{"title":"Reproducibility and reliability of intraocular pressure self-measurement with iCare® HOME2 compared to Goldmann applanation tonometry","authors":"J.-B. Amatu , P. Bastelica , J. Buffault , C. Baudouin , A. Labbé","doi":"10.1016/j.jfo.2025.104517","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>To evaluate the reproducibility and reliability of intraocular pressure (IOP) self-measurements with the iCare® HOME2 tonometer (ICH2, model TA023, Icare Oy, Vanda, Finland) in patients with ocular hypertension or glaucoma.</div></div><div><h3>Patients and methods</h3><div>Fifty-two eyes of 29 patients were included. IOP was assessed consecutively by: air pulse tonometer (AT), Goldmann applanation tonometer (GAT), and ICH2 tonometer by the patient (ICH-P), then by an ophthalmologist (ICH-O). Reproducibility and agreement were assessed using a Bland-Altman test, and reliability was evaluated using an intraclass correlation coefficient (ICC).</div></div><div><h3>Results</h3><div>Mean IOP by ICH-P, ICH-O, AT, and GAT was, respectively, 18.02<!--> <!-->±<!--> <!-->6.53<!--> <!-->mmHg, 17.08<!--> <!-->±<!--> <!-->5.65<!--> <!-->mmHg, 15.60<!--> <!-->±<!--> <!-->4.48<!--> <!-->mmHg and 14.42<!--> <!-->±<!--> <!-->5.00<!--> <!-->mmHg. Bland-Altman analysis between ICH-P and GAT and between ICH-O and GAT showed good reliability, finding a mean bias of 3.59<!--> <!-->mmHg and 3.09<!--> <!-->mmHg respectively, with a 95% limit of agreement of –1.48 to 8.68<!--> <!-->mmHg and –1.35 to 7.53<!--> <!-->mmHg, respectively. The repeatability of the measurements was excellent: ICC's of the measurements on each eye by the AT, ICH-O, ICH-P and GAT were 0.979, 0.991, 0.989 and 0.992, respectively.</div></div><div><h3>Conclusion</h3><div>Although ICH2 appears to overestimate IOP compared with GAT, there was good correlation between ICH2 self-measurements and GAT. iCare HOME2 appears to have good repeatability and reproducibility, and therefore may be a valuable tool to assess IOP in routine glaucoma monitoring.</div></div><div><h3>Objectif</h3><div>Évaluer la reproductibilité et la fiabilité des auto-mesures de la pression intraoculaire (PIO) avec le tonomètre Icare Home 2 (ICH2, modèle TA023, Icare Oy, Vanda, Finlande) chez des patients souffrant d’hypertension oculaire ou de glaucome.</div></div><div><h3>Patients et méthodes</h3><div>Cinquante-deux yeux de 29 patients ont été inclus. La PIO a été évaluée consécutivement par : tonomètre à air pulsé (AT), tonomètre à applanation de Goldmann (GAT), et tonomètre ICH2 par le patient (ICH-P), puis par un ophtalmologiste (ICH-O). La reproductibilité et la concordance des mesures ont été évaluées à l’aide d’un test de Bland-Altman, et la fiabilité a été évaluée à l’aide d’un coefficient de corrélation intra-classe (ICC).</div></div><div><h3>Résultats</h3><div>La PIO moyenne selon l’ICH-P, l’ICH-O, l’AT et le GAT était respectivement de 18,02<!--> <!-->±<!--> <!-->6,53<!--> <!-->mmHg, 17,08<!--> <!-->±<!--> <!-->5,65<!--> <!-->mmHg, 15,60<!--> <!-->±<!--> <!-->4,48<!--> <!-->mmHg et 14,42<!--> <!-->±<!--> <!-->5,00<!--> <!-->mmHg. L’analyse de Bland-Altman entre ICH-P et GAT et entre ICH-O et GAT a montré une bonne fiabilité, avec un biais moyen de 3,59<!--> <!-->mmHg et 3,09<!--> <!-->mmHg respectivement, avec une limite d’accord à 95 % de –1,48 à 8,68<!--> <!-->mmHg et de –1,35 à 7,53<!--> <!-->mmHg. La répétabilité des mesures était excellente : l’ICC des mesures effectuées sur chaque œil par l’AT, l’ICH-O, l’ICH-P et le GAT était de 0,979, 0,991, 0,989 et 0,992, respectivement.</div></div><div><h3>Conclusion</h3><div>Bien que l’ICH2 semble surestimer la PIO par rapport au GAT, il existe une bonne corrélation entre les auto-mesures de l’ICH2 et le GAT. IcareHome2 semble avoir une bonne répétabilité et reproductibilité, et peut donc être un outil précieux pour évaluer la PIO dans la surveillance du glaucome.</div></div>","PeriodicalId":14777,"journal":{"name":"Journal Francais D Ophtalmologie","volume":"48 6","pages":"Article 104517"},"PeriodicalIF":1.2000,"publicationDate":"2025-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal Francais D Ophtalmologie","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0181551225000993","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
To evaluate the reproducibility and reliability of intraocular pressure (IOP) self-measurements with the iCare® HOME2 tonometer (ICH2, model TA023, Icare Oy, Vanda, Finland) in patients with ocular hypertension or glaucoma.
Patients and methods
Fifty-two eyes of 29 patients were included. IOP was assessed consecutively by: air pulse tonometer (AT), Goldmann applanation tonometer (GAT), and ICH2 tonometer by the patient (ICH-P), then by an ophthalmologist (ICH-O). Reproducibility and agreement were assessed using a Bland-Altman test, and reliability was evaluated using an intraclass correlation coefficient (ICC).
Results
Mean IOP by ICH-P, ICH-O, AT, and GAT was, respectively, 18.02 ± 6.53 mmHg, 17.08 ± 5.65 mmHg, 15.60 ± 4.48 mmHg and 14.42 ± 5.00 mmHg. Bland-Altman analysis between ICH-P and GAT and between ICH-O and GAT showed good reliability, finding a mean bias of 3.59 mmHg and 3.09 mmHg respectively, with a 95% limit of agreement of –1.48 to 8.68 mmHg and –1.35 to 7.53 mmHg, respectively. The repeatability of the measurements was excellent: ICC's of the measurements on each eye by the AT, ICH-O, ICH-P and GAT were 0.979, 0.991, 0.989 and 0.992, respectively.
Conclusion
Although ICH2 appears to overestimate IOP compared with GAT, there was good correlation between ICH2 self-measurements and GAT. iCare HOME2 appears to have good repeatability and reproducibility, and therefore may be a valuable tool to assess IOP in routine glaucoma monitoring.
Objectif
Évaluer la reproductibilité et la fiabilité des auto-mesures de la pression intraoculaire (PIO) avec le tonomètre Icare Home 2 (ICH2, modèle TA023, Icare Oy, Vanda, Finlande) chez des patients souffrant d’hypertension oculaire ou de glaucome.
Patients et méthodes
Cinquante-deux yeux de 29 patients ont été inclus. La PIO a été évaluée consécutivement par : tonomètre à air pulsé (AT), tonomètre à applanation de Goldmann (GAT), et tonomètre ICH2 par le patient (ICH-P), puis par un ophtalmologiste (ICH-O). La reproductibilité et la concordance des mesures ont été évaluées à l’aide d’un test de Bland-Altman, et la fiabilité a été évaluée à l’aide d’un coefficient de corrélation intra-classe (ICC).
Résultats
La PIO moyenne selon l’ICH-P, l’ICH-O, l’AT et le GAT était respectivement de 18,02 ± 6,53 mmHg, 17,08 ± 5,65 mmHg, 15,60 ± 4,48 mmHg et 14,42 ± 5,00 mmHg. L’analyse de Bland-Altman entre ICH-P et GAT et entre ICH-O et GAT a montré une bonne fiabilité, avec un biais moyen de 3,59 mmHg et 3,09 mmHg respectivement, avec une limite d’accord à 95 % de –1,48 à 8,68 mmHg et de –1,35 à 7,53 mmHg. La répétabilité des mesures était excellente : l’ICC des mesures effectuées sur chaque œil par l’AT, l’ICH-O, l’ICH-P et le GAT était de 0,979, 0,991, 0,989 et 0,992, respectivement.
Conclusion
Bien que l’ICH2 semble surestimer la PIO par rapport au GAT, il existe une bonne corrélation entre les auto-mesures de l’ICH2 et le GAT. IcareHome2 semble avoir une bonne répétabilité et reproductibilité, et peut donc être un outil précieux pour évaluer la PIO dans la surveillance du glaucome.
期刊介绍:
The Journal français d''ophtalmologie, official publication of the French Society of Ophthalmology, serves the French Speaking Community by publishing excellent research articles, communications of the French Society of Ophthalmology, in-depth reviews, position papers, letters received by the editor and a rich image bank in each issue. The scientific quality is guaranteed through unbiased peer-review, and the journal is member of the Committee of Publication Ethics (COPE). The editors strongly discourage editorial misconduct and in particular if duplicative text from published sources is identified without proper citation, the submission will not be considered for peer review and returned to the authors or immediately rejected.