Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading

Cora M. Märtens, Juliane Schöpfel, S. Bollmann, A. Hannemann, S. Zylla, M. B. Dahl, Friederike Gauß, Josef Schedl, M. Nauck, A. Petersmann
{"title":"Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading","authors":"Cora M. Märtens, Juliane Schöpfel, S. Bollmann, A. Hannemann, S. Zylla, M. B. Dahl, Friederike Gauß, Josef Schedl, M. Nauck, A. Petersmann","doi":"10.1515/cclm-2022-0193","DOIUrl":null,"url":null,"abstract":"Abstract Objectives A carrier prototype by Aerocom® (Schwäbisch Gmünd, Germany) for pneumatic tube systems (PTS) is able to transport 9 blood tubes which are automatically fixed by closing the lid. In this study, we examined the influence of the transport on blood sample quality using the carrier prototype comparing to courier transport and a conventional carrier (AD160, Aerocom®). Methods Triplicate blood samples sets (1 lithium heparin, 1 EDTA, 1 sodium citrate) of 35 probands were split among the transportation methods: 1. courier, 2. conventional carrier, and 3. carrier prototype. After transport 51 measurands from clinical chemistry, hematology and coagulation were measured and compared. Results Overall, 49 of the investigated 51 measurands showed a good concordance among the three transport types, especially between the conventional carrier and the carrier prototype. Focusing on well-known hemolysis sensitive measurands, potassium showed no statistically significant differences. However, between courier and both carrier types lactate dehydrogenase (LDH) and free hemoglobin (fHb) showed statistically significant shifts, whereas the clinical impact of the identified differences was neglectable. The median concentration of fHb, for example, was 0.29 g/L (18 µmol/L), 0.31 g/L (19 µmol/L) and 0.32 g/L (20 µmol/L) for courier transport, conventional carrier and carrier prototype, respectively. These differences cannot be resolved analytically since the minimal difference (MD) for fHb is 0.052 g/L (3.23 µmol/L), at this concentration. Conclusions The carrier prototype by Aerocom® is suitable for transportation of diagnostic blood samples. The overall workflow is improved by decreasing hands-on-time on the ward and laboratory while minimizing the risk of incorrectly packed carriers.","PeriodicalId":10388,"journal":{"name":"Clinical Chemistry and Laboratory Medicine (CCLM)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-05-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Chemistry and Laboratory Medicine (CCLM)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/cclm-2022-0193","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Abstract Objectives A carrier prototype by Aerocom® (Schwäbisch Gmünd, Germany) for pneumatic tube systems (PTS) is able to transport 9 blood tubes which are automatically fixed by closing the lid. In this study, we examined the influence of the transport on blood sample quality using the carrier prototype comparing to courier transport and a conventional carrier (AD160, Aerocom®). Methods Triplicate blood samples sets (1 lithium heparin, 1 EDTA, 1 sodium citrate) of 35 probands were split among the transportation methods: 1. courier, 2. conventional carrier, and 3. carrier prototype. After transport 51 measurands from clinical chemistry, hematology and coagulation were measured and compared. Results Overall, 49 of the investigated 51 measurands showed a good concordance among the three transport types, especially between the conventional carrier and the carrier prototype. Focusing on well-known hemolysis sensitive measurands, potassium showed no statistically significant differences. However, between courier and both carrier types lactate dehydrogenase (LDH) and free hemoglobin (fHb) showed statistically significant shifts, whereas the clinical impact of the identified differences was neglectable. The median concentration of fHb, for example, was 0.29 g/L (18 µmol/L), 0.31 g/L (19 µmol/L) and 0.32 g/L (20 µmol/L) for courier transport, conventional carrier and carrier prototype, respectively. These differences cannot be resolved analytically since the minimal difference (MD) for fHb is 0.052 g/L (3.23 µmol/L), at this concentration. Conclusions The carrier prototype by Aerocom® is suitable for transportation of diagnostic blood samples. The overall workflow is improved by decreasing hands-on-time on the ward and laboratory while minimizing the risk of incorrectly packed carriers.
带固定机构的气动管道系统载体原型的评估,允许自动卸载
Aerocom®(Schwäbisch gm nd,德国)为气动管系统(PTS)设计了一种载体原型,能够通过关闭盖子自动固定输送9根血管。在本研究中,我们使用载体原型比较了快递运输和传统载体(AD160, Aerocom®)运输对血液样本质量的影响。方法将35个先证者的3份血样组(1份肝素锂、1份EDTA、1份柠檬酸钠)按不同的运输方式进行拆分。快递,2。3.常规航母;载体原型。转运后对51项临床化学、血液学、凝血指标进行测定比较。结果调查的51项测量中,有49项在三种输运类型之间表现出较好的一致性,特别是传统载体与载体原型之间的一致性。关注众所周知的溶血敏感指标,钾没有统计学上的显著差异。然而,在信使和两种载体类型之间,乳酸脱氢酶(LDH)和游离血红蛋白(fHb)表现出统计学上的显著变化,而鉴定差异的临床影响可以忽略不计。例如,在快递运输、常规载体和载体原型中,fHb的中位数浓度分别为0.29 g/L(18µmol/L)、0.31 g/L(19µmol/L)和0.32 g/L(20µmol/L)。由于在该浓度下,fHb的最小差异(MD)为0.052 g/L(3.23µmol/L),因此无法用分析方法解决这些差异。结论Aerocom®载体原型适用于诊断性血液样本的运输。通过减少在病房和实验室的动手时间,同时最大限度地降低错误包装载体的风险,改善了整个工作流程。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信