Imprecision Evaluation of Self-Monitoring of Blood Cholesterol (SMBC)Handheld Point of Care Testing Devices: Elemark⢠and Cardiochek PA

J. Bolodeoku., S. Pinkney
{"title":"Imprecision Evaluation of Self-Monitoring of Blood Cholesterol (SMBC)Handheld Point of Care Testing Devices: Elemark⢠and Cardiochek PA","authors":"J. Bolodeoku., S. Pinkney","doi":"10.21767/2386-5180.100290","DOIUrl":null,"url":null,"abstract":"The aim of this study was to determine the imprecision of two cholesterol point of care testing devices, the very established CardioChek PA and recently the Elemark™ which is a mobile blood testing device capable of testing multiple parameters for the management of hypercholesterolaemia undergoing certification. The imprecision study of the Elemark™ and CardioChek PA were performed using blood donations from 3 healthy normal lipidemic subjects. There were 10 repetitions for the within run imprecision and the number of measurement days was 6 for between day imprecision. This study showed comparable within run and between day imprecision between the CardioChek PA and the Elemark™. Based on the imprecision guidelines from the NCEP and CRMLN both devices showed reasonable within run imprecision for their total cholesterol and HDLcholesterol estimations but not the LDL-cholesterol calculations. The between day imprecision for both devices was reasonably consistent with the published within person biological variability. With more accuracy and imprecision studies being carried out in understanding these cholesterol testing point of care devices, their value, or lack of, in the lipid monitoring would become more apparent.","PeriodicalId":8195,"journal":{"name":"Annals of Clinical and Laboratory Research","volume":"1245 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Clinical and Laboratory Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.21767/2386-5180.100290","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5

Abstract

The aim of this study was to determine the imprecision of two cholesterol point of care testing devices, the very established CardioChek PA and recently the Elemark™ which is a mobile blood testing device capable of testing multiple parameters for the management of hypercholesterolaemia undergoing certification. The imprecision study of the Elemark™ and CardioChek PA were performed using blood donations from 3 healthy normal lipidemic subjects. There were 10 repetitions for the within run imprecision and the number of measurement days was 6 for between day imprecision. This study showed comparable within run and between day imprecision between the CardioChek PA and the Elemark™. Based on the imprecision guidelines from the NCEP and CRMLN both devices showed reasonable within run imprecision for their total cholesterol and HDLcholesterol estimations but not the LDL-cholesterol calculations. The between day imprecision for both devices was reasonably consistent with the published within person biological variability. With more accuracy and imprecision studies being carried out in understanding these cholesterol testing point of care devices, their value, or lack of, in the lipid monitoring would become more apparent.
血液胆固醇自我监测(SMBC)手持式护理点检测设备的不精确评估:elemarkⅱ和Cardiochek PA
本研究的目的是确定两种胆固醇护理点检测设备的不精确性,一种是非常成熟的CardioChek PA,另一种是最近的Elemark™,这是一种移动血液检测设备,能够检测高胆固醇血症管理的多个参数,正在接受认证。elmark™和CardioChek PA的不精确研究是使用3名健康正常血脂受试者的献血进行的。运行内不精密度重复10次,运行间不精密度测量天数为6天。该研究显示CardioChek PA和Elemark™在运行和日间的不精确性具有可比性。根据NCEP和CRMLN的不精确指南,这两种设备对总胆固醇和高密度脂蛋白胆固醇的估计都显示出合理的运行内不精确,但对低密度脂蛋白胆固醇的计算却没有。两种装置的日间不精确性与公布的人体内生物变异性相当一致。随着越来越多的准确性和不准确性的研究被开展,以了解这些胆固醇检测点设备,他们的价值,或缺乏,在血脂监测将变得更加明显。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信