{"title":"Quality control test of fixed x-ray units at Bir hospital and National trauma center","authors":"K. Adhikari, K. S. Nyaichyai, B. Chand, Hanson Khadka, Saroj Sharma, T. Khanal","doi":"10.3126/sw.v15i15.45648","DOIUrl":null,"url":null,"abstract":"One of the important goals of the use of X-rays is to diagnose a disease and its extent and minimizing the adverse effect. In diagnostic X-ray, there is an important issue concerning the quality of the image and the exposure of radiation to the patient. Regular practice of quality control in diagnostic X-ray is essential to provide good quality images, decrease hazard to patients, and enhance the appropriate diagnosis. Quality control tests were carried out at six fixed X-ray units at National Academy of Medical Sciences, Bir Hospital and National Trauma Center. Parameters like kVp accuracy, timer accuracy, Reproducibility of kV, time and dose, variable mA and radiation output Linearity with the acceptance limits were checked. Radiation leakage, spatial resolution, contrast resolution and field alignment tests were also carried out. Quality control tests of kVp dose, dose rate, HVL, mAs, beam alignment and leakage radiation shows mixed results. The measured values were within limit in four X-ray units. Until now we do not have any regulations in Nepal on acceptance testing of x-ray units after installation. The need for quality assurance of X-ray needs to be taken to avert detrimental effects to patients and staff. Institutes must introduce a mandatory system for acceptance tests of X-ray when installed and alongside regular quality control programs.","PeriodicalId":21637,"journal":{"name":"Scientific World","volume":"57 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-06-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scientific World","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3126/sw.v15i15.45648","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

One of the important goals of the use of X-rays is to diagnose a disease and its extent and minimizing the adverse effect. In diagnostic X-ray, there is an important issue concerning the quality of the image and the exposure of radiation to the patient. Regular practice of quality control in diagnostic X-ray is essential to provide good quality images, decrease hazard to patients, and enhance the appropriate diagnosis. Quality control tests were carried out at six fixed X-ray units at National Academy of Medical Sciences, Bir Hospital and National Trauma Center. Parameters like kVp accuracy, timer accuracy, Reproducibility of kV, time and dose, variable mA and radiation output Linearity with the acceptance limits were checked. Radiation leakage, spatial resolution, contrast resolution and field alignment tests were also carried out. Quality control tests of kVp dose, dose rate, HVL, mAs, beam alignment and leakage radiation shows mixed results. The measured values were within limit in four X-ray units. Until now we do not have any regulations in Nepal on acceptance testing of x-ray units after installation. The need for quality assurance of X-ray needs to be taken to avert detrimental effects to patients and staff. Institutes must introduce a mandatory system for acceptance tests of X-ray when installed and alongside regular quality control programs.
比尔医院和国家创伤中心固定x光机质量控制试验
使用x射线的一个重要目的是诊断疾病及其程度,并尽量减少不良影响。在诊断x射线,有一个重要的问题,关于图像的质量和辐射暴露给病人。诊断x线质量控制的常规做法对于提供高质量的图像,减少对患者的危害和提高适当的诊断是必不可少的。在国家医学科学院、比尔医院和国家创伤中心的六个固定x光机进行了质量控制测试。kVp精度、计时器精度、kV再现性、时间和剂量、可变mA和辐射输出等参数与接受限呈线性关系。还进行了辐射泄漏、空间分辨率、对比度分辨率和现场对准测试。kVp剂量、剂量率、HVL、mAs、光束对准和泄漏辐射的质量控制试验结果好坏参半。测量值在四个x射线单位的限制范围内。到目前为止,我们在尼泊尔还没有任何关于x射线设备安装后验收测试的规定。必须保证x射线的质量,以避免对病人和工作人员产生有害影响。研究所必须在安装时引入强制性的x射线验收测试系统,以及定期的质量控制程序。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信