Compliance Test of Percentage Depth Dose (PDD) Using LINAC Machine Brand Electa Precise against American Association of Physicists in Medicine (AAPM) PDD with Electron Beam Energy Variation at Radiotherapy Installation RSUP Prof. Ngoerah in Denpasar

IF 0.1 4区 医学 Q4 INFECTIOUS DISEASES
Naumi Paramita Putri Monicasari, N. Ratini, I. W. Sudarsana, Ida Bagus Made Suryatika, I. P. Suardana, I. Widagda, A. Gunawan, R. Irhas
{"title":"Compliance Test of Percentage Depth Dose (PDD) Using LINAC Machine Brand Electa Precise against American Association of Physicists in Medicine (AAPM) PDD with Electron Beam Energy Variation at Radiotherapy Installation RSUP Prof. Ngoerah in Denpasar","authors":"Naumi Paramita Putri Monicasari, N. Ratini, I. W. Sudarsana, Ida Bagus Made Suryatika, I. P. Suardana, I. Widagda, A. Gunawan, R. Irhas","doi":"10.9734/ajmah/2023/v21i10897","DOIUrl":null,"url":null,"abstract":"The purpose of the study was to determine the value of PDD against energy variations in the electron beam using the Elekta Precise Brand LINAC aircraft, and to determine the suitability of the PDD value against energy variations in the electron beam with the standards set by AAPM. PDD value data obtained from the results of research using Sun Nuclear Dosimetry Software (SNC Dosimetry). The results of the study will then be compared with the standard Tolerance value set by AAPM of a maximum of 2%. From the deviation value of each 4 MeV energy is 1.1% - 1.8%, 6 MeV energy is 1.1% - 1.6%, and 8 MeV energy is 1.0% - 1.2%, the deviation error at 4, 6, and 8 MeV energy is in accordance with the tolerance limit set by AAPM which is 2%. So that the Elekta Precise LINAC machine at the Radiotherapy Installation of Prof. Ngoerah Denpasar Hospital is feasible to operate to receive patients.","PeriodicalId":49491,"journal":{"name":"Southeast Asian Journal of Tropical Medicine and Public Health","volume":"44 1","pages":""},"PeriodicalIF":0.1000,"publicationDate":"2023-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Southeast Asian Journal of Tropical Medicine and Public Health","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.9734/ajmah/2023/v21i10897","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
引用次数: 0

Abstract

The purpose of the study was to determine the value of PDD against energy variations in the electron beam using the Elekta Precise Brand LINAC aircraft, and to determine the suitability of the PDD value against energy variations in the electron beam with the standards set by AAPM. PDD value data obtained from the results of research using Sun Nuclear Dosimetry Software (SNC Dosimetry). The results of the study will then be compared with the standard Tolerance value set by AAPM of a maximum of 2%. From the deviation value of each 4 MeV energy is 1.1% - 1.8%, 6 MeV energy is 1.1% - 1.6%, and 8 MeV energy is 1.0% - 1.2%, the deviation error at 4, 6, and 8 MeV energy is in accordance with the tolerance limit set by AAPM which is 2%. So that the Elekta Precise LINAC machine at the Radiotherapy Installation of Prof. Ngoerah Denpasar Hospital is feasible to operate to receive patients.
使用LINAC品牌Electa precision机器对美国医学物理学家协会(AAPM)放射治疗装置中电子束能量变化的百分比深度剂量(PDD)的依从性测试
本研究的目的是使用Elekta Precise Brand LINAC飞机确定PDD值对电子束能量变化的影响,并根据AAPM设定的标准确定PDD值对电子束能量变化的适用性。PDD值数据来自使用Sun核剂量测定软件(SNC剂量测定)的研究结果。然后将研究结果与AAPM设定的最高2%的标准公差值进行比较。从4 MeV能量的偏差值为1.1% ~ 1.8%,6 MeV能量的偏差值为1.1% ~ 1.6%,8 MeV能量的偏差值为1.0% ~ 1.2%来看,4、6、8 MeV能量的偏差误差符合AAPM规定的2%的容差限值。因此,Ngoerah Denpasar教授医院放射治疗装置的Elekta精密LINAC机器可以操作以接收病人。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Southeast Asian Journal of Tropical Medicine and Public Health
Southeast Asian Journal of Tropical Medicine and Public Health PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH-INFECTIOUS DISEASES
CiteScore
0.40
自引率
0.00%
发文量
0
审稿时长
3-8 weeks
期刊介绍: The SEAMEO* Regional Tropical Medicine and Public Health Project was established in 1967 to help improve the health and standard of living of the peoples of Southeast Asia by pooling manpower resources of the participating SEAMEO member countries in a cooperative endeavor to develop and upgrade the research and training capabilities of the existing facilities in these countries. By promoting effective regional cooperation among the participating national centers, it is hoped to minimize waste in duplication of programs and activities. In 1992 the Project was renamed the SEAMEO Regional Tropical Medicine and Public Health Network.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信