Comprehensive analysis of peripheral dose in electron beam therapy with a Varian TrueBeam Linac

Q1 Health Professions
Kapil Dev Maharaj , Mahsheed Sabet , Joshua Dass , Mounir Ibrahim , Talat Mahmood , Pejman Rowshanfarzad
{"title":"Comprehensive analysis of peripheral dose in electron beam therapy with a Varian TrueBeam Linac","authors":"Kapil Dev Maharaj ,&nbsp;Mahsheed Sabet ,&nbsp;Joshua Dass ,&nbsp;Mounir Ibrahim ,&nbsp;Talat Mahmood ,&nbsp;Pejman Rowshanfarzad","doi":"10.1016/j.radmp.2024.10.003","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To investigate the characteristics of peripheral doses outside electron-beam applicators in Varian TrueBeam linacs.</div></div><div><h3>Method</h3><div>Peripheral doses outside the electron applicator were measured for 6-, 9- and 12-MeV beams at the maximum dose depth (<em>D</em><sub>max</sub>) for each energy source and at a source-to-surface distance (SSD) of 100 cm. Measurements were performed using EBT3 films in solid water phantoms. The impact of field size on the penumbra width and peripheral doses was studied using various cutouts, including 3 ​cm ​× ​3 ​cm, 6 ​cm ​× ​6 ​cm, and 10 ​cm ​× ​10 ​cm in a 10 ​cm ​× ​10 ​cm applicator with the gantry and collimator at 0°. The influence of the applicator size was investigated using a circular cutout of 5 ​cm in diameter for various applicator sizes, including 6 ​cm ​× ​6 ​cm, 10 ​cm ​× ​10 ​cm, 15 ​cm ​× ​15 ​cm, 20 cm ​× ​20 ​cm, and 25 ​cm ​× ​25 ​cm, at <em>D</em><sub>max</sub> for each energy, while keeping the gantry and collimator angle at 0°. The measured dose profiles were compared with the Eclipse treatment planning system (TPS) predicted dose profiles. The effect of varying gantry angles (0°, 90°, and 270°) for a 3 ​cm ​× ​3 ​cm cutout in a 10 ​cm ​× ​10 ​cm applicator for each energy source and varying collimator angles (0°, 90°, and 270°) for a 10 ​cm ​× ​10 ​cm field were investigated to determine their effects on the penumbra widths and peripheral doses.</div></div><div><h3>Results</h3><div>Both the penumbra width and peripheral dose values increased with energy across different field sizes, gantry angles, collimator angles, and applicator sizes. Root mean square deviation (RMSD) analysis indicated minimal differences between the measured profiles and TPS data. Peripheral doses remained below 5% of the maximum dose approximately 10–15 ​mm away from the field edges, suggesting the potential for implementing additional shielding where required.</div></div><div><h3>Conclusions</h3><div>This study highlights the importance of considering peripheral doses in electron radiotherapy. It is important to note the impact on healthy tissues beyond the treatment area to ensure patient safety and prevent the long-term side effects of treatment. These findings emphasize the necessity of implementing appropriate measures to minimize peripheral doses.</div></div>","PeriodicalId":34051,"journal":{"name":"Radiation Medicine and Protection","volume":"6 1","pages":"Pages 28-36"},"PeriodicalIF":0.0000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation Medicine and Protection","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666555724000844","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

To investigate the characteristics of peripheral doses outside electron-beam applicators in Varian TrueBeam linacs.

Method

Peripheral doses outside the electron applicator were measured for 6-, 9- and 12-MeV beams at the maximum dose depth (Dmax) for each energy source and at a source-to-surface distance (SSD) of 100 cm. Measurements were performed using EBT3 films in solid water phantoms. The impact of field size on the penumbra width and peripheral doses was studied using various cutouts, including 3 ​cm ​× ​3 ​cm, 6 ​cm ​× ​6 ​cm, and 10 ​cm ​× ​10 ​cm in a 10 ​cm ​× ​10 ​cm applicator with the gantry and collimator at 0°. The influence of the applicator size was investigated using a circular cutout of 5 ​cm in diameter for various applicator sizes, including 6 ​cm ​× ​6 ​cm, 10 ​cm ​× ​10 ​cm, 15 ​cm ​× ​15 ​cm, 20 cm ​× ​20 ​cm, and 25 ​cm ​× ​25 ​cm, at Dmax for each energy, while keeping the gantry and collimator angle at 0°. The measured dose profiles were compared with the Eclipse treatment planning system (TPS) predicted dose profiles. The effect of varying gantry angles (0°, 90°, and 270°) for a 3 ​cm ​× ​3 ​cm cutout in a 10 ​cm ​× ​10 ​cm applicator for each energy source and varying collimator angles (0°, 90°, and 270°) for a 10 ​cm ​× ​10 ​cm field were investigated to determine their effects on the penumbra widths and peripheral doses.

Results

Both the penumbra width and peripheral dose values increased with energy across different field sizes, gantry angles, collimator angles, and applicator sizes. Root mean square deviation (RMSD) analysis indicated minimal differences between the measured profiles and TPS data. Peripheral doses remained below 5% of the maximum dose approximately 10–15 ​mm away from the field edges, suggesting the potential for implementing additional shielding where required.

Conclusions

This study highlights the importance of considering peripheral doses in electron radiotherapy. It is important to note the impact on healthy tissues beyond the treatment area to ensure patient safety and prevent the long-term side effects of treatment. These findings emphasize the necessity of implementing appropriate measures to minimize peripheral doses.
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiation Medicine and Protection
Radiation Medicine and Protection Health Professions-Emergency Medical Services
CiteScore
2.10
自引率
0.00%
发文量
0
审稿时长
103 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信