Application of Multi-Model Adaptive Statistical Iterative Reconstruction-Veo in Ultra-Low Dose Chest CT Examination of Children in Plateau Area.

Q4 Medicine
Xian-Tao Wang, Rui-Ting Bai, Cidanwangjiu, Suolangnima, Nimazhuoga, Bai-Yan Su
{"title":"Application of Multi-Model Adaptive Statistical Iterative Reconstruction-Veo in Ultra-Low Dose Chest CT Examination of Children in Plateau Area.","authors":"Xian-Tao Wang, Rui-Ting Bai, Cidanwangjiu, Suolangnima, Nimazhuoga, Bai-Yan Su","doi":"10.3881/j.issn.1000-503X.16068","DOIUrl":null,"url":null,"abstract":"<p><p>Objective To explore the application value of multi-model adaptive statistical iterative reconstruction-Veo (ASiR-V) in ultra-low dose chest CT examination of children in the plateau area. Methods The children who underwent chest CT examination in Xizang Autonomous Region People's Hospital were enrolled in this study and assigned into two groups according to the scanning conditions.Group A underwent scanning at a tube voltage of 100 kV and ASiR-V 50% reconstruction,and group B underwent scanning at a tube voltage of 80 kV and ASiR-V 0 (Group B1) and ASiR-V 50% (Group B2) reconstruction.The image quality of each group was evaluated objectively and subjectively.The radiation dose and image quality were compared between groups. Results Groups A and B showed the volume CT dose indexes of (2.33±0.62) mGy and (0.86±0.01) mGy and the dose length products of (65.01±25.12) mGy·cm and (23.55±3.38) mGy·cm,respectively,which presented differences between groups (both P<0.001).The image noise in the bilateral upper and middle lung areas in group B2 was lower than that in group B1 but higher than that in group A (all P<0.001).There was no significant difference in image quality score of the lung window among groups (all P>0.05).Groups A,B1,and B2 had no significant differences in ascending aorta (P=0.538) or liver CT value (P=0.175) in the mediastinal window.The signal-to-noise ratios and contrast-to-noise ratios of ascending aorta and liver in group B2 were higher than those in group B1 (all P<0.001) and lower than those in group A (all P<0.05).The image quality score of the mediastinal window followed a descending order of group A>group B2>group B1 (all P<0.001). Conclusion ASiR-V combined with low tube voltage can effectively reduce the radiation dose and guarantee the image quality of chest CT of children in the plateau area.</p>","PeriodicalId":6919,"journal":{"name":"中国医学科学院学报","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-11-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医学科学院学报","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.3881/j.issn.1000-503X.16068","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective To explore the application value of multi-model adaptive statistical iterative reconstruction-Veo (ASiR-V) in ultra-low dose chest CT examination of children in the plateau area. Methods The children who underwent chest CT examination in Xizang Autonomous Region People's Hospital were enrolled in this study and assigned into two groups according to the scanning conditions.Group A underwent scanning at a tube voltage of 100 kV and ASiR-V 50% reconstruction,and group B underwent scanning at a tube voltage of 80 kV and ASiR-V 0 (Group B1) and ASiR-V 50% (Group B2) reconstruction.The image quality of each group was evaluated objectively and subjectively.The radiation dose and image quality were compared between groups. Results Groups A and B showed the volume CT dose indexes of (2.33±0.62) mGy and (0.86±0.01) mGy and the dose length products of (65.01±25.12) mGy·cm and (23.55±3.38) mGy·cm,respectively,which presented differences between groups (both P<0.001).The image noise in the bilateral upper and middle lung areas in group B2 was lower than that in group B1 but higher than that in group A (all P<0.001).There was no significant difference in image quality score of the lung window among groups (all P>0.05).Groups A,B1,and B2 had no significant differences in ascending aorta (P=0.538) or liver CT value (P=0.175) in the mediastinal window.The signal-to-noise ratios and contrast-to-noise ratios of ascending aorta and liver in group B2 were higher than those in group B1 (all P<0.001) and lower than those in group A (all P<0.05).The image quality score of the mediastinal window followed a descending order of group A>group B2>group B1 (all P<0.001). Conclusion ASiR-V combined with low tube voltage can effectively reduce the radiation dose and guarantee the image quality of chest CT of children in the plateau area.

多模型自适应统计迭代重建-Veo在高原地区儿童超低剂量胸部CT检查中的应用
目的 探讨多模型自适应统计迭代重建-Veo(ASiR-V)在高原地区儿童超低剂量胸部CT检查中的应用价值。方法 将在西藏自治区人民医院接受胸部 CT 检查的患儿作为研究对象,根据扫描条件分为两组,A 组在 100 kV 管电压下扫描,ASiR-V 50%重建,B 组在 80 kV 管电压下扫描,ASiR-V 0(B1 组)和 ASiR-V 50%(B2 组)重建,对各组图像质量进行客观和主观评价,比较组间辐射剂量和图像质量。结果 A 组和 B 组的容积 CT 剂量指数分别为(2.33±0.62)mGy 和(0.86±0.01)mGy,剂量长度乘积分别为(65.01±25.12)mGy-cm 和(23.55±3.38)mGy-cm,组间差异均为 P0.05。A组、B1组和B2组纵隔窗升主动脉(P=0.538)和肝脏CT值(P=0.175)差异无学意义。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
中国医学科学院学报
中国医学科学院学报 Medicine-Medicine (all)
CiteScore
0.60
自引率
0.00%
发文量
6813
期刊介绍: Acta Academiae Medicinae Sinicae was founded in February 1979. It is a comprehensive medical academic journal published in China and abroad, supervised by the Ministry of Health of the People's Republic of China and sponsored by the Chinese Academy of Medical Sciences and Peking Union Medical College. The journal mainly reports the latest research results, work progress and dynamics in the fields of basic medicine, clinical medicine, pharmacy, preventive medicine, biomedicine, medical teaching and research, aiming to promote the exchange of medical information and improve the academic level of medicine. At present, the journal has been included in 10 famous foreign retrieval systems and their databases [Medline (PubMed online version), Elsevier, EMBASE, CA, WPRIM, ExtraMED, IC, JST, UPD and EBSCO-ASP]; and has been included in important domestic retrieval systems and databases [China Science Citation Database (Documentation and Information Center of the Chinese Academy of Sciences), China Core Journals Overview (Peking University Library), China Science and Technology Paper Statistical Source Database (China Science and Technology Core Journals) (China Institute of Scientific and Technological Information), China Science and Technology Journal Paper and Citation Database (China Institute of Scientific and Technological Information)].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信