Theresa Eckermann, Michael B Hoffmann, Khaldoon O Al-Nosairy
{"title":"Comparison of DTL and gold cup skin electrodes for recordings of the multifocal electroretinogram.","authors":"Theresa Eckermann, Michael B Hoffmann, Khaldoon O Al-Nosairy","doi":"10.1007/s10633-022-09912-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To compare mfERG recordings with the Dawson-Trick-Litzkow (DTL) and gold cup skin electrode in healthy young and old adults and to test the sensitivity of both electrodes to age-related changes in the responses.</p><p><strong>Methods: </strong>Twenty participants aged 20-27 years (\"young\") and 20 participants aged 60-75 (\"old\") with a visual acuity of ≤ 0 logMAR were included. The mfERG responses were recorded simultaneously using DTL and skin electrodes. P1 amplitudes, peak times and signal-to-noise ratios (SNRs) were compared between both electrodes and across age groups, and correlation analyses were performed. The electrode's performance in discriminating between age groups was assessed via area under curve (AUC) of receiver operating characteristics.</p><p><strong>Results: </strong>Both electrodes reflected the typical waveform of mfERG recordings. For the skin electrode, however, P1 amplitudes were significantly reduced (p < 0.001; reduction by over 70%), P1 peak times were significantly shorter (p < 0.001; by approx. 1.5 ms), and SNRs were reduced [(p < 0.001; logSNR ± SEM DTL young (old) vs gold cup: 0.79 ± 0.13 (0.71 ± 0.15) vs 0.37 ± 0.15 (0.34 ± 0.13)]. All mfERG components showed strong significant correlations (R<sup>2</sup> ≥ 0.253, p < 0.001) between both electrodes for all eccentricities. Both electrodes allowed for the identification of age-related P1 changes, i.e., P1-amplitude reduction and peak-time delay in the older group. There was a trend to higher AUC for the DTL electrode to delineate these differences between age groups, which, however, failed to reach statistical significance.</p><p><strong>Conclusions: </strong>Both electrode types enable successful mfERG recordings. However, in compliant patients, the use of the DTL electrode appears preferable due to the larger amplitudes, higher signal-to-noise ratio and its better reflection of physiological changes, i.e., age effects. Nevertheless, skin electrodes appear a viable alternative for mfERG recordings in patients in whom the use of corneal electrodes is precluded, e.g., children and disabled patients.</p>","PeriodicalId":11207,"journal":{"name":"Documenta Ophthalmologica","volume":"146 1","pages":"67-78"},"PeriodicalIF":2.6000,"publicationDate":"2023-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9911471/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Documenta Ophthalmologica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10633-022-09912-9","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To compare mfERG recordings with the Dawson-Trick-Litzkow (DTL) and gold cup skin electrode in healthy young and old adults and to test the sensitivity of both electrodes to age-related changes in the responses.
Methods: Twenty participants aged 20-27 years ("young") and 20 participants aged 60-75 ("old") with a visual acuity of ≤ 0 logMAR were included. The mfERG responses were recorded simultaneously using DTL and skin electrodes. P1 amplitudes, peak times and signal-to-noise ratios (SNRs) were compared between both electrodes and across age groups, and correlation analyses were performed. The electrode's performance in discriminating between age groups was assessed via area under curve (AUC) of receiver operating characteristics.
Results: Both electrodes reflected the typical waveform of mfERG recordings. For the skin electrode, however, P1 amplitudes were significantly reduced (p < 0.001; reduction by over 70%), P1 peak times were significantly shorter (p < 0.001; by approx. 1.5 ms), and SNRs were reduced [(p < 0.001; logSNR ± SEM DTL young (old) vs gold cup: 0.79 ± 0.13 (0.71 ± 0.15) vs 0.37 ± 0.15 (0.34 ± 0.13)]. All mfERG components showed strong significant correlations (R2 ≥ 0.253, p < 0.001) between both electrodes for all eccentricities. Both electrodes allowed for the identification of age-related P1 changes, i.e., P1-amplitude reduction and peak-time delay in the older group. There was a trend to higher AUC for the DTL electrode to delineate these differences between age groups, which, however, failed to reach statistical significance.
Conclusions: Both electrode types enable successful mfERG recordings. However, in compliant patients, the use of the DTL electrode appears preferable due to the larger amplitudes, higher signal-to-noise ratio and its better reflection of physiological changes, i.e., age effects. Nevertheless, skin electrodes appear a viable alternative for mfERG recordings in patients in whom the use of corneal electrodes is precluded, e.g., children and disabled patients.
期刊介绍:
Documenta Ophthalmologica is an official publication of the International Society for Clinical Electrophysiology of Vision. The purpose of the journal is to promote the understanding and application of clinical electrophysiology of vision. Documenta Ophthalmologica will publish reviews, research articles, technical notes, brief reports and case studies which inform the readers about basic and clinical sciences related to visual electrodiagnosis and means to improve diagnosis and clinical management of patients using visual electrophysiology. Studies may involve animals or humans. In either case appropriate care must be taken to follow the Declaration of Helsinki for human subject or appropriate humane standards of animal care (e.g., the ARVO standards on Animal Care and Use).