{"title":"测量毛细血管血糖浓度:第一滴血真的是正确的一滴血吗?","authors":"Wibke Johannis , Annika Meyer , Dzemal Elezagic, Thomas Streichert","doi":"10.1016/j.pcd.2023.08.005","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Capillary blood glucose (CBG) is fundamental for diabetes mellitus management. However, it is still unclear whether the first or the subsequent blood drops most accurately reflect patients’ blood glucose levels.</p></div><div><h3>Methods</h3><p><span>128 healthy volunteers were included in this prospective cohort study from November 2021 to December 2021. Capillary blood glucose concentrations of the first four blood drops, measured using glucose meters were compared with </span>venous blood<span> concentrations of the respective donors, measured using an in-lab hexokinase<span> method. ANOVA, the Spearman correlation coefficient and Bland-Altman plots were used to analyze the data.</span></span></p></div><div><h3>Results</h3><p>The mean plasma glucose concentration was 90.46 mg/dl with an SD of ± 14.416 (5.02 ± 0.8 mmol/l). There were statistically strong correlations among the glucose concentrations of all capillary blood drops (correlation coefficients of r > 0.8). The first capillary blood drops showed the lowest mean difference to plasma blood glucose concentrations (+4.92 mg/dl; +0.27 mmol/l), followed by the third drop (+7.28 mg/dl; +0.4 mmol/l), second drop (+7.30 mg/dl; +0.4 mmol/l) and fourth drop (+8.87 mg/dl; +0.49 mmol/l).</p></div><div><h3>Conclusion</h3><p>There is a strong correlation and good agreement between the different capillary blood drops, making all blood drops equally suitable for blood glucose monitoring. In practice, the given setting (clinical or patient self-monitoring) should be considered upon choosing a specific blood drop.</p></div>","PeriodicalId":48997,"journal":{"name":"Primary Care Diabetes","volume":null,"pages":null},"PeriodicalIF":2.6000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Measuring capillary blood glucose concentration: Is the first blood drop really the right blood drop?\",\"authors\":\"Wibke Johannis , Annika Meyer , Dzemal Elezagic, Thomas Streichert\",\"doi\":\"10.1016/j.pcd.2023.08.005\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>Capillary blood glucose (CBG) is fundamental for diabetes mellitus management. However, it is still unclear whether the first or the subsequent blood drops most accurately reflect patients’ blood glucose levels.</p></div><div><h3>Methods</h3><p><span>128 healthy volunteers were included in this prospective cohort study from November 2021 to December 2021. Capillary blood glucose concentrations of the first four blood drops, measured using glucose meters were compared with </span>venous blood<span> concentrations of the respective donors, measured using an in-lab hexokinase<span> method. ANOVA, the Spearman correlation coefficient and Bland-Altman plots were used to analyze the data.</span></span></p></div><div><h3>Results</h3><p>The mean plasma glucose concentration was 90.46 mg/dl with an SD of ± 14.416 (5.02 ± 0.8 mmol/l). There were statistically strong correlations among the glucose concentrations of all capillary blood drops (correlation coefficients of r > 0.8). The first capillary blood drops showed the lowest mean difference to plasma blood glucose concentrations (+4.92 mg/dl; +0.27 mmol/l), followed by the third drop (+7.28 mg/dl; +0.4 mmol/l), second drop (+7.30 mg/dl; +0.4 mmol/l) and fourth drop (+8.87 mg/dl; +0.49 mmol/l).</p></div><div><h3>Conclusion</h3><p>There is a strong correlation and good agreement between the different capillary blood drops, making all blood drops equally suitable for blood glucose monitoring. In practice, the given setting (clinical or patient self-monitoring) should be considered upon choosing a specific blood drop.</p></div>\",\"PeriodicalId\":48997,\"journal\":{\"name\":\"Primary Care Diabetes\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2023-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Primary Care Diabetes\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1751991823001432\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENDOCRINOLOGY & METABOLISM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Primary Care Diabetes","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1751991823001432","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
Measuring capillary blood glucose concentration: Is the first blood drop really the right blood drop?
Background
Capillary blood glucose (CBG) is fundamental for diabetes mellitus management. However, it is still unclear whether the first or the subsequent blood drops most accurately reflect patients’ blood glucose levels.
Methods
128 healthy volunteers were included in this prospective cohort study from November 2021 to December 2021. Capillary blood glucose concentrations of the first four blood drops, measured using glucose meters were compared with venous blood concentrations of the respective donors, measured using an in-lab hexokinase method. ANOVA, the Spearman correlation coefficient and Bland-Altman plots were used to analyze the data.
Results
The mean plasma glucose concentration was 90.46 mg/dl with an SD of ± 14.416 (5.02 ± 0.8 mmol/l). There were statistically strong correlations among the glucose concentrations of all capillary blood drops (correlation coefficients of r > 0.8). The first capillary blood drops showed the lowest mean difference to plasma blood glucose concentrations (+4.92 mg/dl; +0.27 mmol/l), followed by the third drop (+7.28 mg/dl; +0.4 mmol/l), second drop (+7.30 mg/dl; +0.4 mmol/l) and fourth drop (+8.87 mg/dl; +0.49 mmol/l).
Conclusion
There is a strong correlation and good agreement between the different capillary blood drops, making all blood drops equally suitable for blood glucose monitoring. In practice, the given setting (clinical or patient self-monitoring) should be considered upon choosing a specific blood drop.
期刊介绍:
The journal publishes original research articles and high quality reviews in the fields of clinical care, diabetes education, nutrition, health services, psychosocial research and epidemiology and other areas as far as is relevant for diabetology in a primary-care setting. The purpose of the journal is to encourage interdisciplinary research and discussion between all those who are involved in primary diabetes care on an international level. The Journal also publishes news and articles concerning the policies and activities of Primary Care Diabetes Europe and reflects the society''s aim of improving the care for people with diabetes mellitus within the primary-care setting.