a comparison device of INCU Analyzer, DHT11, and DHT22 sensors. Based on the measurement, DHT11 produced a value of 46%, while DHT22 produced a value of 55.45% with BPFK standards of 50%-70%. Based on the results of measurements using the INCU Analyzer, the average error value for DHT11 is 16.05%, while DHT22 is 3.47%. Therefore, the results showed that the DHT22 sensor was more accurate to be used in baby incubators because the measurement results were under BPFK standards and produced a low error value. This can be further implemented in a baby incubator making to improve the health and safety of the babies
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信