Photon-counting detector CTA to assess intracranial stents and flow diverters: an in vivo study with ultrahigh-resolution spectral reconstructions.

IF 3.7 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Frederic De Beukelaer, Sophie De Beukelaer, Laura L Wuyts, Omid Nikoubashman, Mohammed El Halal, Iliana Kantzeli, Martin Wiesmann, Hani Ridwan, Charlotte S Weyland
{"title":"Photon-counting detector CTA to assess intracranial stents and flow diverters: an in vivo study with ultrahigh-resolution spectral reconstructions.","authors":"Frederic De Beukelaer, Sophie De Beukelaer, Laura L Wuyts, Omid Nikoubashman, Mohammed El Halal, Iliana Kantzeli, Martin Wiesmann, Hani Ridwan, Charlotte S Weyland","doi":"10.1186/s41747-025-00550-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To define optimal parameters for the evaluation of vessel visibility in intracranial stents (ICS) and flow diverters (FD) using photon-counting detector computed tomography angiography (PCD-CTA) with spectral reconstructions.</p><p><strong>Methods: </strong>We retrospectively analyzed consecutive patients with implanted ICS or FD, who received a PCD-CTA between April 2023 and March 2024. Polyenergetic, virtual monoenergetic, pure lumen, and iodine reconstructions with different keV levels (40, 60, and 80) and reconstruction kernels (body vascular [Bv]48, Bv56, Bv64, Bv72, and Bv76) were evaluated by two radiologists with regions of interests and Likert scales. Reconstructions were compared in descriptive analysis.</p><p><strong>Results: </strong>In total, twelve patients with nine FDs and six ICSs were analyzed. In terms of quantitative image quality, sharper kernels as Bv64 and Bv72 yielded increased image noise and decreased signal-to-noise and contrast-to-noise ratios compared to the smoothest kernel Bv48 (p = 0.001). Among the different keV levels and kernels, readers selected the 40 keV level (p = 0.001) and sharper kernels (in the majority of cases Bv72) as the best to visualize the in-stent vessel lumen. Assessing the different spectral reconstructions virtual monoenergetic and iodine reconstructions proved to be best to evaluate in-stent vessel lumen (p = 0.001).</p><p><strong>Conclusion: </strong>PCD-CTA and spectral reconstructions with sharper reconstruction kernels and a low keV level of 40 seem to be beneficial to achieve optimal image quality for the evaluation of ICS and FD. Iodine and virtual monoenergetic reconstructions were superior to pure lumen and polyenergetic reconstructions to evaluate in-stent vessel lumen.</p><p><strong>Relevance statement: </strong>PCD-CTA offers the opportunity to reduce the need for invasive angiography serving as follow-up examination after intracranial stent (ICS) or flow diverter (FD) implantation.</p><p><strong>Key points: </strong>Neuroimaging of intracranial vessels with implanted stents and flow diverters is limited by artifacts. Twelve patients with nine flow diverters and six intracranial stents underwent photon-counting detector computed tomography angiography (PCD-CTA). In-stent vessel lumen visibility improved using sharp reconstruction kernels and a low keV level. Virtual monoenergetic and iodine reconstructions were best to evaluate in-stent vessel lumen.</p>","PeriodicalId":36926,"journal":{"name":"European Radiology Experimental","volume":"9 1","pages":"10"},"PeriodicalIF":3.7000,"publicationDate":"2025-01-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11780015/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Radiology Experimental","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s41747-025-00550-9","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: To define optimal parameters for the evaluation of vessel visibility in intracranial stents (ICS) and flow diverters (FD) using photon-counting detector computed tomography angiography (PCD-CTA) with spectral reconstructions.

Methods: We retrospectively analyzed consecutive patients with implanted ICS or FD, who received a PCD-CTA between April 2023 and March 2024. Polyenergetic, virtual monoenergetic, pure lumen, and iodine reconstructions with different keV levels (40, 60, and 80) and reconstruction kernels (body vascular [Bv]48, Bv56, Bv64, Bv72, and Bv76) were evaluated by two radiologists with regions of interests and Likert scales. Reconstructions were compared in descriptive analysis.

Results: In total, twelve patients with nine FDs and six ICSs were analyzed. In terms of quantitative image quality, sharper kernels as Bv64 and Bv72 yielded increased image noise and decreased signal-to-noise and contrast-to-noise ratios compared to the smoothest kernel Bv48 (p = 0.001). Among the different keV levels and kernels, readers selected the 40 keV level (p = 0.001) and sharper kernels (in the majority of cases Bv72) as the best to visualize the in-stent vessel lumen. Assessing the different spectral reconstructions virtual monoenergetic and iodine reconstructions proved to be best to evaluate in-stent vessel lumen (p = 0.001).

Conclusion: PCD-CTA and spectral reconstructions with sharper reconstruction kernels and a low keV level of 40 seem to be beneficial to achieve optimal image quality for the evaluation of ICS and FD. Iodine and virtual monoenergetic reconstructions were superior to pure lumen and polyenergetic reconstructions to evaluate in-stent vessel lumen.

Relevance statement: PCD-CTA offers the opportunity to reduce the need for invasive angiography serving as follow-up examination after intracranial stent (ICS) or flow diverter (FD) implantation.

Key points: Neuroimaging of intracranial vessels with implanted stents and flow diverters is limited by artifacts. Twelve patients with nine flow diverters and six intracranial stents underwent photon-counting detector computed tomography angiography (PCD-CTA). In-stent vessel lumen visibility improved using sharp reconstruction kernels and a low keV level. Virtual monoenergetic and iodine reconstructions were best to evaluate in-stent vessel lumen.

求助全文
约1分钟内获得全文 求助全文
来源期刊
European Radiology Experimental
European Radiology Experimental Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
6.70
自引率
2.60%
发文量
56
审稿时长
18 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信