Dose uncertainty due to needle-tip localization error in prostate seed implantation.

IF 1.1 4区 医学 Q4 ONCOLOGY
Zhengzheng Xu, Theodore H Arsenault, Bryan Traughber, Roger Ove, Tarun K Podder
{"title":"Dose uncertainty due to needle-tip localization error in prostate seed implantation.","authors":"Zhengzheng Xu,&nbsp;Theodore H Arsenault,&nbsp;Bryan Traughber,&nbsp;Roger Ove,&nbsp;Tarun K Podder","doi":"10.5114/jcb.2022.123978","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study quantified the dosimetric uncertainty caused by needle-tip detection errors in ultrasound images due to bevel-tip orientation differences, with respect to the location on template grid.</p><p><strong>Material and methods: </strong>Trans-rectal ultrasound (TRUS) system with physical template grid and 18-gauge bevel-tip brachytherapy needles were used. TRUS was set at 6.5 MHz in water phantom, and measurements were taken with 50% and 100% B-mode TRUS gains. Needle-tip localization errors were then retrospectively applied back to 45 prostate seed implant plans to evaluate the important planning parameters for the prostate (D<sub>90</sub>, V<sub>100</sub>, V<sub>150</sub>, and V<sub>200</sub>), urethra (D<sub>10</sub> and D<sub>30</sub>), and rectum (V<sub>100</sub>, D<sub>2cc</sub>, and D<sub>0.1cc</sub>), following the ABS and AAPM TG-137 guidelines.</p><p><strong>Results: </strong>The needle-tip detection errors for 50% and 100% TRUS gains were 3.7 mm (max) and 5.2 mm (max), respectively. The observed significant decrease in prostate coverage (mean D<sub>90</sub> lower by 12.8%, and V<sub>100</sub> lower by 3.9% for smaller prostates) after seed placements were corrected by compensating the needle-tip detection errors. Apex of the prostate was hotter, and the base was cooler. Dosimetric difference for urethral and rectal parameters were not statistically significant.</p><p><strong>Conclusions: </strong>This study revealed that the beveled needle-tip orientation could considerably impact the needle tips detection accuracy, based on which the seeds might be delivered. These errors can lead to significant dosimetric uncertainty in prostate seed implantation.</p>","PeriodicalId":51305,"journal":{"name":"Journal of Contemporary Brachytherapy","volume":"14 6","pages":"582-589"},"PeriodicalIF":1.1000,"publicationDate":"2022-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/bf/26/JCB-14-49848.PMC9924154.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Contemporary Brachytherapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5114/jcb.2022.123978","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study quantified the dosimetric uncertainty caused by needle-tip detection errors in ultrasound images due to bevel-tip orientation differences, with respect to the location on template grid.

Material and methods: Trans-rectal ultrasound (TRUS) system with physical template grid and 18-gauge bevel-tip brachytherapy needles were used. TRUS was set at 6.5 MHz in water phantom, and measurements were taken with 50% and 100% B-mode TRUS gains. Needle-tip localization errors were then retrospectively applied back to 45 prostate seed implant plans to evaluate the important planning parameters for the prostate (D90, V100, V150, and V200), urethra (D10 and D30), and rectum (V100, D2cc, and D0.1cc), following the ABS and AAPM TG-137 guidelines.

Results: The needle-tip detection errors for 50% and 100% TRUS gains were 3.7 mm (max) and 5.2 mm (max), respectively. The observed significant decrease in prostate coverage (mean D90 lower by 12.8%, and V100 lower by 3.9% for smaller prostates) after seed placements were corrected by compensating the needle-tip detection errors. Apex of the prostate was hotter, and the base was cooler. Dosimetric difference for urethral and rectal parameters were not statistically significant.

Conclusions: This study revealed that the beveled needle-tip orientation could considerably impact the needle tips detection accuracy, based on which the seeds might be delivered. These errors can lead to significant dosimetric uncertainty in prostate seed implantation.

前列腺粒子植入术中针尖定位误差引起的剂量不确定性。
目的:本研究量化了超声图像中针尖检测误差所引起的剂量学不确定度,该不确定度是由于针尖在模板网格上的位置不同而引起的。材料和方法:经直肠超声(TRUS)系统,物理模板网格和18号斜尖近距离治疗针。TRUS在水模中设置为6.5 MHz,并在50%和100% b模TRUS增益下进行测量。根据ABS和AAPM TG-137指南,将针尖定位误差回顾性应用到45个前列腺粒子植入计划中,评估前列腺(D90、V100、V150和V200)、尿道(D10和D30)和直肠(V100、D2cc和D0.1cc)的重要计划参数。结果:50%和100% TRUS增益的针尖检测误差分别为3.7 mm和5.2 mm。通过补偿针尖检测误差来纠正种子植入后,前列腺覆盖率显著降低(平均D90降低12.8%,较小前列腺V100降低3.9%)。前列腺顶端较热,底部较冷。尿道和直肠参数的剂量学差异无统计学意义。结论:针尖的倾斜方向对针尖检测精度有较大影响,从而影响种子的输送。这些误差会导致前列腺粒子植入剂量学的不确定性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Contemporary Brachytherapy
Journal of Contemporary Brachytherapy ONCOLOGY-RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
CiteScore
2.40
自引率
14.30%
发文量
54
审稿时长
16 weeks
期刊介绍: The “Journal of Contemporary Brachytherapy” is an international and multidisciplinary journal that will publish papers of original research as well as reviews of articles. Main subjects of the journal include: clinical brachytherapy, combined modality treatment, advances in radiobiology, hyperthermia and tumour biology, as well as physical aspects relevant to brachytherapy, particularly in the field of imaging, dosimetry and radiation therapy planning. Original contributions will include experimental studies of combined modality treatment, tumor sensitization and normal tissue protection, molecular radiation biology, and clinical investigations of cancer treatment in brachytherapy. Another field of interest will be the educational part of the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信