[Feasibility study on measuring anteversion angle of acetabular prosthesis after total hip arthroplasty using arbitrary point method].

Q3 Medicine
Bowen Li, Longyuan Li, Heng Zhang
{"title":"[Feasibility study on measuring anteversion angle of acetabular prosthesis after total hip arthroplasty using arbitrary point method].","authors":"Bowen Li, Longyuan Li, Heng Zhang","doi":"10.7507/1002-1892.202501009","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To explore the reliability and accuracy of the arbitrary point method for measuring the anteversion angle of acetabular prosthesis after total hip arthroplasty (THA) based on pelvic X-ray films.</p><p><strong>Methods: </strong>The clinical data of 23 patients (25 hips) who underwent THA between December 2018 and September 2023 and met the selection criteria were retrospectively analyzed. Among them, there were 16 males and 7 females, with an average age of 57.6 years (range, 34-81 years); 13 hips had THA on the left side and 12 on the right side. There were 19 cases (21 hips) of osteonecrosis of the femoral head, 2 cases (2 hips) of femoral neck fractures, 1 case (1 hip) of developmental dysplasia of the hip, and 1 case (1 hip) of osteoarthritis. After THA, all patients underwent X-ray examination and CT scan. Three physicians measured the anteversion angle of acetabular prosthesis using the arbitrary point method and the CT measurement method respectively, and repeated the measurements three times. The results of the two measurement methods were compared, and the intraclass correlation coefficient (ICC) was employed to assess the reproducibility of the methods.</p><p><strong>Results: </strong>The anteversion angles of acetabular prosthesis were (15.87±7.73)° measured by the arbitrary point method, and (15.31±7.89)° measured by CT measurement method. There was no significant difference between the two methods ( <i>t</i>=1.515, <i>P</i>=0.143). The ICC of the measurement results by the arbitrary point method for the three physicians were 0.97 ( <i>P</i><0.001), 0.96 ( <i>P</i><0.001), and 0.96 ( <i>P</i><0.001), respectively; and the ICC of the measurement results by CT method were 0.93 ( <i>P</i><0.001), 0.93 ( <i>P</i><0.001), and 0.94 ( <i>P</i><0.001), respectively.</p><p><strong>Conclusion: </strong>The arbitrary point method for measuring the anteversion angle of acetabular prosthesis after THA based on pelvic X-ray film is easy to operate, accurate, and has high reproducibility.</p>","PeriodicalId":23979,"journal":{"name":"中国修复重建外科杂志","volume":"39 4","pages":"420-424"},"PeriodicalIF":0.0000,"publicationDate":"2025-04-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12011510/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国修复重建外科杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.7507/1002-1892.202501009","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To explore the reliability and accuracy of the arbitrary point method for measuring the anteversion angle of acetabular prosthesis after total hip arthroplasty (THA) based on pelvic X-ray films.

Methods: The clinical data of 23 patients (25 hips) who underwent THA between December 2018 and September 2023 and met the selection criteria were retrospectively analyzed. Among them, there were 16 males and 7 females, with an average age of 57.6 years (range, 34-81 years); 13 hips had THA on the left side and 12 on the right side. There were 19 cases (21 hips) of osteonecrosis of the femoral head, 2 cases (2 hips) of femoral neck fractures, 1 case (1 hip) of developmental dysplasia of the hip, and 1 case (1 hip) of osteoarthritis. After THA, all patients underwent X-ray examination and CT scan. Three physicians measured the anteversion angle of acetabular prosthesis using the arbitrary point method and the CT measurement method respectively, and repeated the measurements three times. The results of the two measurement methods were compared, and the intraclass correlation coefficient (ICC) was employed to assess the reproducibility of the methods.

Results: The anteversion angles of acetabular prosthesis were (15.87±7.73)° measured by the arbitrary point method, and (15.31±7.89)° measured by CT measurement method. There was no significant difference between the two methods ( t=1.515, P=0.143). The ICC of the measurement results by the arbitrary point method for the three physicians were 0.97 ( P<0.001), 0.96 ( P<0.001), and 0.96 ( P<0.001), respectively; and the ICC of the measurement results by CT method were 0.93 ( P<0.001), 0.93 ( P<0.001), and 0.94 ( P<0.001), respectively.

Conclusion: The arbitrary point method for measuring the anteversion angle of acetabular prosthesis after THA based on pelvic X-ray film is easy to operate, accurate, and has high reproducibility.

[任意点法测量全髋关节置换术后髋臼假体前倾角的可行性研究]。
目的:探讨基于骨盆x线片的任意点法测量全髋关节置换术后髋臼假体前倾角的可靠性和准确性。方法:回顾性分析2018年12月至2023年9月期间符合选择标准的23例(25髋)THA患者的临床资料。其中男性16例,女性7例,平均年龄57.6岁(范围34 ~ 81岁);13髋左侧有THA, 12髋右侧有THA。股骨头坏死19例(21髋),股骨颈骨折2例(2髋),髋关节发育不良1例(1髋),骨关节炎1例(1髋)。术后所有患者均行x线及CT检查。3名医生分别采用任意点法和CT测量法测量髋臼假体前倾角,并重复测量3次。比较两种测量方法的结果,并采用类内相关系数(ICC)评价方法的可重复性。结果:任意点法测髋臼假体前倾角为(15.87±7.73)°,CT测量法测髋臼前倾角为(15.31±7.89)°。两种方法比较差异无统计学意义(t=1.515, P=0.143)。3位医师任意点法测量结果的ICC值均为0.97 (ppppppp)。结论:基于骨盆x线片的任意点法测量THA后髋臼假体前倾角操作简便、准确,重现性高。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
中国修复重建外科杂志
中国修复重建外科杂志 Medicine-Medicine (all)
CiteScore
0.80
自引率
0.00%
发文量
11334
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信