Clinical value of portable 12-lead electrocardiography devices in patients with heart disease: A validation study.

IF 1.3 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
Xiuyan Lu, Qiuhuan Wang, Xiujie Sun, Yibing Shao, Wenbo Jiang
{"title":"Clinical value of portable 12-lead electrocardiography devices in patients with heart disease: A validation study.","authors":"Xiuyan Lu, Qiuhuan Wang, Xiujie Sun, Yibing Shao, Wenbo Jiang","doi":"10.1016/j.jelectrocard.2024.153835","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The present study was conducted to assess the accuracy and reliability of portable 12‑lead electrocardiography (ECG) devices in patients with heart disease.</p><p><strong>Materials and methods: </strong>This single-center, prospective, blinded study enrolled 62 patients between September and October 2023 from the Heart Center of a Class III hospital. In sequential tests on each patient, heart rate (HR) and the PR, QT, QTc and QRS intervals of ECG recordings obtained with a portable 12‑lead device (Weheal, CN) were compared with those obtained via conventional 12‑lead ECG. ECG parameters were read in batches by 3 blinded electrophysiologists. Two-tailed paired t-tests were used to compare the continuous variables. Agreement was evaluated via Bland-Altman plots.</p><p><strong>Results: </strong>Sixty-two patients were included. HR and the QT, QTc and QRS intervals from the portable 12‑lead electrocardiogram recordings were essentially the same as those obtained via conventional ECG. Bland-Altman analysis revealed no significant differences in these values, indicating suitable agreement between the 2 measurements. The PR interval was 176.89 ± 29.53 ms in the portable group and 161.56 ± 17.78 ms in the standard group, which was statistically (p < 0.001) but not clinically significant.</p><p><strong>Conclusions: </strong>ECG recordings obtained with a portable 12‑lead device (Weheal, CN) allow for accurate HR, PR, QT, QTc and QRS assessments. Considering its simplicity, this approach has advantages over conventional ECG and can provide an alternative for evaluating patients outside the hospital. How to improve patients' acceptance of portable ECG machines still needs further research.</p>","PeriodicalId":15606,"journal":{"name":"Journal of electrocardiology","volume":"88 ","pages":"153835"},"PeriodicalIF":1.3000,"publicationDate":"2024-11-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of electrocardiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jelectrocard.2024.153835","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The present study was conducted to assess the accuracy and reliability of portable 12‑lead electrocardiography (ECG) devices in patients with heart disease.

Materials and methods: This single-center, prospective, blinded study enrolled 62 patients between September and October 2023 from the Heart Center of a Class III hospital. In sequential tests on each patient, heart rate (HR) and the PR, QT, QTc and QRS intervals of ECG recordings obtained with a portable 12‑lead device (Weheal, CN) were compared with those obtained via conventional 12‑lead ECG. ECG parameters were read in batches by 3 blinded electrophysiologists. Two-tailed paired t-tests were used to compare the continuous variables. Agreement was evaluated via Bland-Altman plots.

Results: Sixty-two patients were included. HR and the QT, QTc and QRS intervals from the portable 12‑lead electrocardiogram recordings were essentially the same as those obtained via conventional ECG. Bland-Altman analysis revealed no significant differences in these values, indicating suitable agreement between the 2 measurements. The PR interval was 176.89 ± 29.53 ms in the portable group and 161.56 ± 17.78 ms in the standard group, which was statistically (p < 0.001) but not clinically significant.

Conclusions: ECG recordings obtained with a portable 12‑lead device (Weheal, CN) allow for accurate HR, PR, QT, QTc and QRS assessments. Considering its simplicity, this approach has advantages over conventional ECG and can provide an alternative for evaluating patients outside the hospital. How to improve patients' acceptance of portable ECG machines still needs further research.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of electrocardiology
Journal of electrocardiology 医学-心血管系统
CiteScore
2.70
自引率
7.70%
发文量
152
审稿时长
38 days
期刊介绍: The Journal of Electrocardiology is devoted exclusively to clinical and experimental studies of the electrical activities of the heart. It seeks to contribute significantly to the accuracy of diagnosis and prognosis and the effective treatment, prevention, or delay of heart disease. Editorial contents include electrocardiography, vectorcardiography, arrhythmias, membrane action potential, cardiac pacing, monitoring defibrillation, instrumentation, drug effects, and computer applications.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信