{"title":"Nilai Contrast to Noise Ratio (CNR) Radiograf Thorax PA antara menggunakan Grid dengan tanpa Menggunakan Grid","authors":"Angga Yosainto Bequet, Luthfi Rusyadi, F. Fatimah","doi":"10.31983/jimed.v6i2.5653","DOIUrl":null,"url":null,"abstract":"Background: Chest radiographic examination often does not use a grid even though the thorax is an object that has a thickness of more than 10 cm so it must use a grid. This study aims to determine the differences and the degree of change in Contrast to Noise Ratio on PA radiographs between using a grid and without using a grid.Methods: This type of research is a pre-experimental quantitative. The research subject is a radiological image of PA thorax made using the grid (50 patients) and without using the grid (patient). The image is analyzed by pixel value using Dicom softwareResults: There is a difference in contrast values on the PA radiographs between those using the Grid and without using the Grid (p-value 0.001). The average value of the contrast on the chest radiograph using a Grid is 2283.60, while the one without using a Grid is 1878.58. Noise value also shows that there is a difference between chest radiographs of PA images using Grid without using Grid (p-value = 0.001). The average thorax photo noise using Grid is 25.32, paired using Grid is 17.84. Statistical test on the value of Contrast to Noise Ratio (CNR) shows that there is a difference between radiographs of PA chest radiography using Grid without Grid (p-value 0.001). The average CNR radiograph of the PA thorax PA using Grid is 100.79, while the CNR value of the radiograph of thorax PA without using Grid is 125.62.Conclusions: There is a Contrast difference between using a grid and without using a grid (p-value 0.001). There is a Noise difference in PA radiographs between using a grid and without using a grid (p-value = 0.001). There is a difference in Contrast to Noise Ratio on the PA chest radiograph between using a grid and without using a grid (p-value 0.001).","PeriodicalId":205622,"journal":{"name":"Jurnal Imejing Diagnostik (JImeD)","volume":"11 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jurnal Imejing Diagnostik (JImeD)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31983/jimed.v6i2.5653","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

背景:胸片检查通常不使用网格,即使胸部是一个厚度超过10厘米的物体,也必须使用网格。本研究旨在确定使用网格和不使用网格在PA x线片上的对比度噪声比的差异和变化程度。方法:本研究采用实验前定量法。研究对象是使用网格(50例患者)和不使用网格(患者)的PA胸放射图像。使用Dicom软件对图像进行像素值分析。结果:使用Grid和不使用Grid的PA x线片的对比度值存在差异(p值0.001)。使用Grid的胸片对比度平均值为2283.60,未使用Grid的胸片对比度平均值为1878.58。噪声值也显示使用Grid的胸片与不使用Grid的胸片之间存在差异(p值= 0.001)。使用Grid的平均胸腔光噪声为25.32,使用Grid的配对噪声为17.84。对噪声比(CNR)值的统计检验表明,不使用网格的PA胸片与使用网格的胸片之间存在差异(p值0.001)。使用Grid的PA胸片的平均CNR值为100.79,而未使用Grid的PA胸片的CNR值为125.62。结论:使用网格和不使用网格之间存在对比差异(p值0.001)。在使用网格和不使用网格的PA x线片中存在噪声差异(p值= 0.001)。使用网格和不使用网格在PA胸片上的对比度噪声比有差异(p值0.001)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Nilai Contrast to Noise Ratio (CNR) Radiograf Thorax PA antara menggunakan Grid dengan tanpa Menggunakan Grid
Background: Chest radiographic examination often does not use a grid even though the thorax is an object that has a thickness of more than 10 cm so it must use a grid. This study aims to determine the differences and the degree of change in Contrast to Noise Ratio on PA radiographs between using a grid and without using a grid.Methods: This type of research is a pre-experimental quantitative. The research subject is a radiological image of PA thorax made using the grid (50 patients) and without using the grid (patient). The image is analyzed by pixel value using Dicom softwareResults: There is a difference in contrast values on the PA radiographs between those using the Grid and without using the Grid (p-value 0.001). The average value of the contrast on the chest radiograph using a Grid is 2283.60, while the one without using a Grid is 1878.58. Noise value also shows that there is a difference between chest radiographs of PA images using Grid without using Grid (p-value = 0.001). The average thorax photo noise using Grid is 25.32, paired using Grid is 17.84. Statistical test on the value of Contrast to Noise Ratio (CNR) shows that there is a difference between radiographs of PA chest radiography using Grid without Grid (p-value 0.001). The average CNR radiograph of the PA thorax PA using Grid is 100.79, while the CNR value of the radiograph of thorax PA without using Grid is 125.62.Conclusions: There is a Contrast difference between using a grid and without using a grid (p-value 0.001). There is a Noise difference in PA radiographs between using a grid and without using a grid (p-value = 0.001). There is a difference in Contrast to Noise Ratio on the PA chest radiograph between using a grid and without using a grid (p-value 0.001).
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信