Terminal voltage calibration of a 3 MV Tandetron accelerator at Sichuan University, China

IF 1.4 3区 物理与天体物理 Q3 INSTRUMENTS & INSTRUMENTATION
D.H. Xie , Y. Tang , T.Y. Jiao , Z. An , F. Bai , X. Fang , Y.H. Fan , Y.X. Fan , B.S. Gao , J.F. Han , B. Li , G. Liu , W.P. Lin , J.H. Tan , X.D. Tang , C. Wen , P. Wang , Q. Wang
{"title":"Terminal voltage calibration of a 3 MV Tandetron accelerator at Sichuan University, China","authors":"D.H. Xie ,&nbsp;Y. Tang ,&nbsp;T.Y. Jiao ,&nbsp;Z. An ,&nbsp;F. Bai ,&nbsp;X. Fang ,&nbsp;Y.H. Fan ,&nbsp;Y.X. Fan ,&nbsp;B.S. Gao ,&nbsp;J.F. Han ,&nbsp;B. Li ,&nbsp;G. Liu ,&nbsp;W.P. Lin ,&nbsp;J.H. Tan ,&nbsp;X.D. Tang ,&nbsp;C. Wen ,&nbsp;P. Wang ,&nbsp;Q. Wang","doi":"10.1016/j.nimb.2025.165829","DOIUrl":null,"url":null,"abstract":"<div><div>The 3 MV Tandetron accelerator at Sichuan University has been calibrated across a wider terminal voltage range (500–2100 kV), employing the narrow resonance of the <sup>13</sup>C(<span><math><mrow><mi>α</mi><mo>,</mo><mi>n</mi></mrow></math></span>)<sup>16</sup>O reaction along with neutron threshold measurements from <sup>7</sup>Li(<span><math><mrow><mi>p</mi><mo>,</mo><mi>n</mi></mrow></math></span>)<sup>7</sup>Be, <sup>13</sup>C(<span><math><mrow><mi>p</mi><mo>,</mo><mi>n</mi></mrow></math></span>)<sup>13</sup>N, and <sup>19</sup>F(<span><math><mrow><mi>p</mi><mo>,</mo><mi>n</mi></mrow></math></span>)<sup>19</sup>Ne reactions. The calibration demonstrates excellent reproducibility of energy measurements over a two-year period, with about 0.12 keV variation in repeated measurements at <span><math><mrow><msub><mrow><mi>E</mi></mrow><mrow><mi>α</mi></mrow></msub><mo>=</mo><mn>1055</mn><mo>.</mo><mn>63</mn></mrow></math></span> keV. However, an approximate 5 keV shift observed in the <sup>7</sup>Li(<span><math><mrow><mi>p</mi><mo>,</mo><mi>n</mi></mrow></math></span>)<sup>7</sup>Be threshold energy over seven years demonstrates the necessity for periodic recalibration. Given the sub-keV reproducibility demonstrated over a two-year period, we recommend recalibrating at least every two years to ensure ongoing energy accuracy for this accelerator. The calibration parameters are a = 0.9918 ± 0.0005 and b = 4.3104 ± 0.5319 kV, where V = <span><math><mrow><mi>a</mi><mi>⋅</mi><msub><mrow><mi>V</mi></mrow><mrow><mtext>set</mtext></mrow></msub></mrow></math></span> + b. This work establishes a robust calibration framework that ensures reliable accelerator performance for nuclear physics researches and applications.</div></div>","PeriodicalId":19380,"journal":{"name":"Nuclear Instruments & Methods in Physics Research Section B-beam Interactions With Materials and Atoms","volume":"568 ","pages":"Article 165829"},"PeriodicalIF":1.4000,"publicationDate":"2025-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear Instruments & Methods in Physics Research Section B-beam Interactions With Materials and Atoms","FirstCategoryId":"101","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0168583X25002198","RegionNum":3,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

The 3 MV Tandetron accelerator at Sichuan University has been calibrated across a wider terminal voltage range (500–2100 kV), employing the narrow resonance of the 13C(α,n)16O reaction along with neutron threshold measurements from 7Li(p,n)7Be, 13C(p,n)13N, and 19F(p,n)19Ne reactions. The calibration demonstrates excellent reproducibility of energy measurements over a two-year period, with about 0.12 keV variation in repeated measurements at Eα=1055.63 keV. However, an approximate 5 keV shift observed in the 7Li(p,n)7Be threshold energy over seven years demonstrates the necessity for periodic recalibration. Given the sub-keV reproducibility demonstrated over a two-year period, we recommend recalibrating at least every two years to ensure ongoing energy accuracy for this accelerator. The calibration parameters are a = 0.9918 ± 0.0005 and b = 4.3104 ± 0.5319 kV, where V = aVset + b. This work establishes a robust calibration framework that ensures reliable accelerator performance for nuclear physics researches and applications.
四川大学3mv tandem加速器的终端电压校准
利用13C(α,n)16O反应的窄共振以及7Li(p,n)7Be、13C(p,n)13N和19F(p,n)19Ne反应的中子阈值测量,对四川大学的3mv Tandetron加速器在更宽的终端电压范围内(500-2100 kV)进行了校准。校准表明,在两年的时间内,能量测量具有良好的再现性,在Eα=1055.63 keV时,重复测量的变化约为0.12 keV。然而,在7Li(p,n)7Be阈值能量中观察到的近5 keV的位移在7年内表明了周期性重新校准的必要性。鉴于在两年的时间内证明了亚密钥的可重复性,我们建议至少每两年重新校准一次,以确保该加速器的持续能量准确性。标定参数分别为a = 0.9918±0.0005和b = 4.3104±0.5319 kV,其中V = a⋅Vset + b,为核物理研究和应用提供了可靠的加速器标定框架。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.80
自引率
7.70%
发文量
231
审稿时长
1.9 months
期刊介绍: Section B of Nuclear Instruments and Methods in Physics Research covers all aspects of the interaction of energetic beams with atoms, molecules and aggregate forms of matter. This includes ion beam analysis and ion beam modification of materials as well as basic data of importance for these studies. Topics of general interest include: atomic collisions in solids, particle channelling, all aspects of collision cascades, the modification of materials by energetic beams, ion implantation, irradiation - induced changes in materials, the physics and chemistry of beam interactions and the analysis of materials by all forms of energetic radiation. Modification by ion, laser and electron beams for the study of electronic materials, metals, ceramics, insulators, polymers and other important and new materials systems are included. Related studies, such as the application of ion beam analysis to biological, archaeological and geological samples as well as applications to solve problems in planetary science are also welcome. Energetic beams of interest include atomic and molecular ions, neutrons, positrons and muons, plasmas directed at surfaces, electron and photon beams, including laser treated surfaces and studies of solids by photon radiation from rotating anodes, synchrotrons, etc. In addition, the interaction between various forms of radiation and radiation-induced deposition processes are relevant.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信