{"title":"A simple method to calculate P50 from a single blood sample.","authors":"D J Doyle","doi":"10.1007/BF03356585","DOIUrl":null,"url":null,"abstract":"<p><p>Hill's equation relating oxygen tension, saturation and P50 is used as the basis for a simple method to calculate P50 from a single blood sample. The effects of errors of measurement in oxygen tension and saturation are considered using the technique of sensitivity analysis. The method is illustrated using data published by Severinghaus.</p>","PeriodicalId":77181,"journal":{"name":"International journal of clinical monitoring and computing","volume":"14 2","pages":"109-11"},"PeriodicalIF":0.0000,"publicationDate":"1997-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/BF03356585","citationCount":"8","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of clinical monitoring and computing","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/BF03356585","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8
Abstract
Hill's equation relating oxygen tension, saturation and P50 is used as the basis for a simple method to calculate P50 from a single blood sample. The effects of errors of measurement in oxygen tension and saturation are considered using the technique of sensitivity analysis. The method is illustrated using data published by Severinghaus.