Characteristics of Resection Parameters in Robot-Assisted Total Knee Arthroplasty With the Ligament Balancing Workflow.

IF 1.8 2区 医学 Q2 ORTHOPEDICS
Qing-Da Wei, Hao-Ming An, Wang Gu, Wei Sun, Rui Li, Wei Chai
{"title":"Characteristics of Resection Parameters in Robot-Assisted Total Knee Arthroplasty With the Ligament Balancing Workflow.","authors":"Qing-Da Wei, Hao-Ming An, Wang Gu, Wei Sun, Rui Li, Wei Chai","doi":"10.1111/os.14336","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Robotic-assisted total knee arthroplasty (TKA) is a novel orthopedic technique. The workflow of robotic-assisted TKA is quite different from that of traditional manual TKA and may result incompletely different resection parameters. Understanding these parameters may help surgeons better perform robotic-assisted TKA. This study aims to analyze the specific resection parameters of robotic-assisted TKA.</p><p><strong>Methods: </strong>We retrospectively reviewed 85 MAKO-assisted TKA surgeries performed by three surgeons at our institution between May 2021 and November 2023. All patients had unilateral primary knee arthritis, and a Triathlon PS (Stryker) knee prosthesis was used. Intraoperative resection plan, radiological outcomes, and clinical outcomes were collected among them. The angle between the transepicondylar axis (TEA) and the femoral prosthesis axis was defined as rTEA, the angle between the posterior condylar axis (PCA) and the femoral prosthesis axis was defined as rPCA. The t-test and the Chi-square test (or Fisher's exact probability test) were used to determine differences in categorical variables.</p><p><strong>Results: </strong>rTEA averaged 2.7° (range, 0°-6.7°), and rPCA averaged 4.9° (range, 0.2°-9.6°). The mean resection of the medial distal femur was 7.7 mm (range, 3.0-12.5), that of the lateral tibial plateau was 6.4 mm (range, 1.5-13.0), and that of the medial posterior condyle of the femur was 10.6 mm (range, 6.5-17.5), whereas that of the lateral posterior condyle of the femur was 6.7 mm (range, 2.0-13.0).</p><p><strong>Conclusion: </strong>Robotic-assisted TKA using the ligament balancing workflow generally resulted in greater external rotation than reported reference values for conventional manual TKA within the existing literature, with reference to both the TEA and PCA. In addition, tibial resection was generally less, and the joint line was generally shifted upwards.</p>","PeriodicalId":19566,"journal":{"name":"Orthopaedic Surgery","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2025-01-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Orthopaedic Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/os.14336","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Robotic-assisted total knee arthroplasty (TKA) is a novel orthopedic technique. The workflow of robotic-assisted TKA is quite different from that of traditional manual TKA and may result incompletely different resection parameters. Understanding these parameters may help surgeons better perform robotic-assisted TKA. This study aims to analyze the specific resection parameters of robotic-assisted TKA.

Methods: We retrospectively reviewed 85 MAKO-assisted TKA surgeries performed by three surgeons at our institution between May 2021 and November 2023. All patients had unilateral primary knee arthritis, and a Triathlon PS (Stryker) knee prosthesis was used. Intraoperative resection plan, radiological outcomes, and clinical outcomes were collected among them. The angle between the transepicondylar axis (TEA) and the femoral prosthesis axis was defined as rTEA, the angle between the posterior condylar axis (PCA) and the femoral prosthesis axis was defined as rPCA. The t-test and the Chi-square test (or Fisher's exact probability test) were used to determine differences in categorical variables.

Results: rTEA averaged 2.7° (range, 0°-6.7°), and rPCA averaged 4.9° (range, 0.2°-9.6°). The mean resection of the medial distal femur was 7.7 mm (range, 3.0-12.5), that of the lateral tibial plateau was 6.4 mm (range, 1.5-13.0), and that of the medial posterior condyle of the femur was 10.6 mm (range, 6.5-17.5), whereas that of the lateral posterior condyle of the femur was 6.7 mm (range, 2.0-13.0).

Conclusion: Robotic-assisted TKA using the ligament balancing workflow generally resulted in greater external rotation than reported reference values for conventional manual TKA within the existing literature, with reference to both the TEA and PCA. In addition, tibial resection was generally less, and the joint line was generally shifted upwards.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Orthopaedic Surgery
Orthopaedic Surgery ORTHOPEDICS-
CiteScore
3.40
自引率
14.30%
发文量
374
审稿时长
20 weeks
期刊介绍: Orthopaedic Surgery (OS) is the official journal of the Chinese Orthopaedic Association, focusing on all aspects of orthopaedic technique and surgery. The journal publishes peer-reviewed articles in the following categories: Original Articles, Clinical Articles, Review Articles, Guidelines, Editorials, Commentaries, Surgical Techniques, Case Reports and Meeting Reports.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信