{"title":"The impact of technical and environmental conditions on the quality assessment in mammography","authors":"S. Papathanasiou , K. Szczepura , J.D. Thompson","doi":"10.1016/j.radi.2024.10.013","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><div>Mammography requires optimal image quality and that should be ensured during quality assurance. This study investigates the impact of environmental conditions and monitor specifications on quality control procedures within image acquisition rooms in breast imaging departments.</div></div><div><h3>Methods</h3><div>During this study, nine TORMAM test object images acquired under different conditions were evaluated by 16 observers in 12 different environmental conditions (low, medium, and high illumination level, white and grey wall colour and 2 monitors with high and low technical characteristics). Visibility of structures was the key criteria.</div></div><div><h3>Results</h3><div>The number of visible structures per image was dependent on the different environmental conditions, with large variations observed. The wall colour and the illumination level have a statistically significant effect on the number of visible structures. It was statistically proven that the grey wall colour had a positive effect on the visibility of low contrast detail discs.</div></div><div><h3>Conclusion</h3><div>Low ambient light, with a grey wall colour and monitors with high specification allow greatest structure visibility. On the contrary, white wall colour around the monitor and high ambient light had a negative impact on technical evaluation of the images during quality control procedures.</div></div><div><h3>Implications for practice</h3><div>Better standardization of the environmental condition is required in acquisition rooms. Specifically, this research points to the benefit of using a low reflectance wall colour and low illumination level around the monitors.</div></div>","PeriodicalId":47416,"journal":{"name":"Radiography","volume":"31 1","pages":"Pages 131-136"},"PeriodicalIF":2.5000,"publicationDate":"2024-11-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiography","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1078817424003109","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
Mammography requires optimal image quality and that should be ensured during quality assurance. This study investigates the impact of environmental conditions and monitor specifications on quality control procedures within image acquisition rooms in breast imaging departments.
Methods
During this study, nine TORMAM test object images acquired under different conditions were evaluated by 16 observers in 12 different environmental conditions (low, medium, and high illumination level, white and grey wall colour and 2 monitors with high and low technical characteristics). Visibility of structures was the key criteria.
Results
The number of visible structures per image was dependent on the different environmental conditions, with large variations observed. The wall colour and the illumination level have a statistically significant effect on the number of visible structures. It was statistically proven that the grey wall colour had a positive effect on the visibility of low contrast detail discs.
Conclusion
Low ambient light, with a grey wall colour and monitors with high specification allow greatest structure visibility. On the contrary, white wall colour around the monitor and high ambient light had a negative impact on technical evaluation of the images during quality control procedures.
Implications for practice
Better standardization of the environmental condition is required in acquisition rooms. Specifically, this research points to the benefit of using a low reflectance wall colour and low illumination level around the monitors.
RadiographyRADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
4.70
自引率
34.60%
发文量
169
审稿时长
63 days
期刊介绍:
Radiography is an International, English language, peer-reviewed journal of diagnostic imaging and radiation therapy. Radiography is the official professional journal of the College of Radiographers and is published quarterly. Radiography aims to publish the highest quality material, both clinical and scientific, on all aspects of diagnostic imaging and radiation therapy and oncology.