{"title":"Establishment of national diagnostic dose reference levels (DRLs) for routine computed tomography examinations in Jordan","authors":"K. Radaideh, A. Al-Radaideh, R. Ramli, A. Saleh, Rasha Alshayeb","doi":"10.2478/pjmpe-2023-0003","DOIUrl":null,"url":null,"abstract":"Abstract Background: Dose reference levels (DRLs) are used as indicators as well as guidance for dose optimization and to ensure justification of appropriate dose for a given clinical indication. The main aims of this study were to establish local DRLs for each CT imaging protocol as a reference point to evaluate the radiation dose indices and to compare our DRLs with those established in other countries and against the internationally reported guidelines. Materials and methods: 2000 CT dose reports of different adult imaging protocols from January 2021 until April 2022 were collected retrospectively at different hospitals in Jordan. Data were collected from CT scans that were performed using different types and models of CT scanners and included four adult non-enhanced, helical CT imaging protocols; Head, Chest, Abdomen-Pelvis, and Chest-Abdomen-Pelvis. Results: The average doses of CTDIvol, DLP, and effective dose were (65.11 mGy, 1232.71 mGy·cm, 2.83 mSv) for the head scan, (16.6 mGy, 586.6 mGy·cm, 8.21 mSv) for the chest scan, (17.91 mGy, 929.9 mGy·cm, 13.9 mSv) for the abdomen-pelvis scan, and (19.3 mGy, 1152 mGy·cm, 17.25 mSv) for the chest-abdomen-pelvis scan. In comparison with results from different international studies, DLP values measured in the present study were lower for the chest-abdomen-pelvis and abdomen-pelvis CT scans, and higher for the head CT and chest CT scans. Conclusions: It is very important that each country establishes its own DRLs and compares them with those reported by other countries, especially the developed ones. It is also important that these levels are regularly updated.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":"6 1","pages":"26 - 34"},"PeriodicalIF":0.7000,"publicationDate":"2023-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2023-0003","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Abstract Background: Dose reference levels (DRLs) are used as indicators as well as guidance for dose optimization and to ensure justification of appropriate dose for a given clinical indication. The main aims of this study were to establish local DRLs for each CT imaging protocol as a reference point to evaluate the radiation dose indices and to compare our DRLs with those established in other countries and against the internationally reported guidelines. Materials and methods: 2000 CT dose reports of different adult imaging protocols from January 2021 until April 2022 were collected retrospectively at different hospitals in Jordan. Data were collected from CT scans that were performed using different types and models of CT scanners and included four adult non-enhanced, helical CT imaging protocols; Head, Chest, Abdomen-Pelvis, and Chest-Abdomen-Pelvis. Results: The average doses of CTDIvol, DLP, and effective dose were (65.11 mGy, 1232.71 mGy·cm, 2.83 mSv) for the head scan, (16.6 mGy, 586.6 mGy·cm, 8.21 mSv) for the chest scan, (17.91 mGy, 929.9 mGy·cm, 13.9 mSv) for the abdomen-pelvis scan, and (19.3 mGy, 1152 mGy·cm, 17.25 mSv) for the chest-abdomen-pelvis scan. In comparison with results from different international studies, DLP values measured in the present study were lower for the chest-abdomen-pelvis and abdomen-pelvis CT scans, and higher for the head CT and chest CT scans. Conclusions: It is very important that each country establishes its own DRLs and compares them with those reported by other countries, especially the developed ones. It is also important that these levels are regularly updated.
期刊介绍:
Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.