{"title":"The feasibility of a serial 12‑lead ECG wireless patch in the hospital setting","authors":"Mary G. Carey PhD, RN, FAHA, FAAN ,&nbsp;Sudhir K. Mummidi MBBS, RN ,&nbsp;Amanda Kammer BS, RN ,&nbsp;Dillon J. Dzikowicz PhD, RN, PCCN","doi":"10.1016/j.jelectrocard.2024.153828","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Chest pain is the second most common reason to present to the emergency department in the United States, and the ECG is a first-line diagnostic tool for myocardial ischemia assessment. For patients with ongoing symptoms or unclear initial ECGs, guidelines recommend performing multiple standard ECGs at 15–30-min intervals during the first 1–2 h, which improves acute coronary syndrome (ACS) detection by 15 % and accelerates triage of high-risk ACS patients. However, obtaining serial ECG is not consistently practiced due to overcrowding and the limited technical abilities of current 12‑lead ECG machines. This study aimed to evaluate an FDA-approved wireless 12‑lead ECG patch for serial cardiac monitoring in the hospital setting.</div></div><div><h3>Methods</h3><div>Prospectively, ECG patch was applied in the Mason-Likar electrode configuration after obtaining consent. The patch remained in place for at least one hour. Clinical Utility of the ECGs was categorized from 1 to 3: 1 = uninterpretable, 2 = borderline, and 3 = interpretable.</div></div><div><h3>Results</h3><div>Among hospitalized cardiac patients, 28 consented to wear the ECG patch for at least one hour and patients were free to ambulate during the study. Most (70 %) patients were in sinus rhythm, and an episode of asymptomatic TMI was captured. The clinical utility of the ECGs (<em>n</em> = 364) was mostly interpretable, 64 % (<em>n</em> = 231), while 15 % (<em>n</em> = 55) were uninterpretable and 18 % (<em>n</em> = 65) were borderline. Most (69 %) preferred the patch, while 12 % preferred telemetry. The hospitalized cardiac patients reported significantly better ability to ambulate with the ECG patch (Z = −3.607, <em>p</em> &lt; 0.001).</div></div><div><h3>Conclusion</h3><div>Thus, this experiment demonstrated that the ECG patch provides quality serial ECG monitoring and captures TMI of hospitalized cardiac patients without increasing burden.</div></div>","PeriodicalId":15606,"journal":{"name":"Journal of electrocardiology","volume":"88 ","pages":"Article 153828"},"PeriodicalIF":1.3000,"publicationDate":"2024-11-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of electrocardiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S002207362400298X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Chest pain is the second most common reason to present to the emergency department in the United States, and the ECG is a first-line diagnostic tool for myocardial ischemia assessment. For patients with ongoing symptoms or unclear initial ECGs, guidelines recommend performing multiple standard ECGs at 15–30-min intervals during the first 1–2 h, which improves acute coronary syndrome (ACS) detection by 15 % and accelerates triage of high-risk ACS patients. However, obtaining serial ECG is not consistently practiced due to overcrowding and the limited technical abilities of current 12‑lead ECG machines. This study aimed to evaluate an FDA-approved wireless 12‑lead ECG patch for serial cardiac monitoring in the hospital setting.

Methods

Prospectively, ECG patch was applied in the Mason-Likar electrode configuration after obtaining consent. The patch remained in place for at least one hour. Clinical Utility of the ECGs was categorized from 1 to 3: 1 = uninterpretable, 2 = borderline, and 3 = interpretable.

Results

Among hospitalized cardiac patients, 28 consented to wear the ECG patch for at least one hour and patients were free to ambulate during the study. Most (70 %) patients were in sinus rhythm, and an episode of asymptomatic TMI was captured. The clinical utility of the ECGs (n = 364) was mostly interpretable, 64 % (n = 231), while 15 % (n = 55) were uninterpretable and 18 % (n = 65) were borderline. Most (69 %) preferred the patch, while 12 % preferred telemetry. The hospitalized cardiac patients reported significantly better ability to ambulate with the ECG patch (Z = −3.607, p < 0.001).

Conclusion

Thus, this experiment demonstrated that the ECG patch provides quality serial ECG monitoring and captures TMI of hospitalized cardiac patients without increasing burden.
在医院环境中使用串行 12 导联心电图无线贴片的可行性
背景在美国,胸痛是急诊科第二大常见病因,而心电图是评估心肌缺血的一线诊断工具。对于有持续症状或初始心电图不明确的患者,指南建议在最初的 1-2 小时内以 15-30 分钟的间隔进行多次标准心电图检查,这样可将急性冠状动脉综合征(ACS)的检出率提高 15%,并加快高危 ACS 患者的分流。然而,由于过度拥挤和目前 12 导联心电图机的技术能力有限,获取连续心电图的做法并不普遍。本研究旨在评估一种经 FDA 批准的用于医院环境中连续心脏监测的无线 12 导联心电图贴片。贴片放置至少一小时。结果在住院心脏病患者中,有 28 人同意佩戴心电图贴片至少一小时,研究期间患者可以自由活动。大多数患者(70%)处于窦性心律,并记录了一次无症状的 TMI。心电图(364 人)的临床实用性大部分可以解释,占 64%(231 人),15%(55 人)不可解释,18%(65 人)处于边缘状态。大多数人(69%)更喜欢贴片,而 12% 的人更喜欢遥测。结论本实验证明,心电图贴片可提供高质量的连续心电图监测,并在不增加负担的情况下捕捉住院心脏病患者的 TMI。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of electrocardiology
Journal of electrocardiology 医学-心血管系统
CiteScore
2.70
自引率
7.70%
发文量
152
审稿时长
38 days
期刊介绍: The Journal of Electrocardiology is devoted exclusively to clinical and experimental studies of the electrical activities of the heart. It seeks to contribute significantly to the accuracy of diagnosis and prognosis and the effective treatment, prevention, or delay of heart disease. Editorial contents include electrocardiography, vectorcardiography, arrhythmias, membrane action potential, cardiac pacing, monitoring defibrillation, instrumentation, drug effects, and computer applications.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信