{"title":"Angular dependence of the optically stimulated luminescent dosimeter","authors":"Wang Qi, Yuan Jilong, Guo Zhaohui, Cheng Jinsheng","doi":"10.13491/J.ISSN.1004-714X.2021.01.009","DOIUrl":null,"url":null,"abstract":"Objective To study the angular dependence of optically stimulated luminescent dosimeter (OSLD)\n with solid phantoms under SSDL radiation level 60Co radiation field, and to discuss the possibility of OSLD in volumetric modulated\n arc therapy (VMAT) and other rotating irradiation dose audit.\n Methods OSLDs were embedded in the two phantoms with the same size and material, respectively.\n The phantom 1 was set to make the first OSLD perpendicular to the beam, and the phantom\n 2 was set to make the second OSLD parallel to the beam. The OSLDs were irradiated\n at 8 angles: 0°, 45°, 90°, 135°, 180°, –45°, –90, –135°. The counts of the OSLSs were\n read and the response of each angle which normalized to 0° were calculated.\n Results When the OSLDs are perpendicular to the beam, the angular response is between –6.76%\n 〜+1.5%, with the maximum angular dependence at 90° and –90°. When the OSLDs are parallel\n to the beam, the angular response is between –1.74%〜+1.67%, below 2%.\n Conclusion It is better to correct the sensitivity of dosimeters by Element Correction Factors\n (ECF) for dose audit. Under the condition of rotating irradiation, OSLD shoud be set\n parallel to the beam, which can better reduce the influence of angular dependence\n and facilitate further application research of VMAT dose audit.\n 摘要: 目的 结合定制固体模体研宄光致发光剂量计(OSLD)在放疗水平 60Co标准辐射场中的角响应性能, 探讨OSLD在容积旋转调强放射治疗(VMAT)等旋转照射剂量核查中应用的可行性。\n 方法 将OSLD嵌入2个外观尺寸以 及材料均相同的定制固体模体, 模体1使OSLD与射线垂直, 模体2使OSLD与射线平行, 照射8个角度:0°、45°、 90°、135°、180°、–45°、–90°和–135°,\n 读取每个角度的OSLD计数, 计算每个角度相对于0°的角响应值。\n 结果 当OSLD 垂直于射线照射时, 角响应在–6.76%〜+1.5%, 90°和–90°角响应最大;当OSLD平行于射线照射时, 角响应在–1.74%〜 +1.67%,\n 均在± 2%以内。\n 结论 在实际工作中, 应先计算元件校正因子(ECF), 校正各剂量计对射线的灵敏度, 旋转 照射条件下, 将OSLD平行于射线入射方向放置, 能更好地降低角响应影响,\n 便于进一步开展VMAT剂量核查应用研宄。","PeriodicalId":58844,"journal":{"name":"中国辐射卫生","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国辐射卫生","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.13491/J.ISSN.1004-714X.2021.01.009","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective To study the angular dependence of optically stimulated luminescent dosimeter (OSLD)
with solid phantoms under SSDL radiation level 60Co radiation field, and to discuss the possibility of OSLD in volumetric modulated
arc therapy (VMAT) and other rotating irradiation dose audit.
Methods OSLDs were embedded in the two phantoms with the same size and material, respectively.
The phantom 1 was set to make the first OSLD perpendicular to the beam, and the phantom
2 was set to make the second OSLD parallel to the beam. The OSLDs were irradiated
at 8 angles: 0°, 45°, 90°, 135°, 180°, –45°, –90, –135°. The counts of the OSLSs were
read and the response of each angle which normalized to 0° were calculated.
Results When the OSLDs are perpendicular to the beam, the angular response is between –6.76%
〜+1.5%, with the maximum angular dependence at 90° and –90°. When the OSLDs are parallel
to the beam, the angular response is between –1.74%〜+1.67%, below 2%.
Conclusion It is better to correct the sensitivity of dosimeters by Element Correction Factors
(ECF) for dose audit. Under the condition of rotating irradiation, OSLD shoud be set
parallel to the beam, which can better reduce the influence of angular dependence
and facilitate further application research of VMAT dose audit.
摘要: 目的 结合定制固体模体研宄光致发光剂量计(OSLD)在放疗水平 60Co标准辐射场中的角响应性能, 探讨OSLD在容积旋转调强放射治疗(VMAT)等旋转照射剂量核查中应用的可行性。
方法 将OSLD嵌入2个外观尺寸以 及材料均相同的定制固体模体, 模体1使OSLD与射线垂直, 模体2使OSLD与射线平行, 照射8个角度:0°、45°、 90°、135°、180°、–45°、–90°和–135°,
读取每个角度的OSLD计数, 计算每个角度相对于0°的角响应值。
结果 当OSLD 垂直于射线照射时, 角响应在–6.76%〜+1.5%, 90°和–90°角响应最大;当OSLD平行于射线照射时, 角响应在–1.74%〜 +1.67%,
均在± 2%以内。
结论 在实际工作中, 应先计算元件校正因子(ECF), 校正各剂量计对射线的灵敏度, 旋转 照射条件下, 将OSLD平行于射线入射方向放置, 能更好地降低角响应影响,
便于进一步开展VMAT剂量核查应用研宄。
期刊介绍:
Chinese Journal of Radiological Health is one of the Source Journals for Chinese Scientific and Technical Papers and Citations and belongs to the series published by Chinese Preventive Medicine Association (CPMA). It is a national academic journal supervised by National Health Commission of the People’s Republic of China and co-sponsored by Institute of Radiation Medicine, Shandong Academy of Medical Sciences and CPMA, and is a professional academic journal publishing research findings and management experience in the field of radiological health, issued to the public in China and abroad. Under the guidance of the Communist Party of China and the national press and publication policies, the Journal actively publicizes the guidelines and policies of the Party and the state on health work, promotes the implementation of relevant laws, regulations and standards, and timely reports new achievements, new information, new methods and new products in the specialty, with the aim of organizing and promoting the academic communication of radiological health in China and improving the academic level of the specialty, and for the purpose of protecting the health of radiation workers and the public while promoting the extensive use of radioisotopes and radiation devices in the national economy. The main columns include Original Articles, Expert Comments, Experience Exchange, Standards and Guidelines, and Review Articles.