{"title":"用于溶液超声分析的电子测量系统的校准","authors":"A. Barwicz, R. Morawski, J. Dion","doi":"10.1109/IMTC.1990.65973","DOIUrl":null,"url":null,"abstract":"The problem of calibrating a system for measuring the concentration of methanol in water is investigated. The inefficiency of polynomial and inverse-polynomial models used for this purpose is demonstrated. The cubic-spline interpolation is chosen and examined as a flexible means of calibration. A method for experimental validation of the results of calibration is proposed. The proposed method makes possible the utilization of a much greater amount of calibration data for improving the accuracy of calibration as compared with the polynomial and inverse-polynomial models, and requires much less a priori information on the relationships between quantities involved in the measurement process.<<ETX>>","PeriodicalId":404761,"journal":{"name":"7th IEEE Conference on Instrumentation and Measurement Technology","volume":"45 3-4","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1990-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"13","resultStr":"{\"title\":\"Calibration of an electronic measuring system for ultrasonic analysis of solutions\",\"authors\":\"A. Barwicz, R. Morawski, J. Dion\",\"doi\":\"10.1109/IMTC.1990.65973\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The problem of calibrating a system for measuring the concentration of methanol in water is investigated. The inefficiency of polynomial and inverse-polynomial models used for this purpose is demonstrated. The cubic-spline interpolation is chosen and examined as a flexible means of calibration. A method for experimental validation of the results of calibration is proposed. The proposed method makes possible the utilization of a much greater amount of calibration data for improving the accuracy of calibration as compared with the polynomial and inverse-polynomial models, and requires much less a priori information on the relationships between quantities involved in the measurement process.<<ETX>>\",\"PeriodicalId\":404761,\"journal\":{\"name\":\"7th IEEE Conference on Instrumentation and Measurement Technology\",\"volume\":\"45 3-4\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1990-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"13\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"7th IEEE Conference on Instrumentation and Measurement Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/IMTC.1990.65973\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"7th IEEE Conference on Instrumentation and Measurement Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IMTC.1990.65973","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Calibration of an electronic measuring system for ultrasonic analysis of solutions
The problem of calibrating a system for measuring the concentration of methanol in water is investigated. The inefficiency of polynomial and inverse-polynomial models used for this purpose is demonstrated. The cubic-spline interpolation is chosen and examined as a flexible means of calibration. A method for experimental validation of the results of calibration is proposed. The proposed method makes possible the utilization of a much greater amount of calibration data for improving the accuracy of calibration as compared with the polynomial and inverse-polynomial models, and requires much less a priori information on the relationships between quantities involved in the measurement process.<>