Personal Pump Reliability Tester

M. T. Abell, J. Agee, P. A. Hentz, M. Anastas, J. E. Boyd
{"title":"Personal Pump Reliability Tester","authors":"M. T. Abell, J. Agee, P. A. Hentz, M. Anastas, J. E. Boyd","doi":"10.1080/08828032.1989.10390400","DOIUrl":null,"url":null,"abstract":"Abstract A test instrument has been built that measures, displays, and records the flow rate of up to 15 personal sampling pumps. The “Pump Tester” is typically run for eight hours with five or six pumps of one type and displays a graph of flow rate over time for each pump. The flow record for each pump contains up to 240 data points and, although not continuous, provides a useful record of pump performance over long test periods. The numerical flow data can be printed and saved to disk. At the end of the test, the flow-rate graph for each pump is printed. The flow-rate graph can be added to the maintenance and calibration record for the pump to document its performance. Occasionally, erratic pump performance is documented that would not have been detected by flow-rate checks at the beginning and end of an eight-hour period. The Pump Tester can also be used for more formal studies to demonstrate the relative performance of different kinds of pumps under different conditions.","PeriodicalId":8049,"journal":{"name":"Applied Industrial Hygiene","volume":"10 1","pages":"143-146"},"PeriodicalIF":0.0000,"publicationDate":"1989-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applied Industrial Hygiene","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/08828032.1989.10390400","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Abstract A test instrument has been built that measures, displays, and records the flow rate of up to 15 personal sampling pumps. The “Pump Tester” is typically run for eight hours with five or six pumps of one type and displays a graph of flow rate over time for each pump. The flow record for each pump contains up to 240 data points and, although not continuous, provides a useful record of pump performance over long test periods. The numerical flow data can be printed and saved to disk. At the end of the test, the flow-rate graph for each pump is printed. The flow-rate graph can be added to the maintenance and calibration record for the pump to document its performance. Occasionally, erratic pump performance is documented that would not have been detected by flow-rate checks at the beginning and end of an eight-hour period. The Pump Tester can also be used for more formal studies to demonstrate the relative performance of different kinds of pumps under different conditions.
个人泵可靠性测试仪
摘要建立了一种测试仪器,可以测量、显示和记录多达15个个人采样泵的流量。“泵测试仪”通常在一种类型的5或6台泵上运行8小时,并显示每个泵的流量随时间变化的图表。每台泵的流量记录包含多达240个数据点,虽然不是连续的,但可以提供长时间测试期间泵性能的有用记录。数值流数据可以打印并保存到磁盘上。在测试结束时,打印出每个泵的流量曲线图。流量图可以添加到泵的维护和校准记录中,以记录其性能。偶尔会记录到泵的不稳定性能,这是在8小时周期的开始和结束时通过流量检查无法检测到的。泵测试仪也可以用于更正式的研究,以证明不同种类的泵在不同条件下的相对性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信