{"title":"Alteplase Purge Solution for Impella 5.5 Ventricular Assist Device Purge System Occlusion: Case Report and Review of the Literature.","authors":"Benjamin Wang, Jimmy Zheng, Madeline D Silva, Zhuo Shi, Jared Fong, Calvin Diep, William Hiesinger, Karim Sallam","doi":"10.1097/MAT.0000000000002378","DOIUrl":null,"url":null,"abstract":"<p><p>The use of an alteplase (Activase) purge solution to address Impella ventricular assist device \"thrombosis\" or \"purge system occlusion\" has been mainly documented with earlier generation Impella devices (CP, 2.5, 5.0). Here, we report the use of an alteplase purge solution to manage Impella 5.5 purge system occlusion in a 31-year-old male admitted to the cardiac care unit in cardiogenic shock and listed for a heart transplant. Throughout the purge system occlusion, the patient demonstrated hemodynamic stability and overall pump flow and cardiac output were preserved. Initially, there was a lack of response in purge flow and pressure observed at the lower concentration of alteplase purge solution (0.04 mg/ml), yet after using the alteplase 4 mg/50 ml purge (0.08 mg/ml) solution concentration, a response was seen. No bleeding or hemodynamic complications were observed. In addition, a suggested management workflow and review of the case reports and case series published to date regarding Impella purge system occlusion is included in this article.</p>","PeriodicalId":8844,"journal":{"name":"ASAIO Journal","volume":" ","pages":""},"PeriodicalIF":3.1000,"publicationDate":"2025-01-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ASAIO Journal","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1097/MAT.0000000000002378","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
The use of an alteplase (Activase) purge solution to address Impella ventricular assist device "thrombosis" or "purge system occlusion" has been mainly documented with earlier generation Impella devices (CP, 2.5, 5.0). Here, we report the use of an alteplase purge solution to manage Impella 5.5 purge system occlusion in a 31-year-old male admitted to the cardiac care unit in cardiogenic shock and listed for a heart transplant. Throughout the purge system occlusion, the patient demonstrated hemodynamic stability and overall pump flow and cardiac output were preserved. Initially, there was a lack of response in purge flow and pressure observed at the lower concentration of alteplase purge solution (0.04 mg/ml), yet after using the alteplase 4 mg/50 ml purge (0.08 mg/ml) solution concentration, a response was seen. No bleeding or hemodynamic complications were observed. In addition, a suggested management workflow and review of the case reports and case series published to date regarding Impella purge system occlusion is included in this article.
期刊介绍:
ASAIO Journal is in the forefront of artificial organ research and development. On the cutting edge of innovative technology, it features peer-reviewed articles of the highest quality that describe research, development, the most recent advances in the design of artificial organ devices and findings from initial testing. Bimonthly, the ASAIO Journal features state-of-the-art investigations, laboratory and clinical trials, and discussions and opinions from experts around the world.
The official publication of the American Society for Artificial Internal Organs.