Mattia Liccardo, Ersilia Cipolletta, Emma Arezzi, Monica Sicuranza, Maria Rea, Pietro Ricciardi, Giovanni Napolitano
{"title":"External DC Shocks of Symptomatic Supraventricular Tachycardia During Screw-In Leadless Pacemaker Implant: A Case Report.","authors":"Mattia Liccardo, Ersilia Cipolletta, Emma Arezzi, Monica Sicuranza, Maria Rea, Pietro Ricciardi, Giovanni Napolitano","doi":"10.1111/pace.15134","DOIUrl":null,"url":null,"abstract":"<p><p>We present the case of a 64-year-old man who, during the implantation of an active-fixation leadless pacemaker (LP, Aveir VR, Abbott, USA), underwent several external defibrillation shocks up to 240 Joules, due to symptomatic sustained supraventricular tachycardia at 160 bpm. The shocks, delivered both before and after the screwing of the device in the low interventricular septum, did not cause any technical damage to the device, and no complications were observed. The device was then deployed successfully. To our knowledge, this is the first documented case of external cardioversion (ECV) in a patient with an active-fixation LP. The results suggest that ECV in these patients appears to be safe and feasible.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":"315-319"},"PeriodicalIF":1.7000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15134","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/10 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
We present the case of a 64-year-old man who, during the implantation of an active-fixation leadless pacemaker (LP, Aveir VR, Abbott, USA), underwent several external defibrillation shocks up to 240 Joules, due to symptomatic sustained supraventricular tachycardia at 160 bpm. The shocks, delivered both before and after the screwing of the device in the low interventricular septum, did not cause any technical damage to the device, and no complications were observed. The device was then deployed successfully. To our knowledge, this is the first documented case of external cardioversion (ECV) in a patient with an active-fixation LP. The results suggest that ECV in these patients appears to be safe and feasible.
期刊介绍:
Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.