{"title":"Effect of Abutments Design on Wear of Locator Attachments in Effect of Abutments Design on Wear of Locator Attachments in Implant Retained Mandibular Overdenture","authors":"M. Fayed, M. Abdelaziz, Hossam I. Nassar","doi":"10.54623/fdj.80110","DOIUrl":null,"url":null,"abstract":"Purpose: The aim of this study is to compare the effect of abutments design on wear behavior of Locator attachments in implant-retained mandibular overdenture. Materials and Methods: Two epoxy models representing an edentulous mandible were used, two implants were placed in each model. Twenty-four locator attachments were used (twelve in each study group). Cycles of Insertion and removal were performed using a universal testing machine resembling 3 years of clinical use, then each abutment was scanned by scanning electron microscope (SEM) before and after insertion and removal cycles. Data were compared qualitatively using a specialized computer software Results: There was a statistically significant difference in wear between the two studied locator groups P=0000*. Locator F-TX showed 29±1.704 % of surface change. PEEKLOC. Locator showed 21±2.090 %of surface change. Conclusions: PEEKLoc. abutment design showed more wear resistance than the recently introduced Locator F-TX.","PeriodicalId":100562,"journal":{"name":"Future Dental Journal","volume":"7 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Future Dental Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.54623/fdj.80110","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The aim of this study is to compare the effect of abutments design on wear behavior of Locator attachments in implant-retained mandibular overdenture. Materials and Methods: Two epoxy models representing an edentulous mandible were used, two implants were placed in each model. Twenty-four locator attachments were used (twelve in each study group). Cycles of Insertion and removal were performed using a universal testing machine resembling 3 years of clinical use, then each abutment was scanned by scanning electron microscope (SEM) before and after insertion and removal cycles. Data were compared qualitatively using a specialized computer software Results: There was a statistically significant difference in wear between the two studied locator groups P=0000*. Locator F-TX showed 29±1.704 % of surface change. PEEKLOC. Locator showed 21±2.090 %of surface change. Conclusions: PEEKLoc. abutment design showed more wear resistance than the recently introduced Locator F-TX.
基台设计对定位器附着体磨损的影响基台设计对种植下颌覆盖义齿定位器附着体磨损的影响
目的:比较不同基台设计对下颌种植覆盖义齿Locator附着体磨损行为的影响。材料与方法:采用两个环氧树脂模型,每个模型放置两个种植体。使用了24个定位器附件(每个研究组12个)。采用类似临床使用3年的通用试验机进行插入和移除周期,然后在插入和移除周期前后对每个基台进行扫描电镜(SEM)扫描。使用专门的计算机软件对数据进行定性比较。结果:所研究的两组定位器之间的磨损差异具有统计学意义P=0000*。定位仪F-TX显示29±1.704%的表面变化。PEEKLOC。定位仪显示表面变化21±2.090%。结论:PEEKLoc。基台设计比最近推出的Locator F-TX更具耐磨性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信