{"title":"Imaging dose and image quality of kilovoltage imaging implemented on a helical tomotherapy unit.","authors":"Eric D Ehler, Parham Alaei","doi":"10.1016/j.zemedi.2024.12.003","DOIUrl":null,"url":null,"abstract":"<p><p>The purpose of this work was to evaluate the imaging dose for the Accuray Radixact ClearRT system. Low-contrast resolution and CT number consistency was evaluated as well. CTDI measurements were compared to vendor supplied values, and similar measurements were done on a Varian TrueBeam kV cone-beam CT (CBCT) and a Philips Big Bore CT scanner. In-field imaging doses were measured using various protocols in an anthropomorphic phantom, while out-of-field doses were measured 10 cm from the imaging field edge. Comparison of the CTDI and the in-field doses showed considerable disagreement when the patient anatomy size was not congruent with the size of the CTDI phantom. While that is an expected outcome, this work provides an estimate of the differences for a variety of ClearRT protocols when this situation arises. The CNR was measured for all combinations of ClearRT settings for comparisons within the system, as well as for a comparison with a CBCT and fan-beam CT system. The CNR and dose information provided in this work can be used to aid in selecting a ClearRT imaging protocol. The CT number stability was tracked over 27 months; two instances where the CT number constancy exceeded tolerance were observed after service.</p>","PeriodicalId":101315,"journal":{"name":"Zeitschrift fur medizinische Physik","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-01-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zeitschrift fur medizinische Physik","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.zemedi.2024.12.003","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The purpose of this work was to evaluate the imaging dose for the Accuray Radixact ClearRT system. Low-contrast resolution and CT number consistency was evaluated as well. CTDI measurements were compared to vendor supplied values, and similar measurements were done on a Varian TrueBeam kV cone-beam CT (CBCT) and a Philips Big Bore CT scanner. In-field imaging doses were measured using various protocols in an anthropomorphic phantom, while out-of-field doses were measured 10 cm from the imaging field edge. Comparison of the CTDI and the in-field doses showed considerable disagreement when the patient anatomy size was not congruent with the size of the CTDI phantom. While that is an expected outcome, this work provides an estimate of the differences for a variety of ClearRT protocols when this situation arises. The CNR was measured for all combinations of ClearRT settings for comparisons within the system, as well as for a comparison with a CBCT and fan-beam CT system. The CNR and dose information provided in this work can be used to aid in selecting a ClearRT imaging protocol. The CT number stability was tracked over 27 months; two instances where the CT number constancy exceeded tolerance were observed after service.