{"title":"Arrhythmogenicity of monophasic and biphasic PFA waveforms in a porcine model","authors":"Hanife Tugba Kumru PhD, Lars Mattison PhD, Khaldoun G. Tarakji MD, MPH, Atul Verma MD, Daniel C. Sigg MD, PhD","doi":"10.1111/jce.16480","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objective</h3>\n \n <p>The potential risk of inducing ventricular arrhythmias was explored by employing pulsed field ablation (PFA) through the administration of both monophasic and biphasic waveform deliveries.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>PFA was applied to specific locations in the right ventricle (RV, <i>n</i> = 5 sites) in swine (<i>n</i> = 2), utilizing identical settings with consistent amplitude, pulse width, and a number of pulses for both monophasic and biphasic waveforms. PFA deliveries were precisely timed in 10-ms intervals across the entire T wave.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Using monophasic waveforms, ventricular fibrillation (VF) was induced 7/7 times when PFA pulses were delivered during the T wave. Biphasic waveforms resulted in no ventricular arrhythmias when PFA was delivered across the entire duration of T wave.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Our findings indicate that VF can be consistently induced with monophasic pulse waveforms, but not with biphasic pulse waveforms in a porcine model.</p>\n </section>\n </div>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":"35 12","pages":"2487-2490"},"PeriodicalIF":2.3000,"publicationDate":"2024-11-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/jce.16480","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/jce.16480","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
The potential risk of inducing ventricular arrhythmias was explored by employing pulsed field ablation (PFA) through the administration of both monophasic and biphasic waveform deliveries.
Methods
PFA was applied to specific locations in the right ventricle (RV, n = 5 sites) in swine (n = 2), utilizing identical settings with consistent amplitude, pulse width, and a number of pulses for both monophasic and biphasic waveforms. PFA deliveries were precisely timed in 10-ms intervals across the entire T wave.
Results
Using monophasic waveforms, ventricular fibrillation (VF) was induced 7/7 times when PFA pulses were delivered during the T wave. Biphasic waveforms resulted in no ventricular arrhythmias when PFA was delivered across the entire duration of T wave.
Conclusions
Our findings indicate that VF can be consistently induced with monophasic pulse waveforms, but not with biphasic pulse waveforms in a porcine model.
期刊介绍:
Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.