使用1个多探测器计算机断层扫描和5个锥束计算机断层扫描方案的体积测量验证和比较:一项体外研究。

IF 1.7 Q3 DENTISTRY, ORAL SURGERY & MEDICINE
Juliana Andréa Corrêa Travessas, Alessandra Mendonça Dos Santos, Rodrigo Pagliarini Buligon, Nádia Assein Arús, Priscila Fernanda Tiecher da Silveira, Heraldo Luis Dias da Silveira, Mariana Boessio Vizzotto
{"title":"使用1个多探测器计算机断层扫描和5个锥束计算机断层扫描方案的体积测量验证和比较:一项体外研究。","authors":"Juliana Andréa Corrêa Travessas,&nbsp;Alessandra Mendonça Dos Santos,&nbsp;Rodrigo Pagliarini Buligon,&nbsp;Nádia Assein Arús,&nbsp;Priscila Fernanda Tiecher da Silveira,&nbsp;Heraldo Luis Dias da Silveira,&nbsp;Mariana Boessio Vizzotto","doi":"10.5624/isd.20220102","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The purpose of this study was to compare volume measurements obtained using 2 image software packages on Digital Imaging and Communications in Medicine (DICOM) images acquired from 1 multidetector computed tomography and 5 cone-beam computed tomography devices, using different protocols for physical volume measurements.</p><p><strong>Materials and methods: </strong>Four pieces of bovine leg were prepared. Marrow was removed from 3 pieces, leaving cortical bone exposed. The resulting space of 1 piece was filled with water, another was filled with propylene glycol, and the third was left unfilled. The marrow in the fourth sample was left fully intact. Volume measurements were obtained after importing DICOM images into the Dolphin Imaging 11.95 and ITK-SNAP software programs. Data were analyzed using 3-way analysis of variance with a generalized linear model to determine the effects of voxel size, software, and content on percentage mean volume differences between tomographic protocols. A significance level of 0.05 was used.</p><p><strong>Results: </strong>The intraclass correlation coefficients for intraobserver and interobserver reliability were, respectively, 0.915 and 0.764 for the Dolphin software and 0.894 and 0.766 for the ITK-SNAP software. Three sources of statistically significant variation were identified: the interaction between software and content (<i>P</i>=0.001), the main effect of content (<i>P</i>=0.014), and the main effect of software (<i>P</i>=0.001). Voxel size was not associated with statistically significant differences in volume measurements.</p><p><strong>Conclusion: </strong>Both content and software influenced the accuracy of volume measurements, especially when the content had gray values similar to those of the adjacent tissues.</p>","PeriodicalId":51714,"journal":{"name":"Imaging Science in Dentistry","volume":"52 4","pages":"399-408"},"PeriodicalIF":1.7000,"publicationDate":"2022-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/1b/37/isd-52-399.PMC9807801.pdf","citationCount":"0","resultStr":"{\"title\":\"Validation and comparison of volume measurements using 1 multidetector computed tomography and 5 cone-beam computed tomography protocols: An in vitro study.\",\"authors\":\"Juliana Andréa Corrêa Travessas,&nbsp;Alessandra Mendonça Dos Santos,&nbsp;Rodrigo Pagliarini Buligon,&nbsp;Nádia Assein Arús,&nbsp;Priscila Fernanda Tiecher da Silveira,&nbsp;Heraldo Luis Dias da Silveira,&nbsp;Mariana Boessio Vizzotto\",\"doi\":\"10.5624/isd.20220102\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>The purpose of this study was to compare volume measurements obtained using 2 image software packages on Digital Imaging and Communications in Medicine (DICOM) images acquired from 1 multidetector computed tomography and 5 cone-beam computed tomography devices, using different protocols for physical volume measurements.</p><p><strong>Materials and methods: </strong>Four pieces of bovine leg were prepared. Marrow was removed from 3 pieces, leaving cortical bone exposed. The resulting space of 1 piece was filled with water, another was filled with propylene glycol, and the third was left unfilled. The marrow in the fourth sample was left fully intact. Volume measurements were obtained after importing DICOM images into the Dolphin Imaging 11.95 and ITK-SNAP software programs. Data were analyzed using 3-way analysis of variance with a generalized linear model to determine the effects of voxel size, software, and content on percentage mean volume differences between tomographic protocols. A significance level of 0.05 was used.</p><p><strong>Results: </strong>The intraclass correlation coefficients for intraobserver and interobserver reliability were, respectively, 0.915 and 0.764 for the Dolphin software and 0.894 and 0.766 for the ITK-SNAP software. Three sources of statistically significant variation were identified: the interaction between software and content (<i>P</i>=0.001), the main effect of content (<i>P</i>=0.014), and the main effect of software (<i>P</i>=0.001). Voxel size was not associated with statistically significant differences in volume measurements.</p><p><strong>Conclusion: </strong>Both content and software influenced the accuracy of volume measurements, especially when the content had gray values similar to those of the adjacent tissues.</p>\",\"PeriodicalId\":51714,\"journal\":{\"name\":\"Imaging Science in Dentistry\",\"volume\":\"52 4\",\"pages\":\"399-408\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2022-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/1b/37/isd-52-399.PMC9807801.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Imaging Science in Dentistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.5624/isd.20220102\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Imaging Science in Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5624/isd.20220102","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究的目的是比较使用2个图像软件包获得的医学数字成像和通信(DICOM)图像,这些图像来自1个多探测器计算机断层扫描和5个锥束计算机断层扫描设备,使用不同的物理体积测量方案。材料与方法:制备4块牛腿。从3块骨头中取出骨髓,露出皮质骨。1片填充水,另一片填充丙二醇,第三片不填充。第四个样本的骨髓完全完好无损。将DICOM图像导入Dolphin Imaging 11.95和ITK-SNAP软件程序后获得体积测量。采用广义线性模型对数据进行三向方差分析,以确定体素大小、软件和内容对层析成像方案之间的百分比平均体积差异的影响。采用0.05的显著性水平。结果:Dolphin软件的观察者内信度和观察者间信度的类内相关系数分别为0.915和0.764,ITK-SNAP软件的类内相关系数分别为0.894和0.766。确定了三个统计显著差异的来源:软件和内容之间的相互作用(P=0.001),内容的主要影响(P=0.014)和软件的主要影响(P=0.001)。体素大小与体积测量的统计学显著差异无关。结论:内容物和软件都会影响体积测量的准确性,特别是当内容物的灰度值与邻近组织的灰度值相似时。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Validation and comparison of volume measurements using 1 multidetector computed tomography and 5 cone-beam computed tomography protocols: An in vitro study.

Validation and comparison of volume measurements using 1 multidetector computed tomography and 5 cone-beam computed tomography protocols: An in vitro study.

Validation and comparison of volume measurements using 1 multidetector computed tomography and 5 cone-beam computed tomography protocols: An in vitro study.

Validation and comparison of volume measurements using 1 multidetector computed tomography and 5 cone-beam computed tomography protocols: An in vitro study.

Purpose: The purpose of this study was to compare volume measurements obtained using 2 image software packages on Digital Imaging and Communications in Medicine (DICOM) images acquired from 1 multidetector computed tomography and 5 cone-beam computed tomography devices, using different protocols for physical volume measurements.

Materials and methods: Four pieces of bovine leg were prepared. Marrow was removed from 3 pieces, leaving cortical bone exposed. The resulting space of 1 piece was filled with water, another was filled with propylene glycol, and the third was left unfilled. The marrow in the fourth sample was left fully intact. Volume measurements were obtained after importing DICOM images into the Dolphin Imaging 11.95 and ITK-SNAP software programs. Data were analyzed using 3-way analysis of variance with a generalized linear model to determine the effects of voxel size, software, and content on percentage mean volume differences between tomographic protocols. A significance level of 0.05 was used.

Results: The intraclass correlation coefficients for intraobserver and interobserver reliability were, respectively, 0.915 and 0.764 for the Dolphin software and 0.894 and 0.766 for the ITK-SNAP software. Three sources of statistically significant variation were identified: the interaction between software and content (P=0.001), the main effect of content (P=0.014), and the main effect of software (P=0.001). Voxel size was not associated with statistically significant differences in volume measurements.

Conclusion: Both content and software influenced the accuracy of volume measurements, especially when the content had gray values similar to those of the adjacent tissues.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Imaging Science in Dentistry
Imaging Science in Dentistry DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
2.90
自引率
11.10%
发文量
42
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信