Technical features and analytical performances of the Stago Expert Preanalytical Check module for hemolysis, icterus, and lipemia measurement.

IF 1.9 4区 医学 Q2 PATHOLOGY
Pascale Laroche, Joffrey Feriel, Geneviève Tanala Marolahy, Odile Chachlica, François Depasse
{"title":"Technical features and analytical performances of the Stago Expert Preanalytical Check module for hemolysis, icterus, and lipemia measurement.","authors":"Pascale Laroche, Joffrey Feriel, Geneviève Tanala Marolahy, Odile Chachlica, François Depasse","doi":"10.1093/ajcp/aqaf086","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Interference caused by hemolysis, icterus, and lipemia (HIL) are a common source of preanalytical errors during coagulation testing. Stago has developed a specific module on its coagulation analyzers to standardize the detection of HIL.</p><p><strong>Methods: </strong>The aims of this study were to evaluate the accuracy, carryover, and cross-interference of HIL measurement and compare results with those obtained on cobas 8000 modular analyzer series devices (Roche Diagnostics). These evaluations were carried out using Clinical and Laboratory Standards Institute guidance and regulatory study requirements.</p><p><strong>Results: </strong>We demonstrated the absence of sample carryover through the HIL measurement chamber and excellent accuracy between HIL theoretical and measured concentrations, with an R2 above 0.99 in all cases. We did not detect any cross-interference of bilirubin and lipids to hemoglobin, of hemoglobin and lipids to bilirubin, or for hemoglobin and bilirubin to lipids. We found excellent agreement between HIL on the Stago STA R Max3 and cobas 8000 platforms, with κ coefficients ranging from 0.928 to 0.963.</p><p><strong>Conclusions: </strong>These results support the implementation of the Stago Expert Preanalytical Check module for HIL interference measurement in routine coagulation laboratory testing.</p>","PeriodicalId":7506,"journal":{"name":"American journal of clinical pathology","volume":" ","pages":"524-529"},"PeriodicalIF":1.9000,"publicationDate":"2025-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12495518/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of clinical pathology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/ajcp/aqaf086","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"PATHOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Interference caused by hemolysis, icterus, and lipemia (HIL) are a common source of preanalytical errors during coagulation testing. Stago has developed a specific module on its coagulation analyzers to standardize the detection of HIL.

Methods: The aims of this study were to evaluate the accuracy, carryover, and cross-interference of HIL measurement and compare results with those obtained on cobas 8000 modular analyzer series devices (Roche Diagnostics). These evaluations were carried out using Clinical and Laboratory Standards Institute guidance and regulatory study requirements.

Results: We demonstrated the absence of sample carryover through the HIL measurement chamber and excellent accuracy between HIL theoretical and measured concentrations, with an R2 above 0.99 in all cases. We did not detect any cross-interference of bilirubin and lipids to hemoglobin, of hemoglobin and lipids to bilirubin, or for hemoglobin and bilirubin to lipids. We found excellent agreement between HIL on the Stago STA R Max3 and cobas 8000 platforms, with κ coefficients ranging from 0.928 to 0.963.

Conclusions: These results support the implementation of the Stago Expert Preanalytical Check module for HIL interference measurement in routine coagulation laboratory testing.

技术特点和分析性能的Stago专家分析前检查模块溶血,黄疸,和血脂测量。
目的:溶血、黄疸和血脂(HIL)引起的干扰是凝血试验中分析前误差的常见来源。Stago在其凝血分析仪上开发了一个特定的模块,以标准化HIL的检测。方法:本研究的目的是评估HIL测量的准确性、结转性和交叉干扰,并将结果与cobas 8000模块化分析仪系列设备(罗氏诊断)的结果进行比较。这些评估是根据临床和实验室标准协会的指导和监管研究要求进行的。结果:我们证明了HIL测量室没有样品携带,HIL理论浓度和测量浓度之间的准确度很高,在所有情况下R2都在0.99以上。我们没有发现胆红素和脂质对血红蛋白的交叉干扰,血红蛋白和脂质对胆红素的交叉干扰,或者血红蛋白和胆红素对脂质的交叉干扰。我们发现在Stago STA R Max3和cobas 8000平台上的HIL之间具有良好的一致性,κ系数范围为0.928至0.963。结论:这些结果支持在常规凝血实验室检测中实施Stago专家预分析检查模块用于HIL干扰测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
7.70
自引率
2.90%
发文量
367
审稿时长
3-6 weeks
期刊介绍: The American Journal of Clinical Pathology (AJCP) is the official journal of the American Society for Clinical Pathology and the Academy of Clinical Laboratory Physicians and Scientists. It is a leading international journal for publication of articles concerning novel anatomic pathology and laboratory medicine observations on human disease. AJCP emphasizes articles that focus on the application of evolving technologies for the diagnosis and characterization of diseases and conditions, as well as those that have a direct link toward improving patient care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信