{"title":"Tripolar nerve cuff recording: stimulus artifact, EMG and the recorded nerve signal","authors":"J. Struijk, M. Thomsen","doi":"10.1109/IEMBS.1995.579534","DOIUrl":null,"url":null,"abstract":"Properties of nerve cuff recording electrodes were analyzed. Tripolar cuff electrodes have to be described essentially different for (propagating) nerve signals inside the cuff, and electrical muscle activity and stimulus artifacts arising from sources outside the cuff. It was experimentally shown that the signals originating outside the cuff are differently and much stronger influenced by the addition of a balancing resister to the outer cuff electrodes than the nerve signal.","PeriodicalId":20509,"journal":{"name":"Proceedings of 17th International Conference of the Engineering in Medicine and Biology Society","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1995-09-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"72","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of 17th International Conference of the Engineering in Medicine and Biology Society","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IEMBS.1995.579534","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 72
Abstract
Properties of nerve cuff recording electrodes were analyzed. Tripolar cuff electrodes have to be described essentially different for (propagating) nerve signals inside the cuff, and electrical muscle activity and stimulus artifacts arising from sources outside the cuff. It was experimentally shown that the signals originating outside the cuff are differently and much stronger influenced by the addition of a balancing resister to the outer cuff electrodes than the nerve signal.