Dose Evaluation of Head and Neck Cancer IMRT Treatment Planning Based on Gamma Index Analysis of Varian Halcyon 2.0 Linac

IF 0.6 Q4 NUCLEAR SCIENCE & TECHNOLOGY
W. Purwati, F. Suhaimi, W. Wibowo, S. A. Pawiro
{"title":"Dose Evaluation of Head and Neck Cancer IMRT Treatment Planning Based on Gamma Index Analysis of Varian Halcyon 2.0 Linac","authors":"W. Purwati, F. Suhaimi, W. Wibowo, S. A. Pawiro","doi":"10.55981/aij.2023.1203","DOIUrl":null,"url":null,"abstract":"Varian Halcyon 2.0 linear accelerator was launched and became available for clinical use in 2018. Therefore, it is necessary to evaluate the accuracy of exit fluence of the Halcyon 2.0 for quality assurance (QA) of head and neck cancer treatment planning, pretreatment, and treatment. The accuracy of the exit fluence for twenty treatment plannings has been evaluated by conducting gamma analysis for QA pretreatment and treatment in each field and composite field by using criteria for gamma index 3 %/3 mm and 2 %/2 mm. The QA pretreatment results are in the average value for each criterion for each field and composite fields on actual gantry angle and null gantry angle with gamma passing rate (GPR) of over 99 % (range 99.78 %-99.95 %) The total treatments consisted of 2717 fractions. The analysis results of GPR for fields were 99.32 % and 97.74 % for gamma indexes of 3 %/3 mm and 2 %/2 mm, respectively. In addition, the analysis results of GPR for composites were 95.46 % and 81.38 % for gamma indexes of 3 %/3 mm and 2 %/2 mm, respectively. Based on this result, the average GPRs of QA pretreatment are ≈ 99 % of the total pixels. This means the prediction dose of Varian Halcyon 2.0 is accurate. The average GPRs of treatment is nearly  90 %, showing that Varian Halcyon 2.0 is effective for creating treatment plans for complex cases.","PeriodicalId":8647,"journal":{"name":"Atom Indonesia","volume":" ","pages":""},"PeriodicalIF":0.6000,"publicationDate":"2023-03-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Atom Indonesia","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.55981/aij.2023.1203","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"NUCLEAR SCIENCE & TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Varian Halcyon 2.0 linear accelerator was launched and became available for clinical use in 2018. Therefore, it is necessary to evaluate the accuracy of exit fluence of the Halcyon 2.0 for quality assurance (QA) of head and neck cancer treatment planning, pretreatment, and treatment. The accuracy of the exit fluence for twenty treatment plannings has been evaluated by conducting gamma analysis for QA pretreatment and treatment in each field and composite field by using criteria for gamma index 3 %/3 mm and 2 %/2 mm. The QA pretreatment results are in the average value for each criterion for each field and composite fields on actual gantry angle and null gantry angle with gamma passing rate (GPR) of over 99 % (range 99.78 %-99.95 %) The total treatments consisted of 2717 fractions. The analysis results of GPR for fields were 99.32 % and 97.74 % for gamma indexes of 3 %/3 mm and 2 %/2 mm, respectively. In addition, the analysis results of GPR for composites were 95.46 % and 81.38 % for gamma indexes of 3 %/3 mm and 2 %/2 mm, respectively. Based on this result, the average GPRs of QA pretreatment are ≈ 99 % of the total pixels. This means the prediction dose of Varian Halcyon 2.0 is accurate. The average GPRs of treatment is nearly  90 %, showing that Varian Halcyon 2.0 is effective for creating treatment plans for complex cases.
基于Varian Halcyon 2.0 Linac伽玛指数分析的头颈部癌症IMRT治疗计划的剂量评估
瓦里安Halcyon 2.0线性加速器于2018年推出并投入临床使用。因此,有必要评估Halcyon 2.0退出通量的准确性,以确保癌症头颈部治疗计划、预处理和治疗的质量保证(QA)。通过使用伽马指数3%/3mm和2%/2mm的标准对每个场和复合场的QA预处理和处理进行伽马分析,评估了20个处理计划的出口通量的准确性。QA预处理结果为每个场和复合场的每个标准在实际门架角和零门架角上的平均值,伽马通过率(GPR)超过99%(范围99.78%至99.95%)。总处理由2717个部分组成。γ指数为3%/3mm和2%/2mm时,GPR对野外的分析结果分别为99.32%和97.74%。此外,对于3%/3mm和2%/2mm的伽马指数,复合材料的GPR分析结果分别为95.46%和81.38%。基于此结果,QA预处理的平均GPR约为总像素的99%。这意味着Varian Halcyon 2.0的预测剂量是准确的。治疗的平均GPRs接近90%,这表明Varian Halcyon 2.0对于制定复杂病例的治疗计划是有效的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Atom Indonesia
Atom Indonesia NUCLEAR SCIENCE & TECHNOLOGY-
CiteScore
1.00
自引率
0.00%
发文量
20
审稿时长
16 weeks
期刊介绍: The focus of Atom Indonesia is research and development in nuclear science and technology. The scope of this journal covers experimental and analytical research in nuclear science and technology. The topics include nuclear physics, reactor physics, radioactive waste, fuel element, radioisotopes, radiopharmacy, radiation, and neutron scattering, as well as their utilization in agriculture, industry, health, environment, energy, material science and technology, and related fields.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信