{"title":"Dose reduction in the low-energy X-ray diagnostics of extremities.","authors":"P Zaránd, R Pálvölgyi, Z Péntek, I Polgár","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The authors determined the surface absorbed dose (Ds = 67-68 mGy) and the average absorbed dose (Dav = 13-15 mGy) by means of non-screen film radiography of the extremities. The authors recommend the use of xeroradiography (with a tungsten anode tube) or, first of all, a rare-earth screen - film combination (e.g. Kodak min-R + Nuklearmedizin NMB film). The two methods permit the dose values to be decreased by a factor of about 4 and 15, respectively. The dose values are smaller, at least for the 8 cm thick tissue equivalent phantom, when a tungsten anode is used instead of a Mo anode.</p>","PeriodicalId":7308,"journal":{"name":"Acta biochimica et biophysica; Academiae Scientiarum Hungaricae","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1981-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta biochimica et biophysica; Academiae Scientiarum Hungaricae","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The authors determined the surface absorbed dose (Ds = 67-68 mGy) and the average absorbed dose (Dav = 13-15 mGy) by means of non-screen film radiography of the extremities. The authors recommend the use of xeroradiography (with a tungsten anode tube) or, first of all, a rare-earth screen - film combination (e.g. Kodak min-R + Nuklearmedizin NMB film). The two methods permit the dose values to be decreased by a factor of about 4 and 15, respectively. The dose values are smaller, at least for the 8 cm thick tissue equivalent phantom, when a tungsten anode is used instead of a Mo anode.