{"title":"Comparison of two anti-hepatitis C virus enzyme-linked immunosorbent assays: Wellcozyme VK45 and Ortho 2.0.","authors":"H Vrielink, C L van der Poel, H W Reesink, P N Lelie","doi":"10.1159/000223117","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The aim of the present study was to compare 2 anti-HCV ELISA tests with respect to sensitivity and specificity in detecting Hepatitis C antigen.</p><p><strong>Materials and methods: </strong>A 3rd-generation anti-HCV ELISA (Wellcozyme anti-HCV VK45) was compared with a 2nd-generation anti-HCV ELISA (Ortho HCV 2.0) in various serum panels: A) anti-HCV ELISA-positive samples of blood donations (n = 536), B) non-A, non-B hepatitis patients (n = 188), C) multi-transfused patients (n = 79), D) hemodialysis patients (n = 473), and E) random blood donors (n = 1,080).</p><p><strong>Results: </strong>Of 248 cDNA polymerase chain reaction (cDNA-PCR) positive samples in panels A, B, and C, ELISA-VK45 detected 247 (99.6%) and Ortho-2 248 (100%). The cDNA-PCR-positive sample missed by ELISA-VK45 showed isolated anti-C33c reactivity in a 2nd-generation recombinant immunoblot (RIBA-2). Of 281 RIBA-2-positive samples, ELISA-VK45 detected 274 (97.5%) and Ortho-2 279 (99.3%). ELISA-VK45-negative, RIBA-2-positive samples showed combined anti-C100/5-1-1 reactivity in RIBA-2 in 6/7 cases and anti-C22 and C33c reactivity in one. Ortho-2-negative, RIBA-2-positive samples showed combined anti-C100/5-1-1 reactivity in RIBA-2 in 2/2 cases. The specificities of ELISA-VK45 and Ortho-2 were not significantly different in 1,080 blood donors.</p><p><strong>Conclusion: </strong>It is concluded that the ELISA-VK45 lacks sensitivity because a cDNA-PCR-confirmed positive sample was missed in the assay. The specificity of both ELISAs was comparable.</p>","PeriodicalId":13632,"journal":{"name":"Infusionstherapie und Transfusionsmedizin","volume":"22 3","pages":"164-7"},"PeriodicalIF":0.0000,"publicationDate":"1995-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1159/000223117","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Infusionstherapie und Transfusionsmedizin","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1159/000223117","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
Background: The aim of the present study was to compare 2 anti-HCV ELISA tests with respect to sensitivity and specificity in detecting Hepatitis C antigen.
Materials and methods: A 3rd-generation anti-HCV ELISA (Wellcozyme anti-HCV VK45) was compared with a 2nd-generation anti-HCV ELISA (Ortho HCV 2.0) in various serum panels: A) anti-HCV ELISA-positive samples of blood donations (n = 536), B) non-A, non-B hepatitis patients (n = 188), C) multi-transfused patients (n = 79), D) hemodialysis patients (n = 473), and E) random blood donors (n = 1,080).
Results: Of 248 cDNA polymerase chain reaction (cDNA-PCR) positive samples in panels A, B, and C, ELISA-VK45 detected 247 (99.6%) and Ortho-2 248 (100%). The cDNA-PCR-positive sample missed by ELISA-VK45 showed isolated anti-C33c reactivity in a 2nd-generation recombinant immunoblot (RIBA-2). Of 281 RIBA-2-positive samples, ELISA-VK45 detected 274 (97.5%) and Ortho-2 279 (99.3%). ELISA-VK45-negative, RIBA-2-positive samples showed combined anti-C100/5-1-1 reactivity in RIBA-2 in 6/7 cases and anti-C22 and C33c reactivity in one. Ortho-2-negative, RIBA-2-positive samples showed combined anti-C100/5-1-1 reactivity in RIBA-2 in 2/2 cases. The specificities of ELISA-VK45 and Ortho-2 were not significantly different in 1,080 blood donors.
Conclusion: It is concluded that the ELISA-VK45 lacks sensitivity because a cDNA-PCR-confirmed positive sample was missed in the assay. The specificity of both ELISAs was comparable.