Comparison of the isocenter verification methods and calculating imaging offset of medical linear accelerators.

Olga Bąk, K. Świątek, M. Paluszyńska
{"title":"Comparison of the isocenter verification methods and calculating imaging offset of medical linear accelerators.","authors":"Olga Bąk, K. Świątek, M. Paluszyńska","doi":"10.21641/los.2022.19.2.229","DOIUrl":null,"url":null,"abstract":"The aim of the presented work was to compare two methods of verification of the medical accelerator’s isocenter and determine the imaging offset. Two methods were used for verification - Machine Performence Check (MPC) application with the IsoCal phantom and the Artiscan software with the PTW Isoball Phantom. Each of the analyzed methods provides an alternative way to the time-consuming and costly verification of the isocenter on radiochromic or gafchromic dosimetry films. \n                Measurements were performed for ten days and the obtained data were compared for both methods. Using statistical tests, concordance was determined for the results obtained by the two methods. The data were presented in trend function graphs. \n                Analysis of the results shows that it is not possible to unequivocally state that the methods are not equivalent. In addition, the AAPM TG 142 report assumes that MPC is not a sufficient test for checking medical linear accelerator’s isocentre.","PeriodicalId":18098,"journal":{"name":"Letters in Oncology Science","volume":"27 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-12-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Letters in Oncology Science","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.21641/los.2022.19.2.229","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The aim of the presented work was to compare two methods of verification of the medical accelerator’s isocenter and determine the imaging offset. Two methods were used for verification - Machine Performence Check (MPC) application with the IsoCal phantom and the Artiscan software with the PTW Isoball Phantom. Each of the analyzed methods provides an alternative way to the time-consuming and costly verification of the isocenter on radiochromic or gafchromic dosimetry films.                 Measurements were performed for ten days and the obtained data were compared for both methods. Using statistical tests, concordance was determined for the results obtained by the two methods. The data were presented in trend function graphs.                 Analysis of the results shows that it is not possible to unequivocally state that the methods are not equivalent. In addition, the AAPM TG 142 report assumes that MPC is not a sufficient test for checking medical linear accelerator’s isocentre.
医用直线加速器等心验证方法比较及成像偏移量计算。
提出的工作的目的是比较两种方法的验证医学加速器的等心和确定成像偏移。使用两种方法进行验证-使用IsoCal幻影的机器性能检查(MPC)应用程序和使用PTW Isoball幻影的Artiscan软件。每一种分析方法都提供了一种替代方法,以替代耗时且昂贵的放射致色或渐变色剂量测定膜上的等中心验证。测量进行了十天,并比较了两种方法获得的数据。通过统计检验,确定两种方法所得结果的一致性。数据以趋势函数图表示。对结果的分析表明,不可能毫不含糊地说这些方法不等效。此外,AAPM TG 142报告假设MPC不是检查医用直线加速器等心的充分测试。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信