R Park Carmon, Cristine C. Clay, Trudi Stafford, Charles C. Reed
{"title":"A clinical trial of the Bentley BOS-5 pediatric oxygenator.","authors":"R Park Carmon, Cristine C. Clay, Trudi Stafford, Charles C. Reed","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The performance of the Bentley BOS-5 pediatric oxygenator was evaluated on the basis of its response to maintain arterial pH between 7.35 and 7.45, arterial pO(2) between 100 and 200 mm Hg, and arterial pCO(2) between 35 and 45 mm Hg (Texas Heart Institute perfusion protocol). The oxygenator was found to be efficient at all flow rates employed; however, the pO(2) parameter could not be consistently maintained within protocol limits, but could be improved when a mixture of 5% carbon dioxide/95% oxygen was used for the duration of a case.</p>","PeriodicalId":84396,"journal":{"name":"Cardiovascular diseases","volume":"8 4","pages":"562-566"},"PeriodicalIF":0.0000,"publicationDate":"1981-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC288004/pdf/cardiodis00004-0102.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular diseases","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The performance of the Bentley BOS-5 pediatric oxygenator was evaluated on the basis of its response to maintain arterial pH between 7.35 and 7.45, arterial pO(2) between 100 and 200 mm Hg, and arterial pCO(2) between 35 and 45 mm Hg (Texas Heart Institute perfusion protocol). The oxygenator was found to be efficient at all flow rates employed; however, the pO(2) parameter could not be consistently maintained within protocol limits, but could be improved when a mixture of 5% carbon dioxide/95% oxygen was used for the duration of a case.
Bentley BOS-5儿童氧合器的性能评估基于其维持动脉pH在7.35至7.45之间,动脉pO(2)在100至200 mm Hg之间,动脉pCO(2)在35至45 mm Hg之间的反应(德克萨斯心脏研究所灌注方案)。氧合器被发现在所有流速下都是有效的;然而,pO(2)参数不能始终保持在协议限制内,但在病例持续时间内使用5%二氧化碳/95%氧气的混合物可以改善。