Assessment of a New CT Detector and Filtration Technology: Part 2-Image Quality in Phantoms, Cadavers, and Patients.

IF 1 4区 医学 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Izabella Barreto, Graham Stoddard, Anahita Heshmat, Tara Massini, Ibrahim Tuna, Lynn Rill, Manuel Arreola
{"title":"Assessment of a New CT Detector and Filtration Technology: Part 2-Image Quality in Phantoms, Cadavers, and Patients.","authors":"Izabella Barreto, Graham Stoddard, Anahita Heshmat, Tara Massini, Ibrahim Tuna, Lynn Rill, Manuel Arreola","doi":"10.1097/RCT.0000000000001711","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The purpose of this work was to evaluate the image quality of a commercial CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO).</p><p><strong>Methods: </strong>CT number, noise, contrast-to-noise ratio (CNR), modulation transfer function (MTF), and noise power spectrum (NPS) were assessed using the ACR CT Accreditation phantom scanned with various acquisitions at 80 kV, 100 kV, 120 kV, and 135 kV, each with multiple CTDIvol values of 20 mGy, 40 mGy, and 65 mGy. Artifacts were evaluated in an anthropomorphic head phantom, a cadaver head, and in patient studies. Two neuroradiologists assessed image quality features in various patients who were examined with unenhanced brain CT on both scanners.</p><p><strong>Results: </strong>Compared with the conventional scanner, for the same CTDIvol, the PVO scanner produced 20.3% less image noise (P < 0.001), 18.9% higher CNR (P < 0.01), and 24.6% higher spatial resolution (P < 0.001). Streak artifacts were less severe with the PVO scanner for the phantom, cadaver, and patient scans (P < 0.05). Radiologists scored the PVO scanner as significantly better for visualization of the cerebrospinal fluid space over the cerebral sulci in high convexity, image noise in gray and white matter, and artifacts in the posterior fossa. They also significantly preferred the PVO scanner for visualization of the border between brain gray and white matter, cerebrospinal fluid space around the mesencephalon, and overall diagnostic acceptability.</p><p><strong>Conclusions: </strong>For matched CTDIvol values, the scanner equipped with PVO technology produced better objective and subjective image quality metrics in brain CT imaging compared with a conventional CT scanner without PVO. In clinical settings, PVO may allow for lower doses while enhancing imaging through dense areas, improving visualization of subtle details, and offering more effective options for examining obese patients.This research received financial support from Canon Medical Systems USA. The study design and data were fully controlled by the coauthors, of which none are employees or consultants of Canon Medical Systems USA.</p>","PeriodicalId":15402,"journal":{"name":"Journal of Computer Assisted Tomography","volume":" ","pages":""},"PeriodicalIF":1.0000,"publicationDate":"2025-01-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Computer Assisted Tomography","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RCT.0000000000001711","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The purpose of this work was to evaluate the image quality of a commercial CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO).

Methods: CT number, noise, contrast-to-noise ratio (CNR), modulation transfer function (MTF), and noise power spectrum (NPS) were assessed using the ACR CT Accreditation phantom scanned with various acquisitions at 80 kV, 100 kV, 120 kV, and 135 kV, each with multiple CTDIvol values of 20 mGy, 40 mGy, and 65 mGy. Artifacts were evaluated in an anthropomorphic head phantom, a cadaver head, and in patient studies. Two neuroradiologists assessed image quality features in various patients who were examined with unenhanced brain CT on both scanners.

Results: Compared with the conventional scanner, for the same CTDIvol, the PVO scanner produced 20.3% less image noise (P < 0.001), 18.9% higher CNR (P < 0.01), and 24.6% higher spatial resolution (P < 0.001). Streak artifacts were less severe with the PVO scanner for the phantom, cadaver, and patient scans (P < 0.05). Radiologists scored the PVO scanner as significantly better for visualization of the cerebrospinal fluid space over the cerebral sulci in high convexity, image noise in gray and white matter, and artifacts in the posterior fossa. They also significantly preferred the PVO scanner for visualization of the border between brain gray and white matter, cerebrospinal fluid space around the mesencephalon, and overall diagnostic acceptability.

Conclusions: For matched CTDIvol values, the scanner equipped with PVO technology produced better objective and subjective image quality metrics in brain CT imaging compared with a conventional CT scanner without PVO. In clinical settings, PVO may allow for lower doses while enhancing imaging through dense areas, improving visualization of subtle details, and offering more effective options for examining obese patients.This research received financial support from Canon Medical Systems USA. The study design and data were fully controlled by the coauthors, of which none are employees or consultants of Canon Medical Systems USA.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.50
自引率
0.00%
发文量
230
审稿时长
4-8 weeks
期刊介绍: The mission of Journal of Computer Assisted Tomography is to showcase the latest clinical and research developments in CT, MR, and closely related diagnostic techniques. We encourage submission of both original research and review articles that have immediate or promissory clinical applications. Topics of special interest include: 1) functional MR and CT of the brain and body; 2) advanced/innovative MRI techniques (diffusion, perfusion, rapid scanning); and 3) advanced/innovative CT techniques (perfusion, multi-energy, dose-reduction, and processing).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信