{"title":"Complete blood count in neonatal Intensive care Unit (NICU): Performance comparison between POCT Sight OLO® and Sysmex XN-9100™ hematology analyzers","authors":"Francesca Nencini , Alessandro Bonari , Sara Ciullini Mannurita , Alessandra Mongia , Francesca Romano , Maria Garieri , Edda Russo , Silvia Sastrucci , Graziella Marrani , Martina Tonelli , Stefano Salti , Niccola Funel , Amedeo Amedei , Carlo Dani , Alessandra Fanelli","doi":"10.1016/j.plabm.2025.e00453","DOIUrl":null,"url":null,"abstract":"<div><h3>Background and aims</h3><div>The use of a POCT (Point Of Care Test) could help in reducing the impact of pre-analytical errors in particular in challenging newborn samples.</div><div>The study purpose is to compare the POCT Sight OLO® hematology analyzer, validated for >3 months patients, with the reference system Sysmex XN-9100™ in Neonatal Intensive Care Unit (NICU).</div></div><div><h3>Material and methods</h3><div>The two analyzers were compared through Passing-Bablok regression analysis and Bland-Altman plot.</div></div><div><h3>Results</h3><div>We analyzed 65 blood samples, in detail 38 from adults and 27 from newborns.</div><div>The regression analysis results performed in the newborn and adult patients showed a good agreement between the two instruments. The evaluation of the Bland-Altman plots showed comparable values of bias <10 % for the most of parameters.</div><div>The evaluation of sample flags for the presence of distributional and morphological abnormalities showed a partial accordance between the two approaches, but the POCT exhibited good performance compared to the final report revised by the laboratory specialist.</div></div><div><h3>Conclusions</h3><div>The comparison of the two instruments demonstrated that they provide comparable blood counts, also in patients aged <3 months. The POCT allows having reliable analytical data and faster turning around time, particularly useful in NICU.</div></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"44 ","pages":"Article e00453"},"PeriodicalIF":1.7000,"publicationDate":"2025-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11787615/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S235255172500006X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background and aims
The use of a POCT (Point Of Care Test) could help in reducing the impact of pre-analytical errors in particular in challenging newborn samples.
The study purpose is to compare the POCT Sight OLO® hematology analyzer, validated for >3 months patients, with the reference system Sysmex XN-9100™ in Neonatal Intensive Care Unit (NICU).
Material and methods
The two analyzers were compared through Passing-Bablok regression analysis and Bland-Altman plot.
Results
We analyzed 65 blood samples, in detail 38 from adults and 27 from newborns.
The regression analysis results performed in the newborn and adult patients showed a good agreement between the two instruments. The evaluation of the Bland-Altman plots showed comparable values of bias <10 % for the most of parameters.
The evaluation of sample flags for the presence of distributional and morphological abnormalities showed a partial accordance between the two approaches, but the POCT exhibited good performance compared to the final report revised by the laboratory specialist.
Conclusions
The comparison of the two instruments demonstrated that they provide comparable blood counts, also in patients aged <3 months. The POCT allows having reliable analytical data and faster turning around time, particularly useful in NICU.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.