Performance of Fluid Infusion Systems in the Changing Atmospheric Pressures Encountered in Aviation.

IF 0.9 4区 医学 Q4 BIOPHYSICS
Ka Siu Fan, Megan Paterson, Fariba Shojaee-Moradie, Antonios Manoli, Victoria Edwards, Vivienne Lee, Ewan Hutchison, Robert M Gifford, Iain T Parsons, Gerd Koehler, Chantal Mathieu, Julia K Mader, Bruce R King, David Russell-Jones, Easa Consortium
{"title":"Performance of Fluid Infusion Systems in the Changing Atmospheric Pressures Encountered in Aviation.","authors":"Ka Siu Fan, Megan Paterson, Fariba Shojaee-Moradie, Antonios Manoli, Victoria Edwards, Vivienne Lee, Ewan Hutchison, Robert M Gifford, Iain T Parsons, Gerd Koehler, Chantal Mathieu, Julia K Mader, Bruce R King, David Russell-Jones, Easa Consortium","doi":"10.3357/AMHP.6477.2024","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>With the increasing use of aeromedical transport for critically ill patients, it is essential to understand the impact of pressure changes on drug infusion delivery systems. As airplanes ascend and descend, gases/bubbles are released from solutions when ambient pressure decreases and dissolves when pressure increases. This may affect mechanical fluid delivery systems and cause clinically significant changes, especially within a critical care setting. We aimed to evaluate the impact of pressure changes on volumetric pumps and syringe drivers.</p><p><strong>Methods: </strong>An in vitro study of six volumetric pumps and eight syringe drivers was conducted in a hypobaric chamber to mimic pressure changes during flights. Infusion devices were set to deliver water at 0.2 ml ⋅ h-1 and infused volumes were measured. There were 15 open-ended syringes also studied.</p><p><strong>Results: </strong>During ascent, syringe drivers and volumetric pumps over-delivered 173 µL and 38 µL of fluid. During descent, syringe drivers under-delivered by 166 µL, whereas volumetric pumps under-delivered by 9 µL. Syringe drivers experienced statistically significant changes in fluid delivery during both ascent and descent. In volumetric pumps, only the descent phase infusion differed significantly from other phases. The volume of fluid expansion is dependent on volume and the mechanical properties of the fluid.</p><p><strong>Discussion: </strong>Decreasing ambient pressure causes bubble formation, which displaces fluid, and increasing ambient pressure causes bubble reabsorption in mechanical infusion devices. Hence, atmospheric pressure changes during air travel may alter fluid delivery from medical fluid delivery systems and affect critically ill patients who require both aeromedical evacuation and accurate infusion of drugs. Fan KS, Paterson M, Shojaee-Moradie F, Manoli A, Edwards V, Lee V, Hutchison E, Gifford RM, Parsons IT, Koehler G, Mathieu C, Mader JK, King BR, Russell-Jones D; EASA Consortium. Performance of fluid infusion systems in the changing atmospheric pressures encountered in aviation. Aerosp Med Hum Perform. 2025; 96(1):4-11.</p>","PeriodicalId":7463,"journal":{"name":"Aerospace medicine and human performance","volume":"96 1","pages":"4-11"},"PeriodicalIF":0.9000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Aerospace medicine and human performance","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3357/AMHP.6477.2024","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOPHYSICS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: With the increasing use of aeromedical transport for critically ill patients, it is essential to understand the impact of pressure changes on drug infusion delivery systems. As airplanes ascend and descend, gases/bubbles are released from solutions when ambient pressure decreases and dissolves when pressure increases. This may affect mechanical fluid delivery systems and cause clinically significant changes, especially within a critical care setting. We aimed to evaluate the impact of pressure changes on volumetric pumps and syringe drivers.

Methods: An in vitro study of six volumetric pumps and eight syringe drivers was conducted in a hypobaric chamber to mimic pressure changes during flights. Infusion devices were set to deliver water at 0.2 ml ⋅ h-1 and infused volumes were measured. There were 15 open-ended syringes also studied.

Results: During ascent, syringe drivers and volumetric pumps over-delivered 173 µL and 38 µL of fluid. During descent, syringe drivers under-delivered by 166 µL, whereas volumetric pumps under-delivered by 9 µL. Syringe drivers experienced statistically significant changes in fluid delivery during both ascent and descent. In volumetric pumps, only the descent phase infusion differed significantly from other phases. The volume of fluid expansion is dependent on volume and the mechanical properties of the fluid.

Discussion: Decreasing ambient pressure causes bubble formation, which displaces fluid, and increasing ambient pressure causes bubble reabsorption in mechanical infusion devices. Hence, atmospheric pressure changes during air travel may alter fluid delivery from medical fluid delivery systems and affect critically ill patients who require both aeromedical evacuation and accurate infusion of drugs. Fan KS, Paterson M, Shojaee-Moradie F, Manoli A, Edwards V, Lee V, Hutchison E, Gifford RM, Parsons IT, Koehler G, Mathieu C, Mader JK, King BR, Russell-Jones D; EASA Consortium. Performance of fluid infusion systems in the changing atmospheric pressures encountered in aviation. Aerosp Med Hum Perform. 2025; 96(1):4-11.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Aerospace medicine and human performance
Aerospace medicine and human performance PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH -MEDICINE, GENERAL & INTERNAL
CiteScore
1.10
自引率
22.20%
发文量
272
期刊介绍: The peer-reviewed monthly journal, Aerospace Medicine and Human Performance (AMHP), formerly Aviation, Space, and Environmental Medicine, provides contact with physicians, life scientists, bioengineers, and medical specialists working in both basic medical research and in its clinical applications. It is the most used and cited journal in its field. It is distributed to more than 80 nations.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信