Compressed SENSE acceleration factor influence on magnetic resonance image quality in patients with endometrial cancer.

IF 3.3 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Olivera Nikolić, Marijana Basta Nikolić, Miloš Pantelić, Sonja Lukač, Viktor Till, Sanja Stojanović, Una Molnar
{"title":"Compressed SENSE acceleration factor influence on magnetic resonance image quality in patients with endometrial cancer.","authors":"Olivera Nikolić, Marijana Basta Nikolić, Miloš Pantelić, Sonja Lukač, Viktor Till, Sanja Stojanović, Una Molnar","doi":"10.1016/j.ejmp.2025.104899","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To investigate the impact of compressed sensing - sensitivity encoding (CS-SENSE) acceleration factor on the diagnostic performance of magnetic resonance imaging (MRI) within standard female pelvis protocol in patients with endometrial cancer.</p><p><strong>Methods: </strong>T2-weighted turbo spin echo (TSE) sequence from standard female pelvic MRI protocol was chosen due to its long acquisition time and essential role in the evaluation of morphological characteristics of the female pelvic anatomical structures. Fully sampled reference scans and multiple prospectively 2x to 5x under-sampled CS-SENSE scans were acquired. Retrospectively, under-sampled scans were compared to fully sampled scans and visually assessed for image quality and diagnostic quality by two independent radiologists dedicated to urogenital imaging with different experience levels.</p><p><strong>Results: </strong>Images obtained with CS-SENSE accelerated acquisition were of diagnostically acceptable quality at up to 3x acceleration for T2 TSE in both axial and sagittal planes (with an acquisition time reduction of 64 %). Among all evaluated uterine structures, the junctional zone proved to be most sensitive to the influence of the acceleration factor. Statistical analysis showed statistically significant differences between image interpretation qualities between two radiologists of different experience levels (p < 0.05).</p><p><strong>Conclusion: </strong>CS-SENSE accelerated T2 TSE sequence of the female pelvis shows image quality similar to that of conventional acquisitions with reduced acquisition time. CS-SENSE can moderately reduce scan time, providing many benefits without losing the image quality.</p>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"130 ","pages":"104899"},"PeriodicalIF":3.3000,"publicationDate":"2025-01-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.ejmp.2025.104899","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: To investigate the impact of compressed sensing - sensitivity encoding (CS-SENSE) acceleration factor on the diagnostic performance of magnetic resonance imaging (MRI) within standard female pelvis protocol in patients with endometrial cancer.

Methods: T2-weighted turbo spin echo (TSE) sequence from standard female pelvic MRI protocol was chosen due to its long acquisition time and essential role in the evaluation of morphological characteristics of the female pelvic anatomical structures. Fully sampled reference scans and multiple prospectively 2x to 5x under-sampled CS-SENSE scans were acquired. Retrospectively, under-sampled scans were compared to fully sampled scans and visually assessed for image quality and diagnostic quality by two independent radiologists dedicated to urogenital imaging with different experience levels.

Results: Images obtained with CS-SENSE accelerated acquisition were of diagnostically acceptable quality at up to 3x acceleration for T2 TSE in both axial and sagittal planes (with an acquisition time reduction of 64 %). Among all evaluated uterine structures, the junctional zone proved to be most sensitive to the influence of the acceleration factor. Statistical analysis showed statistically significant differences between image interpretation qualities between two radiologists of different experience levels (p < 0.05).

Conclusion: CS-SENSE accelerated T2 TSE sequence of the female pelvis shows image quality similar to that of conventional acquisitions with reduced acquisition time. CS-SENSE can moderately reduce scan time, providing many benefits without losing the image quality.

压缩感知加速因子对子宫内膜癌患者磁共振图像质量的影响。
目的:探讨压缩感敏编码(CS-SENSE)加速因子对子宫内膜癌女性骨盆标准方案磁共振成像(MRI)诊断效能的影响。方法:选择来自标准女性骨盆MRI方案的t2加权涡轮自旋回波(TSE)序列,因为它的采集时间长,对评估女性骨盆解剖结构的形态学特征具有重要作用。获得了完全采样的参考扫描和多次2x至5倍欠采样的CS-SENSE扫描。回顾性地,将欠采样扫描与全采样扫描进行比较,并由两名具有不同经验水平的独立的泌尿生殖系统成像放射科医生对图像质量和诊断质量进行视觉评估。结果:CS-SENSE加速采集获得的图像在轴位和矢状面T2 TSE高达3倍的加速下具有诊断可接受的质量(采集时间减少64%)。在所有被评估的子宫结构中,交界区对加速因子的影响最为敏感。结论:CS-SENSE女性骨盆加速T2 TSE序列图像质量与常规采集图像质量相近,但采集时间缩短。CS-SENSE可以适度减少扫描时间,在不损失图像质量的情况下提供许多好处。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信