在新生儿筛查TSH, 17-OHP和IRT的自动时间分辨荧光免疫测定中,商用瓶装纯净水作为替代仪器给水

E. Villanueva, Roel Martin Bumalay, J. M. Rivera
{"title":"在新生儿筛查TSH, 17-OHP和IRT的自动时间分辨荧光免疫测定中,商用瓶装纯净水作为替代仪器给水","authors":"E. Villanueva, Roel Martin Bumalay, J. M. Rivera","doi":"10.21141/PJP.2018.011","DOIUrl":null,"url":null,"abstract":"Objective. The study was undertaken to determine if commercially bottled purified water can be used as substitute instrument feed water for three (3) newborn screening immunoassays. Methdology. A total of 294 control samples and 300 patient samples were included in this study. Accuracy and precision studies using control samples, and parallel testing using patient samples, were done to compare the use of clinical laboratory reagent water (CLRW) and commercially bottled purified water (CBPW) in the performance of automated time-resolved fluorescent immunoassay of thyroid stimulating hormone (TSH), 17α-OH-progesterone (17-OHP) and immunoreactive trypsinogen (IRT). Results. The use of CBPW as instrument feed water for measurements of TSH, 17-OHP and IRT levels by automated time-resolved fluorescent immunoassay using AutoDELFIA (Perkin-Elmer) in NBS has an acceptable accuracy and precision compared to using CLRW. The parallel testing using patient samples showed that, overall, the performance of using CBPW in automated time-resolved fluorescent immunoassay for TSH, 17-OHP, and IRT is acceptable, compared with using CLRW as instrument feed water. Conclusion. Commercially bottled purified water can be used as substitute when setting up a laboratory water purification system is too expensive for a laboratory, or as back up to clinical laboratory reagent water when there is breakdown of the installed water purification system to be used as instrument feed water in automated time-resolved fluorescent immunoassay of TSH, 17-OHP and IRT in NBS using AutoDELFIA (Perkin-Elmer).","PeriodicalId":166708,"journal":{"name":"Philippine Journal of Pathology","volume":"27 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2018-11-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Commercially Bottled Purified Water as an Alternative Instrument Feed Water in Automated Time-Resolved Fluorescent Immunoassay for TSH, 17-OHP and IRT in Neonatal Screening\",\"authors\":\"E. Villanueva, Roel Martin Bumalay, J. M. Rivera\",\"doi\":\"10.21141/PJP.2018.011\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective. The study was undertaken to determine if commercially bottled purified water can be used as substitute instrument feed water for three (3) newborn screening immunoassays. Methdology. A total of 294 control samples and 300 patient samples were included in this study. Accuracy and precision studies using control samples, and parallel testing using patient samples, were done to compare the use of clinical laboratory reagent water (CLRW) and commercially bottled purified water (CBPW) in the performance of automated time-resolved fluorescent immunoassay of thyroid stimulating hormone (TSH), 17α-OH-progesterone (17-OHP) and immunoreactive trypsinogen (IRT). Results. The use of CBPW as instrument feed water for measurements of TSH, 17-OHP and IRT levels by automated time-resolved fluorescent immunoassay using AutoDELFIA (Perkin-Elmer) in NBS has an acceptable accuracy and precision compared to using CLRW. The parallel testing using patient samples showed that, overall, the performance of using CBPW in automated time-resolved fluorescent immunoassay for TSH, 17-OHP, and IRT is acceptable, compared with using CLRW as instrument feed water. Conclusion. Commercially bottled purified water can be used as substitute when setting up a laboratory water purification system is too expensive for a laboratory, or as back up to clinical laboratory reagent water when there is breakdown of the installed water purification system to be used as instrument feed water in automated time-resolved fluorescent immunoassay of TSH, 17-OHP and IRT in NBS using AutoDELFIA (Perkin-Elmer).\",\"PeriodicalId\":166708,\"journal\":{\"name\":\"Philippine Journal of Pathology\",\"volume\":\"27 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2018-11-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Philippine Journal of Pathology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.21141/PJP.2018.011\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Philippine Journal of Pathology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.21141/PJP.2018.011","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目标。进行这项研究是为了确定商业瓶装纯净水是否可以用作三(3)个新生儿筛查免疫测定的替代仪器喂养水。Methdology。本研究共纳入294份对照样本和300份患者样本。使用对照样品进行准确性和精密度研究,并使用患者样品进行平行测试,比较使用临床实验室试剂水(CLRW)和市售瓶装纯净水(CBPW)在促甲状腺激素(TSH)、17α- oh孕酮(17-OHP)和免疫反应性胰蛋白酶原(IRT)的自动时间分辨荧光免疫测定中的性能。结果。与使用CLRW相比,使用CBPW作为仪器给水,使用AutoDELFIA (Perkin-Elmer)在NBS进行自动时间分辨荧光免疫测定,测量TSH, 17-OHP和IRT水平,具有可接受的准确性和精密度。使用患者样本的平行测试表明,总体而言,与使用CLRW作为仪器进水相比,使用CBPW进行TSH、17-OHP和IRT的自动时间分辨荧光免疫分析的性能是可以接受的。结论。市售瓶装纯净水可作为实验室水净化系统过于昂贵时的替代品,或在安装的水净化系统发生故障时作为临床实验室试剂水的备用,使用AutoDELFIA (Perkin-Elmer)作为NBS中TSH, 17-OHP和IRT的自动时间分辨荧光免疫测定的仪器给水。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Commercially Bottled Purified Water as an Alternative Instrument Feed Water in Automated Time-Resolved Fluorescent Immunoassay for TSH, 17-OHP and IRT in Neonatal Screening
Objective. The study was undertaken to determine if commercially bottled purified water can be used as substitute instrument feed water for three (3) newborn screening immunoassays. Methdology. A total of 294 control samples and 300 patient samples were included in this study. Accuracy and precision studies using control samples, and parallel testing using patient samples, were done to compare the use of clinical laboratory reagent water (CLRW) and commercially bottled purified water (CBPW) in the performance of automated time-resolved fluorescent immunoassay of thyroid stimulating hormone (TSH), 17α-OH-progesterone (17-OHP) and immunoreactive trypsinogen (IRT). Results. The use of CBPW as instrument feed water for measurements of TSH, 17-OHP and IRT levels by automated time-resolved fluorescent immunoassay using AutoDELFIA (Perkin-Elmer) in NBS has an acceptable accuracy and precision compared to using CLRW. The parallel testing using patient samples showed that, overall, the performance of using CBPW in automated time-resolved fluorescent immunoassay for TSH, 17-OHP, and IRT is acceptable, compared with using CLRW as instrument feed water. Conclusion. Commercially bottled purified water can be used as substitute when setting up a laboratory water purification system is too expensive for a laboratory, or as back up to clinical laboratory reagent water when there is breakdown of the installed water purification system to be used as instrument feed water in automated time-resolved fluorescent immunoassay of TSH, 17-OHP and IRT in NBS using AutoDELFIA (Perkin-Elmer).
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信