Variaciones en los parámetros cuantitativos y en la puntuación de la escala de Deauville aplicando diferentes algoritmos de reconstrucción en las imágenes PET/TC con [18F]FDG de pacientes con linfoma
IF 1.6 4区 医学Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
M. Genc , N. Yildirim , N. Coskun , E. Ozdemir , S. Turkolmez
{"title":"Variaciones en los parámetros cuantitativos y en la puntuación de la escala de Deauville aplicando diferentes algoritmos de reconstrucción en las imágenes PET/TC con [18F]FDG de pacientes con linfoma","authors":"M. Genc , N. Yildirim , N. Coskun , E. Ozdemir , S. Turkolmez","doi":"10.1016/j.remn.2023.07.002","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction and Objectives</h3><p>18F-FDG PET with the Deauville score (DS) is a unique semiquantitative method for lymphoma. However, type of standard uptake values (max, mean, and peak) reconstruction algorithms could affect DS. We compared the Bayesian Penalized Likelihood reconstruction algorithm (BPL) with Ordered Subsets Expectation Maximization (OSEM) on quantitative parameters and DS in lymphoma. We investigated the effect of the size of the lymph node on quantitative variation.</p></div><div><h3>Patients and Methods</h3><p>Raw PET data of 255 lymphoma patients were reconstructed separately with Q.Clear (GE Healthcare), a BPL, and SharpIR (GE Healthcare), an OSEM algorithm. In both images, each patient's liver, mediastinal blood pool, and SUVs (SUVmax, SUVmean, and SUVpeak) of a total of 487 lesions selected from the patients were performed. DSmax, DSmean, and DSpeak were compared.</p></div><div><h3>Results</h3><p>In our study, DS increased significantly with BPL (p<!--> <!--><<!--> <!-->0.001), and the DS increased to 4-5 in 30 patients evaluated as 1-2-3 with OSEM. It was found that the quantitative values of the lymph nodes increased statistically with BPL (p<!--> <!--><<!--> <!-->0.001), and the liver from the reference regions were significantly decreased (p<!--> <!--><<!--> <!-->0.001). In addition, difference in lymph node was independently associated with size of lesion and was significantly more pronounced in small lesions (p<!--> <!--><<!--> <!-->0.001). The effects of BPL algorithm were more pronounced in SUVmax than in SUVmean and SUVpeak. DS-mean and DS-peak scores were less changed by BPL than DS-max.</p></div><div><h3>Conclusion</h3><p>Different reconstruction algorithms in FDG PET/CT affect the quantitative evaluation. That variation may affect the change in DS in lymphoma patients, thus affecting patient management.</p></div>","PeriodicalId":48986,"journal":{"name":"Revista Espanola De Medicina Nuclear E Imagen Molecular","volume":"42 6","pages":"Pages 388-392"},"PeriodicalIF":1.6000,"publicationDate":"2023-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista Espanola De Medicina Nuclear E Imagen Molecular","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2253654X23000835","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction and Objectives
18F-FDG PET with the Deauville score (DS) is a unique semiquantitative method for lymphoma. However, type of standard uptake values (max, mean, and peak) reconstruction algorithms could affect DS. We compared the Bayesian Penalized Likelihood reconstruction algorithm (BPL) with Ordered Subsets Expectation Maximization (OSEM) on quantitative parameters and DS in lymphoma. We investigated the effect of the size of the lymph node on quantitative variation.
Patients and Methods
Raw PET data of 255 lymphoma patients were reconstructed separately with Q.Clear (GE Healthcare), a BPL, and SharpIR (GE Healthcare), an OSEM algorithm. In both images, each patient's liver, mediastinal blood pool, and SUVs (SUVmax, SUVmean, and SUVpeak) of a total of 487 lesions selected from the patients were performed. DSmax, DSmean, and DSpeak were compared.
Results
In our study, DS increased significantly with BPL (p < 0.001), and the DS increased to 4-5 in 30 patients evaluated as 1-2-3 with OSEM. It was found that the quantitative values of the lymph nodes increased statistically with BPL (p < 0.001), and the liver from the reference regions were significantly decreased (p < 0.001). In addition, difference in lymph node was independently associated with size of lesion and was significantly more pronounced in small lesions (p < 0.001). The effects of BPL algorithm were more pronounced in SUVmax than in SUVmean and SUVpeak. DS-mean and DS-peak scores were less changed by BPL than DS-max.
Conclusion
Different reconstruction algorithms in FDG PET/CT affect the quantitative evaluation. That variation may affect the change in DS in lymphoma patients, thus affecting patient management.
期刊介绍:
The Revista Española de Medicina Nuclear e Imagen Molecular (Spanish Journal of Nuclear Medicine and Molecular Imaging), was founded in 1982, and is the official journal of the Spanish Society of Nuclear Medicine and Molecular Imaging, which has more than 700 members.
The Journal, which publishes 6 regular issues per year, has the promotion of research and continuing education in all fields of Nuclear Medicine as its main aim. For this, its principal sections are Originals, Clinical Notes, Images of Interest, and Special Collaboration articles.