Comparison of fluorescence intensity of protoporphyrin IX as observed on the screen of different cystoscopic systems.

Hideo Fukuhara, Takahiro Nishimura, Yu Shimojo, Keiji Inoue
{"title":"Comparison of fluorescence intensity of protoporphyrin IX as observed on the screen of different cystoscopic systems.","authors":"Hideo Fukuhara, Takahiro Nishimura, Yu Shimojo, Keiji Inoue","doi":"10.1016/j.pdpdt.2024.104425","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Photodynamic diagnosis-assisted transurethral resection of bladder tumor (PDD-TURBT) is an essential and useful intervention used for removing non-invasive bladder cancer (NMIBC). Several PDD devices can be used in clinical practice, but few reports have compared them.In this study, we examined the differences in detected fluorescence intensity for each PDD device.</p><p><strong>Methods: </strong>The PDD device was used from OLYMPUS, Richard Wolf, and Karl Storz. Light power and illuminance from endoscope were measured. For each PDD device, fluorescence intensity was measured at 3 cm and 8 cm distance from the protoporphyrin IX (PpIX) phantom to the cystoscope.</p><p><strong>Results: </strong>The variation of excitation light power with observation distance were 35.0 %, 27.0 %, and 28.2 % for OLYMPUS, Richard Wolf, and Karl Storz, respectively. The variation of fluorescence intensity with observation distance were 132.8%, 120.5%, and 49.8% for OLYMPUS, Richard Wolf, and Karl Storz, respectively.</p><p><strong>Conclusions: </strong>PDD devices used in clinical practice show both some performance differences between the PDD systems themselves, and also differences in observed fluorescence intensity.</p>","PeriodicalId":94170,"journal":{"name":"Photodiagnosis and photodynamic therapy","volume":" ","pages":"104425"},"PeriodicalIF":0.0000,"publicationDate":"2024-11-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Photodiagnosis and photodynamic therapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.pdpdt.2024.104425","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Photodynamic diagnosis-assisted transurethral resection of bladder tumor (PDD-TURBT) is an essential and useful intervention used for removing non-invasive bladder cancer (NMIBC). Several PDD devices can be used in clinical practice, but few reports have compared them.In this study, we examined the differences in detected fluorescence intensity for each PDD device.

Methods: The PDD device was used from OLYMPUS, Richard Wolf, and Karl Storz. Light power and illuminance from endoscope were measured. For each PDD device, fluorescence intensity was measured at 3 cm and 8 cm distance from the protoporphyrin IX (PpIX) phantom to the cystoscope.

Results: The variation of excitation light power with observation distance were 35.0 %, 27.0 %, and 28.2 % for OLYMPUS, Richard Wolf, and Karl Storz, respectively. The variation of fluorescence intensity with observation distance were 132.8%, 120.5%, and 49.8% for OLYMPUS, Richard Wolf, and Karl Storz, respectively.

Conclusions: PDD devices used in clinical practice show both some performance differences between the PDD systems themselves, and also differences in observed fluorescence intensity.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信