Automatic patient centering in computed tomography: a systematic review and meta-analysis.

IF 4.7 2区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Yasser H Hadi, Lauren Keaney, Andrew England, Niamh Moore, Mark McEntee
{"title":"Automatic patient centering in computed tomography: a systematic review and meta-analysis.","authors":"Yasser H Hadi, Lauren Keaney, Andrew England, Niamh Moore, Mark McEntee","doi":"10.1007/s00330-024-11170-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To comprehensively examine the influence of auto-patient centering technologies on positioning accuracy, radiation dose, image quality, and time efficiency of computed tomography (CT) scans.</p><p><strong>Materials and methods: </strong>A systematic search of peer-reviewed English publications was performed between January 2000 and November 2023 in PubMed, Embase, CINAHL, Scopus, and Web of Science. Two postgraduate students and an academic lecturer independently reviewed the articles to verify adherence to the inclusion criteria. The QUADAS-2 tool was employed to evaluate study quality. We derived summary estimates on positioning accuracy, radiation dose reduction, image quality, and time efficiency using proportion and meta-analysis methodologies.</p><p><strong>Results: </strong>Nine studies were identified comparing automatic and manual CT positioning. Automatic positioning improved accuracy by reducing vertical offsets to 7 mm and 4 mm for thorax and abdominal CTs, compared to 19 mm and 18 mm with manual methods. Most studies showed significant reductions in radiation dose, ranging from 5.71 to 31%. Image quality results were mixed, automatic methods generally produced images with less noise, but differences were minimal. Time efficiency was better, with automatic positioning reducing preparation time from 0.48 min versus 0.67 min for manual positioning.</p><p><strong>Conclusions: </strong>This review confirms that automatic patient-centering technologies enhance positioning accuracy and decrease preparation times in CT scans. While reductions in radiation doses and some improvements in image quality were observed, the evidence remains mixed. Findings support integrating these technologies into clinical practice to optimize patient care.</p><p><strong>Key points: </strong>Question Does automatic patient centering in CT enhance positioning accuracy, reduce radiation exposure, and improve image quality? Findings Findings indicate that automatic centering can optimize image quality, reduce examination times and contribute to overall improvements in imaging efficiency. Clinical relevance Automatic patient centering in CT improves positioning accuracy, minimizes radiation exposure, enhances image quality, and accelerates imaging workflows, contributing to safer, more efficient imaging procedures that benefit patient care.</p>","PeriodicalId":12076,"journal":{"name":"European Radiology","volume":" ","pages":""},"PeriodicalIF":4.7000,"publicationDate":"2024-11-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00330-024-11170-z","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To comprehensively examine the influence of auto-patient centering technologies on positioning accuracy, radiation dose, image quality, and time efficiency of computed tomography (CT) scans.

Materials and methods: A systematic search of peer-reviewed English publications was performed between January 2000 and November 2023 in PubMed, Embase, CINAHL, Scopus, and Web of Science. Two postgraduate students and an academic lecturer independently reviewed the articles to verify adherence to the inclusion criteria. The QUADAS-2 tool was employed to evaluate study quality. We derived summary estimates on positioning accuracy, radiation dose reduction, image quality, and time efficiency using proportion and meta-analysis methodologies.

Results: Nine studies were identified comparing automatic and manual CT positioning. Automatic positioning improved accuracy by reducing vertical offsets to 7 mm and 4 mm for thorax and abdominal CTs, compared to 19 mm and 18 mm with manual methods. Most studies showed significant reductions in radiation dose, ranging from 5.71 to 31%. Image quality results were mixed, automatic methods generally produced images with less noise, but differences were minimal. Time efficiency was better, with automatic positioning reducing preparation time from 0.48 min versus 0.67 min for manual positioning.

Conclusions: This review confirms that automatic patient-centering technologies enhance positioning accuracy and decrease preparation times in CT scans. While reductions in radiation doses and some improvements in image quality were observed, the evidence remains mixed. Findings support integrating these technologies into clinical practice to optimize patient care.

Key points: Question Does automatic patient centering in CT enhance positioning accuracy, reduce radiation exposure, and improve image quality? Findings Findings indicate that automatic centering can optimize image quality, reduce examination times and contribute to overall improvements in imaging efficiency. Clinical relevance Automatic patient centering in CT improves positioning accuracy, minimizes radiation exposure, enhances image quality, and accelerates imaging workflows, contributing to safer, more efficient imaging procedures that benefit patient care.

计算机断层扫描中的患者自动居中:系统回顾和荟萃分析。
目的:全面研究患者自动定中心技术对定位精度、辐射剂量、图像质量和时间效率的影响:全面研究患者自动定中心技术对计算机断层扫描(CT)的定位精度、辐射剂量、图像质量和时间效率的影响:在 PubMed、Embase、CINAHL、Scopus 和 Web of Science 中对 2000 年 1 月至 2023 年 11 月期间经同行评审的英文出版物进行了系统检索。两名研究生和一名学术讲师对文章进行了独立审阅,以核实是否符合纳入标准。采用 QUADAS-2 工具评估研究质量。我们使用比例和荟萃分析方法对定位准确性、辐射剂量减少、图像质量和时间效率进行了总结评估:结果:九项研究对自动和手动 CT 定位进行了比较。胸部和腹部 CT 的自动定位将垂直偏移量分别减少到 7 毫米和 4 毫米,从而提高了定位的准确性,而手动方法的垂直偏移量分别为 19 毫米和 18 毫米。大多数研究显示辐射剂量明显减少,从 5.71% 到 31% 不等。图像质量参差不齐,自动方法生成的图像一般噪点较少,但差异很小。时间效率更高,自动定位可将准备时间从 0.48 分钟减少到 0.67 分钟:本综述证实,患者自动对中技术提高了 CT 扫描中定位的准确性并缩短了准备时间。虽然辐射剂量有所减少,图像质量也有所改善,但证据仍然参差不齐。研究结果支持将这些技术融入临床实践,以优化患者护理:问题 CT 患者自动居中是否能提高定位准确性、减少辐射量并改善图像质量?研究结果 研究结果表明,自动居中可优化图像质量、缩短检查时间并有助于全面提高成像效率。临床意义 CT 患者自动居中可提高定位准确性、减少辐射暴露、提高图像质量并加快成像工作流程,从而有助于更安全、更高效的成像程序,有利于患者护理。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
European Radiology
European Radiology 医学-核医学
CiteScore
11.60
自引率
8.50%
发文量
874
审稿时长
2-4 weeks
期刊介绍: European Radiology (ER) continuously updates scientific knowledge in radiology by publication of strong original articles and state-of-the-art reviews written by leading radiologists. A well balanced combination of review articles, original papers, short communications from European radiological congresses and information on society matters makes ER an indispensable source for current information in this field. This is the Journal of the European Society of Radiology, and the official journal of a number of societies. From 2004-2008 supplements to European Radiology were published under its companion, European Radiology Supplements, ISSN 1613-3749.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信