An In Vitro Study of Torque Application in Abutment Screw Tightening at the Implant-Abutment Junction in Simulated Osseointegrated Implants.

Jan C Kwan
{"title":"An In Vitro Study of Torque Application in Abutment Screw Tightening at the Implant-Abutment Junction in Simulated Osseointegrated Implants.","authors":"Jan C Kwan","doi":"10.11607/jomi.11322","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate initial and removal torque values (ITV and RTV) in abutment screw tightening and their impact on torque application at the implant-abutment junction (IAJ), utilizing a biomechanical friction resistance model simulating various bone densities, with and without counter-torquing measures.</p><p><strong>Materials and method: </strong>Sixty implant-abutment assemblies were divided into three groups, each simulating different levels of frictional resistance. The endpoint for abutment screw tightening were either reaching a preset torque value of 35 Ncm or detecting a change in the implant's angular position. ITV, RTV, and changes in angular position were recorded. One-way ANOVA with Tukey post hoc were used.</p><p><strong>Results: </strong>For all frictional resistance groups, abutment screws tightened with a counter-torquing device attained 35 Ncm (p> 0.05). Abutment screws tightened without a counter torquing device in low and medium frictional resistances were unable to attain 35 Ncm prior to implant rotation. ITVs of abutment screws without counter-torque were not significantly different in the medium and high frictional resistance groups (p > 0.05), except in the low frictional resistance group (p < 0.001). No significant difference in the change in implant angular displacement for all groups (p > 0.05) regardless of counter-torquing measures; however, a high variance was observed in the low and medium frictional resistance groups.</p><p><strong>Conclusion: </strong>Under the experimental conditions, counter-torque is important to ensure the desired torque value is applied to the implant-abutment connection in low frictional resistance. RTV is proportionally related to ITV and can be used to estimate the ITV at the IAJ.</p>","PeriodicalId":94230,"journal":{"name":"The International journal of oral & maxillofacial implants","volume":"0 0","pages":"1-27"},"PeriodicalIF":0.0000,"publicationDate":"2025-03-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The International journal of oral & maxillofacial implants","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11607/jomi.11322","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate initial and removal torque values (ITV and RTV) in abutment screw tightening and their impact on torque application at the implant-abutment junction (IAJ), utilizing a biomechanical friction resistance model simulating various bone densities, with and without counter-torquing measures.

Materials and method: Sixty implant-abutment assemblies were divided into three groups, each simulating different levels of frictional resistance. The endpoint for abutment screw tightening were either reaching a preset torque value of 35 Ncm or detecting a change in the implant's angular position. ITV, RTV, and changes in angular position were recorded. One-way ANOVA with Tukey post hoc were used.

Results: For all frictional resistance groups, abutment screws tightened with a counter-torquing device attained 35 Ncm (p> 0.05). Abutment screws tightened without a counter torquing device in low and medium frictional resistances were unable to attain 35 Ncm prior to implant rotation. ITVs of abutment screws without counter-torque were not significantly different in the medium and high frictional resistance groups (p > 0.05), except in the low frictional resistance group (p < 0.001). No significant difference in the change in implant angular displacement for all groups (p > 0.05) regardless of counter-torquing measures; however, a high variance was observed in the low and medium frictional resistance groups.

Conclusion: Under the experimental conditions, counter-torque is important to ensure the desired torque value is applied to the implant-abutment connection in low frictional resistance. RTV is proportionally related to ITV and can be used to estimate the ITV at the IAJ.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信